Analysis of Trace Nitrogen Species in Benzene
Applications | 2007 | Agilent TechnologiesInstrumentation
Monitoring trace nitrogen species in high-purity benzene is critical for petrochemical processing and catalyst performance. As processing catalysts become more selective, allowable limits for impurities are driven ever lower. Reliable detection of sub-ppm nitrogen compounds ensures product quality and informs catalyst life-cycle management.
This overview evaluates a gas chromatographic method coupled with a nitrogen chemiluminescence detector (NCD) for quantifying trace nitrogen species in benzene. Key goals include demonstrating detector specificity against a benzene matrix, establishing calibration linearity, and assessing detection limits and repeatability for common nitrogen contaminants such as acetonitrile, morpholine, 1-methyl-2-pyrrolidinone, and 4-formylmorpholine.
The analysis employs an Agilent 255 Nitrogen Chemiluminescence Detector interfaced to a conventional gas chromatograph. Chromatographic separation uses a 10 m HP-1 capillary column (0.53 mm i.d., 2.65 µm film) with helium carrier at 7.4 mL/min. Injector and detector temperatures are set at 220 °C and 800 °C, respectively. A split ratio of 1:4 and 2 µL injection volume enable reproducible sample introduction. The NCD’s selective chemiluminescence response ensures negligible hydrocarbon interference.
– Blank benzene runs show no response, confirming high selectivity.
– Morpholine yields a linear detector response across the 0.5–5 µg/mL range (as nitrogen) with excellent correlation.
– Signal-to-noise ratios support reliable detection down to 0.2 ppm nitrogen; large-volume injection could further lower limits for high-boiling species.
– Peak shapes improve at higher concentrations and with thicker film or deactivated columns, reducing tailing and enhancing integration precision (≈10 % RSD at 0.5–1 ppm).
The NCD approach delivers high sensitivity and specificity, eliminating hydrocarbon interference common in universal detectors like FID. Its compatibility with existing GC systems allows laboratories to add a second detection channel or maintain dedicated setups. The technique supports quality control in manufacturing and research settings where trace nitrogen levels influence product performance.
Advances may include broader application of large-volume or cryogenic focusing techniques to push detection limits below sub-ppb levels. Development of specialized stationary phases and deactivation treatments can further mitigate adsorption effects and improve peak symmetry. Coupling NCD with automated sampling platforms may expand its utility in high-throughput environments and on-line process monitoring.
The Agilent 255 NCD paired with GC offers a robust, selective solution for trace nitrogen analysis in benzene. Demonstrated linearity, low detection limits, and strong repeatability position this method as a valuable tool for industrial and research laboratories focused on strict impurity control.
GC
IndustriesEnergy & Chemicals
ManufacturerAgilent Technologies
Summary
Importance of the Topic
Monitoring trace nitrogen species in high-purity benzene is critical for petrochemical processing and catalyst performance. As processing catalysts become more selective, allowable limits for impurities are driven ever lower. Reliable detection of sub-ppm nitrogen compounds ensures product quality and informs catalyst life-cycle management.
Objectives and Study Overview
This overview evaluates a gas chromatographic method coupled with a nitrogen chemiluminescence detector (NCD) for quantifying trace nitrogen species in benzene. Key goals include demonstrating detector specificity against a benzene matrix, establishing calibration linearity, and assessing detection limits and repeatability for common nitrogen contaminants such as acetonitrile, morpholine, 1-methyl-2-pyrrolidinone, and 4-formylmorpholine.
Methodology and Instrumentation
The analysis employs an Agilent 255 Nitrogen Chemiluminescence Detector interfaced to a conventional gas chromatograph. Chromatographic separation uses a 10 m HP-1 capillary column (0.53 mm i.d., 2.65 µm film) with helium carrier at 7.4 mL/min. Injector and detector temperatures are set at 220 °C and 800 °C, respectively. A split ratio of 1:4 and 2 µL injection volume enable reproducible sample introduction. The NCD’s selective chemiluminescence response ensures negligible hydrocarbon interference.
Main Results and Discussion
– Blank benzene runs show no response, confirming high selectivity.
– Morpholine yields a linear detector response across the 0.5–5 µg/mL range (as nitrogen) with excellent correlation.
– Signal-to-noise ratios support reliable detection down to 0.2 ppm nitrogen; large-volume injection could further lower limits for high-boiling species.
– Peak shapes improve at higher concentrations and with thicker film or deactivated columns, reducing tailing and enhancing integration precision (≈10 % RSD at 0.5–1 ppm).
Benefits and Practical Applications of the Method
The NCD approach delivers high sensitivity and specificity, eliminating hydrocarbon interference common in universal detectors like FID. Its compatibility with existing GC systems allows laboratories to add a second detection channel or maintain dedicated setups. The technique supports quality control in manufacturing and research settings where trace nitrogen levels influence product performance.
Future Trends and Potential Applications
Advances may include broader application of large-volume or cryogenic focusing techniques to push detection limits below sub-ppb levels. Development of specialized stationary phases and deactivation treatments can further mitigate adsorption effects and improve peak symmetry. Coupling NCD with automated sampling platforms may expand its utility in high-throughput environments and on-line process monitoring.
Conclusion
The Agilent 255 NCD paired with GC offers a robust, selective solution for trace nitrogen analysis in benzene. Demonstrated linearity, low detection limits, and strong repeatability position this method as a valuable tool for industrial and research laboratories focused on strict impurity control.
Instrumentation Used
- Gas chromatograph with split injector
- Agilent 255 Nitrogen Chemiluminescence Detector
- 10 m HP-1 capillary column (0.53 mm i.d., 2.65 µm film)
- Helium carrier gas
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Dual Plasma Sulfur and Nitrogen Chemiluminescence Detectors
2006|Agilent Technologies|Brochures and specifications
Dual Plasma Sulfur and Nitrogen Chemiluminescence Detectors Unsurpassed Stability, Selectivity, and Sensitivity for your GC Analysis Now with New Dual Plasma Technology Overview Dual Plasma Technology The Agilent Technologies Sulfur Chemiluminescence Detector (355 SCD) and Nitrogen Chemiluminescence Detector (255 NCD)…
Key words
sulfur, sulfurplasma, plasmadual, dualfid, fidthiophene, thiophenechemiluminescence, chemiluminescencetechnology, technologymillivolts, millivoltsnotorious, notoriousnitrogen, nitrogenequimolar, equimolarair, aircompounds, compoundsminutes, minutesburner
Agilent Model 255 Nitrogen Chemiluminescence Detector (NCD) Analysis of Adhesive Samples Using the NCD
2007|Agilent Technologies|Applications
Agilent Model 255 Nitrogen Chemiluminescence Detector (NCD) Analysis of Adhesive Samples Using the NCD Technical Overview Introduction The Agilent Model 255 NCD can easily replace a NPD for the analysis of 2-pyrrolidone and 1-vinyl2-pyrrolidinone in adhesive samples. A short-term evaluation…
Key words
ncd, ncdadhesive, adhesivenpd, npdppm, ppmequimolar, equimolardiluted, dilutednumber, numberconcentration, concentrationruns, runsrecalibrated, recalibrateddilution, dilutionmaintenance, maintenancesample, samplefactor, factorevaluation
Analysis of Trace 2-Ethylhexyl Nitrate in Diesel Using ChemiluminescenceDetector (NCD)
2010|Agilent Technologies|Applications
Analysis of Trace 2-Ethylhexyl Nitrate in Diesel Using Chemiluminescence Detector (NCD) Application Brief ChunXiao Wang and Roger Firor HPI An increase in the use of fleet diesel vehicles has helped define requirements for diesel fuel for light duty engines. One…
Key words
diesel, dieselfiror, firorchunxiao, chunxiaoroger, rogerncd, ncdwang, wangchemist, chemistppm, ppmnitrogen, nitrogencetane, cetanehpi, hpifleet, fleetjudge, judgeequimolar, equimolarselectivity
GC-SCD Analysis of Fuels and Petrochemicals
2021|Shimadzu|Guides
GC-SCD Analysis of Fuels and Petrochemicals Application Notebook GC-SCD Analysis of Fuels and Petrochemicals Application Notebook There is perhaps no technique more specific, selective, and sensitive for analysis of sulfur compounds than gas chromatography with sulfur chemiluminescence detection (GC-SCD). As…
Key words
sulfur, sulfurthiophene, thiophenesulfide, sulfidescd, scddisulfide, disulfidedbds, dbdsinsulating, insulatinggas, gascompounds, compoundsbenzene, benzenemercaptan, mercaptanequimolar, equimolarsensitivity, sensitivityarea, areanews