Analysis of Phenols
Applications | 2023 | ShimadzuInstrumentation
Phenolic compounds are common environmental contaminants and intermediates in many industrial processes. Their potential toxicity and persistence in water and soil call for reliable analytical methods capable of detecting trace levels. Gas chromatography coupled with flame ionization detection (GC-FID) provides a robust approach for separating and quantifying multiple phenols in a single run, supporting environmental monitoring and quality control.
This work evaluates an analytical procedure for the simultaneous analysis of eleven phenolic compounds at 200 ng/mL using an inert mid-polarity capillary GC column. The primary objectives are to demonstrate baseline separation, assess reproducibility, and verify a temperature program suitable for routine environmental testing.
A 1 µL sample of mixed phenols (200 ng/mL) is injected in split mode (1:40). The GC temperature program starts at 50 °C (4 min hold), then ramps at 6 °C/min to 250 °C. Helium is used as the carrier gas at 20 cm/s linear velocity. The injector operates at 250 °C, while the valve remains at ambient temperature. Flame ionization detection records the eluting compounds.
The selected temperature program achieves clear resolution of all eleven phenolic analytes, including positional isomers and nitro-substituted derivatives. Peak shapes are sharp and symmetrical, indicating minimal adsorption on the inert column surface. The method demonstrates consistent retention times and reproducible peak areas, suitable for quantitative analysis at low nanogram levels.
Advancements may include coupling GC to mass spectrometry (GC-MS) for enhanced selectivity and confirmatory analysis. Further miniaturization and faster temperature ramps could reduce analysis time. Automated sample preparation and on-line extraction systems will streamline workflows for high-throughput monitoring of phenolic pollutants.
The presented GC-FID method using the SH-200 column offers a reliable and reproducible approach for the simultaneous analysis of multiple phenolic compounds at trace levels. Its robustness and sensitivity make it well suited for environmental laboratories and industrial quality control settings.
GC, Consumables, GC columns
IndustriesEnvironmental
ManufacturerShimadzu
Summary
Significance of the topic
Phenolic compounds are common environmental contaminants and intermediates in many industrial processes. Their potential toxicity and persistence in water and soil call for reliable analytical methods capable of detecting trace levels. Gas chromatography coupled with flame ionization detection (GC-FID) provides a robust approach for separating and quantifying multiple phenols in a single run, supporting environmental monitoring and quality control.
Goals and overview of the study
This work evaluates an analytical procedure for the simultaneous analysis of eleven phenolic compounds at 200 ng/mL using an inert mid-polarity capillary GC column. The primary objectives are to demonstrate baseline separation, assess reproducibility, and verify a temperature program suitable for routine environmental testing.
Methodology and analytical procedure
A 1 µL sample of mixed phenols (200 ng/mL) is injected in split mode (1:40). The GC temperature program starts at 50 °C (4 min hold), then ramps at 6 °C/min to 250 °C. Helium is used as the carrier gas at 20 cm/s linear velocity. The injector operates at 250 °C, while the valve remains at ambient temperature. Flame ionization detection records the eluting compounds.
Used Instrumentation
- Gas chromatograph: Shimadzu GC with FID detector
- Column: SH-200 inert mid-polarity capillary column (30 m × 0.32 mm I.D., 0.25 µm film thickness; P/N 227-36184-01)
- Carrier gas: Helium, 20 cm/s linear velocity
- Injection mode: Split (1:40), 1 µL injection volume; injector temperature 250 °C
Main results and discussion
The selected temperature program achieves clear resolution of all eleven phenolic analytes, including positional isomers and nitro-substituted derivatives. Peak shapes are sharp and symmetrical, indicating minimal adsorption on the inert column surface. The method demonstrates consistent retention times and reproducible peak areas, suitable for quantitative analysis at low nanogram levels.
Benefits and practical applications
- High sensitivity: FID provides stable response for a wide range of phenols.
- Inert column surface: Reduces analyte adsorption and improves reproducibility.
- Efficient separation: Eleven compounds resolved in a single 35-minute run.
- Routine monitoring: Applicable to environmental water analysis, industrial effluent testing, and quality assurance.
Future trends and potential applications
Advancements may include coupling GC to mass spectrometry (GC-MS) for enhanced selectivity and confirmatory analysis. Further miniaturization and faster temperature ramps could reduce analysis time. Automated sample preparation and on-line extraction systems will streamline workflows for high-throughput monitoring of phenolic pollutants.
Conclusion
The presented GC-FID method using the SH-200 column offers a reliable and reproducible approach for the simultaneous analysis of multiple phenolic compounds at trace levels. Its robustness and sensitivity make it well suited for environmental laboratories and industrial quality control settings.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Analysis of Amines & Phenols
2023|Shimadzu|Applications
ERAS-1000-0420 GC FID SH Series SH-5 Amine Analysis of Amines & Phenols 420 Keywords: environmental analysis, food, chemical 1. Diethylamine 2. Pyridine 3. Morpholine 4. Phenol 5. Aniline 6. 2-Chlorophenol 7. Diethylenetriamine 8. Octylamine 9. 1-Methyl-2-pyrrolidinone 10. 2-Nitrophenol 11. 2,6-Dimethylaniline…
Key words
fid, fidphenols, phenolsamines, aminessplit, splitseries, seriesratio, ratiocarrier, carriersensitivity, sensitivitydetector, detectorinstrument, instrumentgas, gascolumn, columnanalysis
Analysis of Free Fatty acids (SH-200)
2023|Shimadzu|Applications
ERAS-1000-0508 GC FID SH Series SH-200 Analysis of Free Fatty acids 508 Keywords: Organic acid, Food science 1. Acetic acid 2. Propionic acid 3. Isobutyric acid 4. n-Butyric acid 5. Isovaleric acid 6. n-Valeric acid 7. Isocaproic acid 8. Caproic…
Key words
vent, ventfatty, fattyfid, fidacids, acidssplit, splitseries, seriesfree, freecarrier, carrierdetector, detectorinstrument, instrumentgas, gascolumn, columnanalysis
Analysis of Aldehydes
2023|Shimadzu|Applications
ERAS-1000-0412 GC FID SH Series SH-Wax Analysis of Aldehydes 412 Keywords: environment, primary aldehydes 1. Ethanal 2. Propanal 3. Butenal 4. Pentanal 5. Hexanal 6. Heptanal 7. Octanal 8. Nonanal 9. Decanal 10. Undecanal (On-column conc.: 250 ng) Instrument :…
Key words
aldehydes, aldehydeswax, waxfid, fidsplit, splitseries, seriesratio, ratiocarrier, carrierdetector, detectorinstrument, instrumentgas, gascolumn, columnanalysis
Analysis of Residual solvents USP 467 Class 2 MixB in pharmaceuticals
2023|Shimadzu|Applications
ERAS-1000-0490 GC FID SH Series SH-PolarWax Analysis of Residual solvents USP 467 Class 2 MixB in pharmaceuticals 490 Keywords: Procedure B, Water soluble 1 8 1. Hexane 2. 1,2-Dimethoxyethane 3. Trichloroethene 4. Chloroform 5. Methylbutylketone 6. Nitromethane 7. Pyridine 8.…
Key words
vial, vialfid, fidpolarwax, polarwaxtime, timestirring, stirringpressure, pressureflush, flushrate, rateloading, loadingflow, flowheat, heatneedle, needlesplit, splitair, airoff