Importance of the Topic


Volatile organic compounds (VOCs) in industrial wastewater pose significant environmental and health hazards when released untreated. Monitoring these contaminants ensures regulatory compliance under the U.S. Clean Water Act and protects aquatic ecosystems and public health. Advances in instrumentation and sample preparation enhance sensitivity, throughput, and laboratory efficiency for VOC analysis.

Objectives and Study Overview


This study demonstrates a complete analytical workflow for quantifying 31 target VOCs in wastewater according to U.S. EPA Method 624.1. Key objectives included:
  • Establishing method linearity over 0.5–100 ppb.
  • Determining method detection limits (MDLs) and calculation of relative standard deviations (RSDs).
  • Verifying accuracy and precision via an Initial Demonstration of Capability (IDC).
  • Assessing long-term robustness over 209 injections without user intervention.

Methodology and Instrumentation


Sample preparation involved seven-point aqueous calibration standards with internal and surrogate standards at 25 ppb. A Teledyne LABS Tekmar Lumin Purge & Trap concentrator paired with the AQUATek LVA autosampler removed moisture and preconcentrated VOCs. Separation was achieved on a Thermo Scientific TRACE 1610 GC with a TraceGOLD TG-VMS column and HeSaver-H2Safer split/splitless inlet. Detection used a Thermo Scientific ISQ 7610 single quadrupole MS in full-scan mode (35–260 amu) with optional SIM for enhanced selectivity. Chromeleon CDS controlled the entire workflow.

Results and Discussion


  • Linearity: All compounds exhibited RSDs < 20% over 0.5–100 ppb, meeting EPA criteria.
  • MDLs: Determined from seven replicates of 0.5 ppb standards; most compounds showed MDLs < 0.15 ppb.
  • Accuracy and Precision: Mid-point checks (20 ppb, n=7) yielded RSDs < 20% and recoveries of 81–103%.
  • Robustness: 209 consecutive injections over three days produced an average RSD of 6.5% without maintenance or user intervention.
  • Helium Saver Impact: HeSaver-H2Safer technology reduced helium consumption by fourfold, extending cylinder lifetime and lowering costs.

Benefits and Practical Applications


The described method offers high sensitivity, reproducibility, and regulatory compliance for routine wastewater monitoring. Automated purge-and-trap sample handling and integrated software control streamline operations, minimize water on the column, and maximize throughput in environmental testing and contract laboratories.

Future Trends and Opportunities


Emerging trends include:
  • Increased adoption of automated, self-optimizing GC-MS systems to further reduce downtime.
  • Expansion of purge-and-trap protocols to cover a broader range of polar and semi-volatile analytes.
  • Integration of data analytics and real-time reporting for rapid decision making in wastewater treatment plants.

Conclusion


The combined Tekmar Lumin P&T – AQUATek LVA autosampler with Thermo Scientific TRACE 1610 GC and ISQ 7610 MS fulfills U.S. EPA Method 624.1 requirements, delivering robust, sensitive, and cost-effective VOC analysis in wastewater. Its automation and helium-saving features support high-throughput environmental testing.

Reference


  1. U.S. EPA. Control of Volatile Organic Compound Emissions from Industrial Wastewater. EPA453/D-93-056.
  2. U.S. EPA. Clean Water Act Overview. 33 U.S.C. §1251 et seq.
  3. Thermo Fisher Scientific. AppsLab Library for EPA Method 624.1.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Analysis of volatile organic compounds in drinking water according to U.S. EPA Method 524.4
Application note | 003028 Environmental Analysis of volatile organic compounds in drinking water according to U.S. EPA Method 524.4 Goal Authors Adam Ladak , Terry Jeffers , and Demonstration of an analytical method that meets the requirements outlined in U.S.…
Key words
lin, linavgcalfact, avgcalfactaquatek, aquatekether, etherlva, lvabutylbenzene, butylbenzenemethyl, methylbutyl, butylbake, bakeprecision, precisionethyl, ethylamyl, amyllumin, luminchloride, chloridepurge
Adhering to U.S. EPA Method 524.2 for the analysis of volatile organic compounds in drinking water
Application note | 002952 Environmental Adhering to U.S. EPA Method 524.2 for the analysis of volatile organic compounds in drinking water Goal Authors Adam Ladak , Terry Jeffers , and Demonstration of a routine analytical method that meets the requirements…
Key words
aquatek, aquateklumin, luminlva, lvabutylbenzene, butylbenzenetekmar, tekmarmdl, mdlprecision, precisionmethyl, methylbake, bakepurge, purgechloride, chloridemethacrylate, methacrylateppb, ppbtert, tertvariable
Trace analysis of epichlorohydrin in drinking water using GC-MS coupled with purge and trap
Application note | 003894 Environmental Trace analysis of epichlorohydrin in drinking water using GC-MS coupled with purge and trap Authors Goal Adam Ladak1, Terry Jeffers2, and Amy Nutter 3 Thermo Fisher Scientific, UK 1 Demonstration of an analytical method for…
Key words
epichlorohydrin, epichlorohydrinlva, lvaaquatek, aquatekech, echpurge, purgelumin, luminbake, bakevariable, variabletekmar, tekmardrinking, drinkingdesorb, desorbtrap, trapmandated, mandatedtime, timefluorobenzene
US EPA Method 624 with the Teledyne Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MSD
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1904 P+T VOC ENV US EPA Method 624 with the Teledyne Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MSD Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA…
Key words
aquatek, aquateklva, lvatemp, temppurge, purgebake, bakeconcentrator, concentratordesorb, desorblumin, luminvariable, variabletime, timetekmar, tekmarteledyne, teledyneppb, ppbtable, tablesurr
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike