Sub-ppt Atmospheric Measurements Using PTV-GC-FID and Real-Time Digital Signal Processing
Applications | | GL SciencesInstrumentation
Accurate measurement of trace-level atmospheric hydrocarbons is vital for understanding oxidative capacity and chemical processes in clean environments such as marine air. Traditional grab-sample approaches and analogue filtering have limited sensitivity and integration reliability at sub-ppt concentrations.
This study evaluates the coupling of a digital signal processing (DSP) unit with a programmed temperature vaporisation GC-FID system to enable reliable in-situ sub-ppt hydrocarbon detection. Performance was assessed using standard mixtures and continuous field sampling at Mace Head Observatory.
Air samples (700 mL) were drawn over a cooled adsorbent trap (70 mL min⁻¹), thermally desorbed at 400 °C, and separated by temperature-programmed GC. The FID electrometer output was digitized at high sampling rate and filtered to remove high-frequency noise without peak broadening.
Analysis of a ppb-level hydrocarbon mixture (27 compounds) showed a minor 2–5 % reduction in peak height after DSP, with integration accuracy correlation R² = 0.9995. In clean marine air, signal-to-noise ratios improved from 2:1 to better than 6:1. A six-day field test with hourly samples of propane (80–500 ppt) yielded R² ≈ 1 and a unity slope, while 1,3-butadiene (2–20 ppt) gave R² = 0.88 and slope 0.75, revealing raw data overestimation of ~25 % at low S/N. Manual reprocessing dropped from 23 to 6 runs when using DSP.
Integration of DSP directly into commercial detectors, expansion to other analytical platforms (e.g., LC, mass spectrometry), and deployment in real-time atmospheric monitoring networks are promising developments to further improve trace-level detection accuracy.
Applying digital signal processing to GC-FID electrometer outputs significantly enhances trace-level hydrocarbon detection by reducing noise, lowering detection limits, and improving integration accuracy, thereby supporting robust automated atmospheric measurements.
GC
IndustriesEnvironmental
ManufacturerGL Sciences
Summary
Significance of the Topic
Accurate measurement of trace-level atmospheric hydrocarbons is vital for understanding oxidative capacity and chemical processes in clean environments such as marine air. Traditional grab-sample approaches and analogue filtering have limited sensitivity and integration reliability at sub-ppt concentrations.
Objectives and Study Overview
This study evaluates the coupling of a digital signal processing (DSP) unit with a programmed temperature vaporisation GC-FID system to enable reliable in-situ sub-ppt hydrocarbon detection. Performance was assessed using standard mixtures and continuous field sampling at Mace Head Observatory.
Instrumentation Used
- PTV injector (OPTIC 400) with activated charcoal trap, cooled to −20 °C
- GC-94 gas chromatograph with 50 m × 0.53 mm i.d. Al₂O₃ PLOT column
- Flame ionisation detector operating at 10 Hz sampling frequency
- Digital signal processing unit (ID/10) in parallel with EZChrom PC data capture
Methodology
Air samples (700 mL) were drawn over a cooled adsorbent trap (70 mL min⁻¹), thermally desorbed at 400 °C, and separated by temperature-programmed GC. The FID electrometer output was digitized at high sampling rate and filtered to remove high-frequency noise without peak broadening.
Results and Discussion
Analysis of a ppb-level hydrocarbon mixture (27 compounds) showed a minor 2–5 % reduction in peak height after DSP, with integration accuracy correlation R² = 0.9995. In clean marine air, signal-to-noise ratios improved from 2:1 to better than 6:1. A six-day field test with hourly samples of propane (80–500 ppt) yielded R² ≈ 1 and a unity slope, while 1,3-butadiene (2–20 ppt) gave R² = 0.88 and slope 0.75, revealing raw data overestimation of ~25 % at low S/N. Manual reprocessing dropped from 23 to 6 runs when using DSP.
Benefits and Practical Applications
- Lowered detection limits from ~3–5 ppt to ~0.5–0.8 ppt for C6 species
- Improved peak integration reliability by up to 4-fold for low S/N levels
- Reduced manual intervention and enhanced data quality for field monitoring
Future Trends and Opportunities
Integration of DSP directly into commercial detectors, expansion to other analytical platforms (e.g., LC, mass spectrometry), and deployment in real-time atmospheric monitoring networks are promising developments to further improve trace-level detection accuracy.
Conclusion
Applying digital signal processing to GC-FID electrometer outputs significantly enhances trace-level hydrocarbon detection by reducing noise, lowering detection limits, and improving integration accuracy, thereby supporting robust automated atmospheric measurements.
References
- A.C. Lewis et al., J. Chem. Soc. Faraday Trans., 1997, 93, 2921.
- P.G. Simmonds et al., Atmos. Environ., 1996, 30, 4041.
- A.V. Oppenheim, R.W. Schafer, Digital Signal Processing, Prentice Hall, 1975.
- A. Antoniou, Digital Filters: Analysis, Design and Applications, 2nd ed., McGraw-Hill, 1993.
- K.D. Jackson et al., J. Automatic Chemistry, 1997, 19, 145-152.
- A.C. Lewis et al., J. High Res. Chrom., 1996, 19, 686.
- Thomas Swan, Consett, UK, Internal Report, 1997.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Atmospheric Monitoring of Volatile Organic Compounds Using Programmed Temperature Vaporization Injection
|Thermo Fisher Scientific|Applications
Atmospheric Monitoring of Volatile Organic Compounds Using Programmed Temperature Vaporization Injection Alastair C. Lewis*, Keith D. Bartle, James B. McQuaid, Michael J. Pilling, and Paul W. Seakins School of Chemistry, University of Leeds, Leeds, LS2 9JT, UK Peter Ridgeon Ai…
Key words
vocs, vocssorbent, sorbenttube, tubeatmospheric, atmospherictrap, trapptv, ptvvaporization, vaporizationready, readydesorbing, desorbingforest, forestonto, ontomonitoring, monitoringprogrammed, programmedisoprene, isopreneinjector
In-Situ Field Measurements of Isoprene, Its Oxidation Products and Selected Monoterpenes in a Eucalyptus Forest
|Thermo Fisher Scientific|Applications
Application Note No. 021 In-Situ Field Measurements of Isoprene, Its Oxidation Products and Selected Monoterpenes in a Eucalyptus Forest AC Lewis, P.W Seakins, A.M Denha, K.D Bartle, M.J Pilling School of Chemistry, University of Leeds, Leeds, LS2 9JT, U.K. Abstract…
Key words
forest, forestisoprene, isoprenecanopy, canopyeucalyptus, eucalyptusbiogenic, biogenicambient, ambientsite, sitemonoterpenes, monoterpenessorbent, sorbentsampling, samplingtree, treeanthrpogenic, anthrpogenichimalaya, himalayamacr, macrpennsyvannia
Applications of PTV Injectors for Problem Solving in the Petrochemical Industry Part 1:- Thermal Desorption with GC and GC-MS
|Agilent Technologies|Applications
Application Note No. 039 Applications of PTV Injectors for Problem Solving in the Petrochemical Industry Part 1:- Thermal Desorption with GC and GC-MS 1 J.S. Lancaster, 1T.P. Lynch, 1A K Grosser, and 2P.G. McDowell 1 Hull Research & Technology Centre,…
Key words
ptv, ptvplasticisers, plasticiserspolyethylene, polyethylenepvc, pvcdesorption, desorptiondiup, diupoxygenated, oxygenatedphthalate, phthalatedup, dupthermal, thermalflue, fluefilm, filminjection, injectionprogramme, programmeinjectors
Heating Up Simulated Sistillation
|GL Sciences|Applications
Allen K. Vickers Sr. Applications Chemist The American Society for Testing and Materials (ASTM) has established guidelines for simulated distillation (SimDist) analyses, which include samples that have boiling points in the range of about -44°F to 1139°F. These include ASTM…
Key words
boiling, boilinghtsd, htsdpoint, pointpolyimide, polyimidedis, disupper, upperhas, hastemperature, temperaturethick, thickdistribution, distributiondbht, dbhtsimulated, simulatedend, endaebp, aebppatience