Clarity Control Module OQ VALIDATION DETECTOR

Manuals | 2024 | DataApexInstrumentation
Software
Industries
Other
Manufacturer
DataApex

Summary

Significance of the Topic



The virtual OQ Validation detector integrated into the Clarity chromatography platform provides a standardized approach to operational qualification (OQ) testing by replaying pre-recorded chromatograms. This method ensures reproducible evaluation of data acquisition workflows without requiring physical detector hardware, reducing downtime and simplifying routine system checks.

Objectives and Study Overview



This application note details the implementation of the OQ Validation detector control module within Clarity. Key objectives include:
  • Describing installation and configuration of the virtual detector
  • Outlining method setup parameters for chromatogram replay
  • Demonstrating report generation of OQ validation runs

The document targets analytical laboratories seeking to automate qualification procedures for liquid or gas chromatography systems.

Used Instrumentation



The virtual detector is purely software-based and interfaces via the DataApex UNI control module. No physical detector is required. System requirements:
  • Clarity chromatography station with LC Control (A24) or GC Control (A23) license
  • DataApex UNI Ruby script DATAAPEXOQVALIDATIONDET.RB

All functions execute within the Clarity environment on Windows platforms.

Methodology



Installation proceeds by adding the OQ Validation detector driver in the System Configuration dialog and assigning it to an instrument tab. Key steps:
  • Select and add the driver via the Available Control Modules dialog
  • Configure digital I/O start and ready signals for synchronization
  • Specify the OQ script path in the DataApex UNI Setup dialog and assign a custom device name and signal units

Method Setup details:
  • Choose a preloaded chromatogram for replay
  • Set replay sample rate (recommended 10 Hz for a 3 min run)
  • Use the Hardware Configuration dialog to verify device name and units

Report Setup enables inclusion of OQ detector settings and parameters in method reports.

Key Findings and Discussion



The virtual detector approach successfully simulates real-time data acquisition, preserving instrument method details in the resulting chromatograms. Replay at 10 Hz generates sufficient data density for qualification criteria. Report integration ensures transparent documentation of OQ parameters, enhancing audit readiness. The module’s reliance on existing UNI infrastructure streamlines deployment across multiple instruments.

Benefits and Practical Applications



Adoption of the OQ Validation detector delivers the following advantages:
  • Elimination of hardware dependencies for routine qualification
  • Faster execution and reduced maintenance overhead
  • Consistent, reproducible test results across systems and laboratories
  • Automated reporting of method and validation parameters for compliance

Industries such as pharmaceuticals, environmental monitoring, and food safety can integrate this module into QA/QC workflows to ensure chromatography systems meet performance specifications.

Future Trends and Potential Applications



Advancements may include integration with cloud-based data management, automated deviation detection through machine learning models, and extension to other virtual instruments (e.g., pumps, autosamplers). Combining virtual validation modules with digital twins of chromatographic systems could further enhance predictive maintenance and remote diagnostics.

Conclusion



The DataApex OQ Validation detector module offers a practical, software-driven solution for operational qualification of chromatography systems. By replaying known chromatograms at controlled sampling rates and embedding method parameters within results, laboratories can streamline validation routines, maintain compliance, and reduce instrument downtime.

References



1. DataApex Ltd., "OQ Validation Detector Control Module Manual," Clarity Station Documentation, Rev. M236/90A, May 2024.
2. DataApex Ltd., DataApex UNI Ruby Script Repository, UTILS/UNI_DRIVERS/DATAAPEX, accessed 2024.
3. Microsoft Corporation, Windows Operating System Requirements for Clarity Chromatography Station.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Clarity Control Module ADVANTEC CHF122SC
ADVANTEC CHF122SC Clarity Control Module ENG Code/Rev.: M161/90B Date: 2024-11-13 DataApex Ltd. Phone: +420 251 013 400 Petrzilkova 2583/13 [email protected] 158 00 Prague 5 www.dataapex.com Czech Republic ® are trademarks of DataApex Ltd. Microsoft® and WindowsTM are Clarity®, DataApex® and…
Key words
fraction, fractionsignal, signalfulfilled, fulfilledvial, vialstart, startstop, stopdetector, detectorclarity, claritydialog, dialogsetup, setupderivative, derivativedecreases, decreasescollector, collectorfirst, firstcollection
Clarity Control Module DEMO GCXGC MODULATOR
DEMO GCXGC MODULATOR Clarity Control Module ENG Code/Rev.: M205/100A Date: 2025-05-12 DataApex Ltd. Phone: +420 251 013 400 Petrzilkova 2583/13 [email protected] 158 00 Prague 5 www.dataapex.com Czech Republic ® are trademarks of DataApex Ltd. Microsoft® and WindowsTM are Clarity®, DataApex®…
Key words
demo, demogcxgc, gcxgcmodulator, modulatorclarity, clarityuni, unidataapex, dataapexdialog, dialogdevice, deviceitem, itemsetup, setupruby, rubycontrol, controlcontents, contentsmodulation, modulationtext
Clarity Control Module CTC PAL3 SYSTEM
CTC PAL3 SYSTEM Clarity Control Module ENG Code/Rev.: M224/100A Date: 2024-10-17 DataApex Ltd. Phone: +420 251 013 400 Petrzilkova 2583/13 [email protected] 158 00 Prague 5 www.dataapex.com Czech Republic ® are trademarks of DataApex Ltd. Microsoft® and WindowsTM are Clarity®, DataApex®…
Key words
clarity, clarityicf, icfconfiguration, configurationlan, lancommand, commandmonitor, monitoragilent, agilentctc, ctcdialog, dialogping, pingsubnet, subnetdevice, devicetab, tabcommunication, communicationversion
Clarity Software - GETTING STARTED
GETTING STARTED Clarity Software ENG Code/Rev.: M006/80L Date: 2/10/2021 Phone: +420 251 013 400 Fax: +420 251 013 401 [email protected] www.dataapex.com DataApex Ltd. Petrzilkova 2583/13 158 00 Prague 5 The Czech Republic ® are trademarks of DataApex Ltd. Microsoft ®…
Key words
getting, gettingstarted, startedclarity, claritywindow, windowstart, startcommand, commandsequence, sequenceyou, youdialog, dialogoutput, outputdigital, digitalopen, openfile, fileicon, iconcolibrick
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike