Streamlining environmental monitoring: Advanced analysis of PBDEs using triple quadrupole GC-MS/MS
Applications | 2025 | Thermo Fisher ScientificInstrumentation
Polybrominated diphenyl ethers are persistent flame retardants that accumulate in environmental compartments and biota. Due to their toxicity and longevity, regulatory agencies worldwide set strict limits for their presence in water, soil, and food. Rapid and reliable analytical methods are essential for environmental monitoring, risk assessment, and regulatory compliance. Consolidating multiple analyses into a single streamlined configuration enhances laboratory efficiency and data consistency.
The primary goal of the study was to implement a unified gas chromatography tandem mass spectrometry configuration capable of analyzing a broad spectrum of PBDE congeners in compliance with European Standard 16694 2025 and other international regulations. By employing one column and one MS instrument, the method aims to reduce operational complexity, cut turnaround times, and maintain high sensitivity and selectivity for both regulated PBDEs and emerging brominated flame retardants.
The analytical method uses a TRACE 1610 GC coupled to a TSQ 9610 triple quadrupole mass spectrometer with NeverVent technology and an advanced electron ionization source. A TraceGOLD TG-Contaminants column (15 m × 0.25 mm × 0.1 µm) provides high inertness and resolution. A large volume programmable temperature vaporizing injector ensures precise sample introduction. The analysis runs in less than 17 minutes in selected reaction monitoring mode. Calibration in iso-octane with analytical protectant covers 25 compounds over a range from 0.02 to 50 micrograms per liter.
Linearity was assessed over five days using five calibration curves for each PBDE congener. All calibration functions met R2 values above 0.990. Limits of quantification as low as 0.02 micrograms per liter were achieved for key congeners. Relative amount deviation remained below 40 percent at LOQ and below 20 percent at higher levels. Chromatographic resolution allowed baseline separation of critical isomeric pairs within the 17-minute run. Signal to noise ratios at LOQ levels demonstrated excellent sensitivity for late-eluting and high molecular weight analytes such as PBDE 209 and DBDPE.
Consolidation of multiple regulatory methods into one GC-MS/MS platform reduces sample preparation and instrument downtime. The single configuration supports a wide calibration range and meets global compliance standards. Laboratories gain higher throughput, lower consumable costs, and flexible inclusion of emerging micropollutants and novel brominated flame retardants without changing hardware.
Analytical demands will expand to cover next generation flame retardants and transformation products. Future developments may include shorter columns with advanced stationary phases, automated data processing workflows, and integration of high resolution MS for non-target screening. Miniaturized injection systems and AI-driven spectral deconvolution will further enhance throughput and analytical depth.
The unified triple quadrupole GC-MS/MS method provides robust, sensitive, and regulatory compliant analysis of PBDEs and related flame retardants. Its single-column approach reduces complexity and operational costs while delivering high precision and low detection limits. This streamlined platform is well suited for environmental laboratories seeking efficient monitoring solutions.
GC/MSD, GC/MS/MS, GC/QQQ
IndustriesEnvironmental
ManufacturerThermo Fisher Scientific
Summary
Significance of the topic
Polybrominated diphenyl ethers are persistent flame retardants that accumulate in environmental compartments and biota. Due to their toxicity and longevity, regulatory agencies worldwide set strict limits for their presence in water, soil, and food. Rapid and reliable analytical methods are essential for environmental monitoring, risk assessment, and regulatory compliance. Consolidating multiple analyses into a single streamlined configuration enhances laboratory efficiency and data consistency.
Objectives and overview
The primary goal of the study was to implement a unified gas chromatography tandem mass spectrometry configuration capable of analyzing a broad spectrum of PBDE congeners in compliance with European Standard 16694 2025 and other international regulations. By employing one column and one MS instrument, the method aims to reduce operational complexity, cut turnaround times, and maintain high sensitivity and selectivity for both regulated PBDEs and emerging brominated flame retardants.
Methodology and instrumentation
The analytical method uses a TRACE 1610 GC coupled to a TSQ 9610 triple quadrupole mass spectrometer with NeverVent technology and an advanced electron ionization source. A TraceGOLD TG-Contaminants column (15 m × 0.25 mm × 0.1 µm) provides high inertness and resolution. A large volume programmable temperature vaporizing injector ensures precise sample introduction. The analysis runs in less than 17 minutes in selected reaction monitoring mode. Calibration in iso-octane with analytical protectant covers 25 compounds over a range from 0.02 to 50 micrograms per liter.
Main results and discussion
Linearity was assessed over five days using five calibration curves for each PBDE congener. All calibration functions met R2 values above 0.990. Limits of quantification as low as 0.02 micrograms per liter were achieved for key congeners. Relative amount deviation remained below 40 percent at LOQ and below 20 percent at higher levels. Chromatographic resolution allowed baseline separation of critical isomeric pairs within the 17-minute run. Signal to noise ratios at LOQ levels demonstrated excellent sensitivity for late-eluting and high molecular weight analytes such as PBDE 209 and DBDPE.
Benefits and practical applications
Consolidation of multiple regulatory methods into one GC-MS/MS platform reduces sample preparation and instrument downtime. The single configuration supports a wide calibration range and meets global compliance standards. Laboratories gain higher throughput, lower consumable costs, and flexible inclusion of emerging micropollutants and novel brominated flame retardants without changing hardware.
Future trends and opportunities
Analytical demands will expand to cover next generation flame retardants and transformation products. Future developments may include shorter columns with advanced stationary phases, automated data processing workflows, and integration of high resolution MS for non-target screening. Miniaturized injection systems and AI-driven spectral deconvolution will further enhance throughput and analytical depth.
Conclusion
The unified triple quadrupole GC-MS/MS method provides robust, sensitive, and regulatory compliant analysis of PBDEs and related flame retardants. Its single-column approach reduces complexity and operational costs while delivering high precision and low detection limits. This streamlined platform is well suited for environmental laboratories seeking efficient monitoring solutions.
Used instrumentation
- TRACE 1610 Gas Chromatograph
- TSQ 9610 Triple Quadrupole Mass Spectrometer with AEI source
- TraceGOLD TG-Contaminants GC column 15 m × 0.25 mm × 0.1 µm
- iConnect Programmable Temperature Vaporizing injector
- TriPlus RSH SMART autosampler
- SureSTART 2 mL glass vials and screw caps
- High-temperature transfer line nut
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Low-level consistent analysis of PBDEs in environmental and food matrices using triple quadrupole GC-MS/MS
2021|Thermo Fisher Scientific|Applications
Application note | 000493 Environmental Low-level consistent analysis of PBDEs in environmental and food matrices using triple quadrupole GC-MS/MS Authors Goal Kjell Hope , Giulia Riccardino , The aim of this application note is to demonstrate the performance of the…
Key words
pbdes, pbdesntr, ntrcalculated, calculatedation, ationpbde, pbdecon, conmin, minloq, loqinjection, injectiontracegold, tracegoldidl, idlensured, ensuredmea, meaoret, oreticonnect
Fast, ultra-sensitive analysis of PBDEs in food using advanced GC-MS/MS technology
2018|Thermo Fisher Scientific|Applications
APPLICATION NOTE 10674 Fast, ultra-sensitive analysis of PBDEs in food using advanced GC-MS/MS technology Authors Aaron Lamb1, Cristian Cojocariu1, and Sean Panton2 Thermo Fisher Scientific, Runcorn, UK 1 2 Fera Science, Ltd., York, UK Keywords Food safety, polybrominated diphenyl ethers…
Key words
pbde, pbdedeviation, deviationhrms, hrmsaei, aeipbdes, pbdescounts, countsratio, ratioion, ionpenta, pentatetra, tetrabde, bdehexa, hexapass, passcongeners, congenershepta
Improved laboratory productivity with a single GC-MS/MS configuration for multipurpose environmental analysis
2024|Thermo Fisher Scientific|Applications
Application note | 002866 Environmental Improved laboratory productivity with a single GC-MS/MS configuration for multipurpose environmental analysis Authors Goal Delphine Thomas, Bénédicte Gauriat This application note will highlight a single instrument configuration, based on triple Thermo Fisher Scientific Customer Solution…
Key words
counts, countsmin, mintracegold, tracegoldcontaminants, contaminantspesticides, pesticidesderivatives, derivativesmicropollutants, micropollutantssrm, srmthermo, thermotime, timecompounds, compoundsclasses, classesallows, allowsaei, aeipbde
Overcoming analytical challenges for polybrominated diphenyl ethers (PBDEs) analysis in environmental samples using gas chromatography – Orbitrap mass spectrometry
2018|Thermo Fisher Scientific|Applications
APPLICATION NOTE 10644 Overcoming analytical challenges for polybrominated diphenyl ethers (PBDEs) analysis in environmental samples using gas chromatography – Orbitrap mass spectrometry Authors Goal Jane Cooper,1 Jordi Parera,2 Esteban Abad,2 Richard Law,¹ and Cristian Cojocariu1 1 Thermo Fisher Scientific, Runcorn,…
Key words
bde, bdepbdes, pbdespbde, pbdecounts, countsexactive, exactivebdes, bdesisotopic, isotopicquantification, quantificationwere, weredibdes, dibdesmonobde, monobdetribdes, tribdesnumber, numberheptabdes, heptabdesenvironmental