A streamlined laboratory workflow for the analysis of common contaminants according to the U.S. EPA 8270E and 8081B methods using GC-MS/MS
Applications | 2024 | Thermo Fisher ScientificInstrumentation
Environmental laboratories must detect a wide range of semivolatile organic compounds (SVOCs) and organochlorine pesticides in various matrices to protect public health and comply with U.S. EPA regulations. Traditional workflows based on multiple detectors (ECD, NPD, FPD) involve extensive sample preparation, multiple instrument setups, and complex chromatographic separations, which reduce throughput and increase cost. A unified GC-MS/MS approach offers greater selectivity, sensitivity, and automation to address these challenges.
This application note demonstrates a modernized workflow that consolidates EPA 8270E (SVOCs) and EPA 8081B (organochlorine pesticides) analyses onto a single Thermo Scientific TSQ 9610 GC-MS/MS platform with timed-selected reaction monitoring (t-SRM). The goal is to simplify sample preparation, harmonize instrument configurations, and achieve reliable performance across target analytes with improved productivity in high-throughput environmental labs.
• Instrumentation:
• Chromatographic column: TraceGOLD™ TG-5SilMS (30 m × 0.25 mm × 0.25 µm) with integrated 5 m guard
• Acquisition: timed-SRM transitions for hundreds of analytes in a single run
• Automated sample preparation: offline calibration dilutions and internal standard additions using RSH SMART to reduce analyst exposure to solvents
• Instrument performance tests (DFTPP tuning, tailing, degradation) met EPA 8270E criteria without manual tuning adjustments.
• Chromatographic resolution for critical isomer pairs exceeded 50% as required by EPA methods.
• Linearity (0.5–250 µg/L for SVOCs; 0.5–1000 µg/L for pesticides) achieved RF %RSD < 20% for all analytes.
• Instrument detection limits (IDLs) for SVOCs ranged from 5 to 45 ng on column (corresponding to 0.02–0.18 µg/L), with peak area RSD < 10%.
• Pesticide IDLs (n=10 injections at 1 µg/L) were 0.10–0.68 µg/L with area RSD < 20% (except one isomer at 24%).
• Excellent signal-to-noise ratios (S/N >100) observed for BHC isomers at 0.5 µg/L, indicating potential for even lower detection limits.
• Single-platform workflow reduces need for multiple GC detectors, spare parts inventory, and operator training.
• Automated sample prep and instrument maintenance (NeverVent technology, consumable usage alerts) enhance safety and uptime.
• Broad dynamic range of the XLXR detector handles low- and high-concentration analytes in one run.
• Centralized software control across GC, MS, and LC instruments streamlines data handling in regulated environments.
• Method consolidation accelerates sample throughput and cost savings in routine environmental testing.
• Expansion of unified workflows to additional EPA and global environmental methods (e.g., PCBs, emerging contaminants).
• Integration of high-resolution accurate mass (HRAM) detection for nontarget screening and confirmation.
• Further reduction of limits of detection through optimized ion source design and SRM scheduling.
• Remote and cloud-based monitoring of instrument health and sequence execution.
• Application of artificial intelligence and machine learning for automated data interpretation and anomaly detection.
The Thermo Scientific TSQ 9610 GC-MS/MS with timed-SRM and NeverVent™ AEI ion source, combined with TRACE 1610 GC and automated sample handling, provides a robust, sensitive, and simplified workflow for EPA 8270E and 8081B analyses. This unified approach streamlines laboratory operations, enhances data quality, and offers scalability for high-throughput environmental testing.
GC/MSD, GC/MS/MS, GC/QQQ
IndustriesEnvironmental
ManufacturerThermo Fisher Scientific
Summary
Importance of the Topic
Environmental laboratories must detect a wide range of semivolatile organic compounds (SVOCs) and organochlorine pesticides in various matrices to protect public health and comply with U.S. EPA regulations. Traditional workflows based on multiple detectors (ECD, NPD, FPD) involve extensive sample preparation, multiple instrument setups, and complex chromatographic separations, which reduce throughput and increase cost. A unified GC-MS/MS approach offers greater selectivity, sensitivity, and automation to address these challenges.
Study Objectives and Overview
This application note demonstrates a modernized workflow that consolidates EPA 8270E (SVOCs) and EPA 8081B (organochlorine pesticides) analyses onto a single Thermo Scientific TSQ 9610 GC-MS/MS platform with timed-selected reaction monitoring (t-SRM). The goal is to simplify sample preparation, harmonize instrument configurations, and achieve reliable performance across target analytes with improved productivity in high-throughput environmental labs.
Methodology and Instrumentation
• Instrumentation:
- TRACE 1610 GC with user-installable injector/detector modules and NeverVent™ AEI ion source
- iConnect™ split/splitless inlet and TriPlus RSH SMART autosampler for liquid, headspace, and SPME automation
- TSQ 9610 triple quadrupole MS with XLXR™ detector
- Chromeleon™ CDS v7.3.2 with Environmental Analysis Extension Pack for sequence control, data processing, and 21 CFR Part 11 compliance
• Chromatographic column: TraceGOLD™ TG-5SilMS (30 m × 0.25 mm × 0.25 µm) with integrated 5 m guard
• Acquisition: timed-SRM transitions for hundreds of analytes in a single run
• Automated sample preparation: offline calibration dilutions and internal standard additions using RSH SMART to reduce analyst exposure to solvents
Key Results and Discussion
• Instrument performance tests (DFTPP tuning, tailing, degradation) met EPA 8270E criteria without manual tuning adjustments.
• Chromatographic resolution for critical isomer pairs exceeded 50% as required by EPA methods.
• Linearity (0.5–250 µg/L for SVOCs; 0.5–1000 µg/L for pesticides) achieved RF %RSD < 20% for all analytes.
• Instrument detection limits (IDLs) for SVOCs ranged from 5 to 45 ng on column (corresponding to 0.02–0.18 µg/L), with peak area RSD < 10%.
• Pesticide IDLs (n=10 injections at 1 µg/L) were 0.10–0.68 µg/L with area RSD < 20% (except one isomer at 24%).
• Excellent signal-to-noise ratios (S/N >100) observed for BHC isomers at 0.5 µg/L, indicating potential for even lower detection limits.
Benefits and Practical Applications
• Single-platform workflow reduces need for multiple GC detectors, spare parts inventory, and operator training.
• Automated sample prep and instrument maintenance (NeverVent technology, consumable usage alerts) enhance safety and uptime.
• Broad dynamic range of the XLXR detector handles low- and high-concentration analytes in one run.
• Centralized software control across GC, MS, and LC instruments streamlines data handling in regulated environments.
• Method consolidation accelerates sample throughput and cost savings in routine environmental testing.
Future Trends and Opportunities
• Expansion of unified workflows to additional EPA and global environmental methods (e.g., PCBs, emerging contaminants).
• Integration of high-resolution accurate mass (HRAM) detection for nontarget screening and confirmation.
• Further reduction of limits of detection through optimized ion source design and SRM scheduling.
• Remote and cloud-based monitoring of instrument health and sequence execution.
• Application of artificial intelligence and machine learning for automated data interpretation and anomaly detection.
Conclusion
The Thermo Scientific TSQ 9610 GC-MS/MS with timed-SRM and NeverVent™ AEI ion source, combined with TRACE 1610 GC and automated sample handling, provides a robust, sensitive, and simplified workflow for EPA 8270E and 8081B analyses. This unified approach streamlines laboratory operations, enhances data quality, and offers scalability for high-throughput environmental testing.
References
- U.S. EPA SW-846 Test Method 8081B: Organochlorine Pesticides by Gas Chromatography, Rev. 2 (2007).
- U.S. EPA SW-846 Test Method 8270E: Semivolatile Organic Compounds by GC-MS, Rev. 6 (2018).
- U.S. EPA SW-846 Test Method 8141B: Organophosphorus Compounds by Gas Chromatography, Rev. 2 (2007).
- Thermo Fisher Scientific BR52235-EN, TriPlus RSH SMART autosampler brochure.
- Thermo Fisher Scientific BR74090-EN, TRACE 1610 Series GC brochure.
- Thermo Fisher Scientific BR000161-EN, TSQ 9610 GC-MS/MS brochure.
- Thermo Fisher Scientific BR001975-EN, Chromeleon CDS v7.3.2 brochure.
- U.S. Code of Federal Regulations Part 136, Method Detection Limit revisions.
- Thermo Fisher Scientific TN 10494, IDL determination with ISQ LT GC-MS.
- IUPAC Analytical Compendium: Detection Limit definitions.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Staying Ahead in a Rapidly Changing World - Application Compendium
2023|Agilent Technologies|Guides
GC/MS solutions that help analytical labs see more, further, and faster Staying Ahead in a Rapidly Changing World Application Compendium Table of Contents Introduction 3 Analysis of Semivolatile Organic Compounds Using Hydrogen Carrier Gas and HydroInert Ion Source by Gas…
Key words
wide, widefalse, falselinear, linearavg, avgdmrm, dmrmphthalate, phthalatescan, scanacquisition, acquisitiontime, timemin, minbenzo, benzoinlet, inletmethyl, methylcounts, countsfluoranthene
Analysis of Semivolatile Organic Compounds with Hydrogen Carrier Gas and HydroInert Source by Gas Chromatography/Triple Quadrupole Mass Spectrometry (GC/MS/MS)
2022|Agilent Technologies|Applications
Application Note Environmental Analysis of Semivolatile Organic Compounds with Hydrogen Carrier Gas and HydroInert Source by Gas Chromatography/Triple Quadrupole Mass Spectrometry (GC/MS/MS) Author Abstract Angela Smith Henry, PhD Agilent Technologies, Inc. Gas chromatography/mass spectrometry (GC/MS) is integral to the analysis…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluoranthenequadratic, quadraticpass, passaramite, aramitemethanesulfonate, methanesulfonatenitrosodi, nitrosodianthracene, anthracenepyrene, pyrenebenzyl, benzylbutyl, butylparathion, parathionnitroso, nitrosohydroinert
Automated dispersive liquid-liquid micro extraction (DLLME) for GC-MS/MS analysis of semi-volatile compounds in water
2024|Thermo Fisher Scientific|Applications
Application note | 003270 Environmental Automated dispersive liquid-liquid micro extraction (DLLME) for GC-MS/MS analysis of semi-volatile compounds in water Goal Authors To demonstrate the analytical performance of the automated dispersive liquid-liquid Paolo Morelli , Cedric Wissel , 1 2 Paolo…
Key words
hch, hchbenzo, benzodllme, dllmefluoranthene, fluoranthenecounts, countsmin, minethyl, ethylazinphos, azinphospyrene, pyrenesolvent, solventchlordane, chlordaneparathion, parathionpirimicarb, pirimicarbatrazine, atrazineendosulfan
Analysis of Semivolatile Organic Compounds Using Hydrogen Carrier Gas and the Agilent HydroInert Source by Gas Chromatography/Mass Spectrometry
2022|Agilent Technologies|Applications
Application Note Environmental Analysis of Semivolatile Organic Compounds Using Hydrogen Carrier Gas and the Agilent HydroInert Source by Gas Chromatography/Mass Spectrometry Author Angela Smith Henry, PhD Agilent Technologies, Inc. Abstract Gas chromatography/mass spectrometry (GC/MS) is integral to the analysis of…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluoranthenepass, passnitrobenzene, nitrobenzeneanthracene, anthracenearamite, aramitelinear, linearmethanesulfonate, methanesulfonatediallate, diallatepyrene, pyreneazobenzene, azobenzenebenzyl, benzylnitroso, nitrosohydroinert