GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Peak Shape Problems: No Peaks

Technical notes |  | PhenomenexInstrumentation
GC, Consumables, GC columns
Industries
Manufacturer
Phenomenex

Summary

Importance of the Topic


Gas chromatography (GC) is a cornerstone technique for separating and quantifying volatile compounds. When peaks disappear completely or partially, analytical results become unreliable, impacting quality control, research outcomes and regulatory compliance.

Objectives and Study Overview


This technical note investigates three common scenarios of missing peaks in GC runs and proposes practical solutions. By examining causes related to sample injection, column integrity and detector performance, the study aims to guide analysts through systematic troubleshooting.

Methodology and Instrumentation Used


The approach involves categorizing symptoms into three cases and mapping each to likely causes:
  • All peaks missing
  • No peaks after solvent peak
  • Some peaks missing

Key instrumentation components considered include the syringe, inlet/liner, column and detector (e.g., flame ionization detector), along with carrier gas flow control.

Main Results and Discussion


1. All peaks missing:
  • Defective syringe → replace with a proven device.
  • Leaking septum or inlet leaks → locate and fix leaks.
  • Carrier gas flow issues → adjust flow rate.
  • Broken or misinstalled column → trim or reconnect with a press-fit, or reinstall.
  • Detector malfunction or disconnection → verify detector operation and output connection.
  • Incorrect injector temperature → verify with a thermometer; adjust within column limits.

2. No peaks after solvent peak:
  • Excessive sample volume → reduce injection amount or increase split ratio.
  • FID flame extinguished by solvent → ensure correct detector temperature.
  • High carrier gas flow → lower flow rate.
  • Column too hot → lower oven temperature.
  • Inadequate separation of solvent and analytes → select alternative solvent or column.

3. Some peaks missing:
  • Active compounds interacting with inlet or column surfaces → clean or replace inlet liner, use inert columns.

Overall, detector performance is critical but injection hardware and column condition must also be evaluated.

Benefits and Practical Applications


Implementing these troubleshooting steps helps laboratories restore GC performance quickly, minimize downtime and maintain data integrity. The guidance supports method development, routine QC/QA workflows and instrument maintenance protocols.

Future Trends and Potential Applications


Advances in inert surface technologies, real-time detector diagnostics and AI-driven fault detection promise to automate troubleshooting. Emerging sensor integration may enable predictive maintenance and reduce manual interventions.

Conclusion


A systematic evaluation of injection, column and detector factors effectively resolves missing peak issues. Adopting these best practices enhances reliability and efficiency in GC analyses.

References


No external references were provided in the source document.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
GC TROUBLESHOOTING GUIDE
GC TROUBLESHOOTING GUIDE
2017|Phenomenex|Guides
GC TROUBLESHOOTING GUIDE www.phenomenex.com/GC Access GC Resources 24/7 Explore Online Tips & Tools You asked for a better online experience. One filled with technical resources, easy-to-find products, and useful tools to make your GC work easier. Good news… it’s here!…
Key words
phenomenex, phenomenexsymptom, symptomremedy, remedycolumn, columnproblems, problemssuggested, suggestedcause, causepolar, polarpeak, peaktemperature, temperaturepossible, possiblebleed, bleedseptum, septumpeaks, peaksshape
Capillary GC Troubleshooting Guide: How to Locate Problems and Solve Them
Bulletin 853B Capillary GC Troubleshooting Guide: How to Locate Problems and Solve Them The real task in correcting a problem with your capillary GC system is identifying the cause of the problem without wasting time. The systematic approach to troubleshooting…
Key words
supelco, supelcoproblem, problemcolumn, columnsymptom, symptomremedy, remedynormal, normalcause, causepurifier, purifierseptum, septumgas, gastest, testpossible, possiblerecorder, recordercapillary, capillarydetector
Packed Column GC Troubleshooting Guide: How to Locate Problems and Solve Them
595 North Harrison Road Bellefonte, PA 16823-0048 USA Telephone 800-247-6628 ● 814-359-3441 Fax 800-447-3044 ● 814-359-3044 email: [email protected] sigma-aldrich.com/supelco Bulletin 792C Packed Column GC Troubleshooting Guide: How to Locate Problems and Solve Them By applying a systematic approach to troubleshooting,…
Key words
problem, problemsupelco, supelconormal, normalsymptom, symptomremedy, remedycolumn, columnpurifier, purifierdetector, detectorrecorder, recordercause, causeglass, glasstoo, toogas, gaspossible, possibletemperature
Packed Column GC Troubleshooting Guide: How to Locate Problems and Solve Them
Bulletin 792C Packed Column GC Troubleshooting Guide: How to Locate Problems and Solve Them 792C By applying a systematic approach to troubleshooting, you can solve many GC problems on your own. The real task is identifying the cause of a…
Key words
problem, problemsupelco, supelconormal, normalsymptom, symptomremedy, remedycolumn, columnpurifier, purifierdetector, detectorcause, causegas, gasrecorder, recordertoo, tooglass, glasspossible, possiblewrench
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike