Novel Column Chemistry Raises the Bar on Sensitivity and Data Accuracy in the Analysis of Semivolatile Organic Compounds
Applications | 2024 | Agilent TechnologiesInstrumentation
The analysis of semivolatile organic compounds (SVOCs) by gas chromatography mass spectrometry is crucial for environmental monitoring and regulatory compliance. Advances in column chemistry directly improve sensitivity data accuracy and method robustness when detecting trace level pollutants in complex matrices.
This study evaluates a novel low bleed inert GC column coupled with an advanced ion source on a triple quadrupole GC MS system. It compares performance metrics such as bleed reduction inertness thermal stability sensitivity and chromatographic consistency against conventional columns under high throughput and challenging sample conditions.
This evaluation employed an Agilent 8890 gas chromatograph with an Inert Extractor source on a 5977B single quadrupole MSD and on a 7010D triple quadrupole GC TQ upgraded with the high efficiency ion source HES 2.0. Columns tested included the J W DB 5Q low bleed phase a DB 5ms UI phase and two standard 5ms type columns. Calibration standards of a representative SVOC mix were prepared at 10 to 1000 pg on column in dichloromethane. A heavy soil extract matrix was used to assess robustness. GC conditions featured a pulsed splitless inlet at 300 C with UI inlet liners a temperature program from 40 C to 340 C and helium carrier gas at constant flow. MS parameters were configured in electron impact ionization with scan mode on the 5977B and dynamic multiple reaction monitoring with concurrent scan acquisition on the 7010D.
The combination of low bleed inert column chemistry and advanced ion source delivers lower limits of detection enhanced quantitation accuracy extended maintenance intervals and simplified method upgrades. This approach supports high throughput laboratories engaged in environmental monitoring industrial QC and regulatory compliance.
Continued innovation in stationary phase design and ion source technology will further lower detection limits and expand capabilities for simultaneous targeted and non targeted workflows. Integration of self tuning maintenance feedback and real time diagnostic features will streamline method development. These advancements will benefit emerging contaminant analysis and complex matrix applications.
The study demonstrates that the Agilent J W DB 5Q GC column paired with the HES 2 0 source on a triple quadrupole instrument markedly improves sensitivity data accuracy and robustness for semivolatile organic compound analysis meeting and exceeding EPA method performance criteria.
Consumables, GC columns, GC/MSD, GC/MS/MS, GC/QQQ, GC/SQ
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Significance of the Topic
The analysis of semivolatile organic compounds (SVOCs) by gas chromatography mass spectrometry is crucial for environmental monitoring and regulatory compliance. Advances in column chemistry directly improve sensitivity data accuracy and method robustness when detecting trace level pollutants in complex matrices.
Objectives and Study Overview
This study evaluates a novel low bleed inert GC column coupled with an advanced ion source on a triple quadrupole GC MS system. It compares performance metrics such as bleed reduction inertness thermal stability sensitivity and chromatographic consistency against conventional columns under high throughput and challenging sample conditions.
Methodology and Instrumentation Used
This evaluation employed an Agilent 8890 gas chromatograph with an Inert Extractor source on a 5977B single quadrupole MSD and on a 7010D triple quadrupole GC TQ upgraded with the high efficiency ion source HES 2.0. Columns tested included the J W DB 5Q low bleed phase a DB 5ms UI phase and two standard 5ms type columns. Calibration standards of a representative SVOC mix were prepared at 10 to 1000 pg on column in dichloromethane. A heavy soil extract matrix was used to assess robustness. GC conditions featured a pulsed splitless inlet at 300 C with UI inlet liners a temperature program from 40 C to 340 C and helium carrier gas at constant flow. MS parameters were configured in electron impact ionization with scan mode on the 5977B and dynamic multiple reaction monitoring with concurrent scan acquisition on the 7010D.
Key Results and Discussion
- Bleed and Baseline Stability
The DB 5Q column exhibited significantly reduced high temperature bleed compared to a conventional 5ms type phase resulting in a stable low baseline that enhances detection of low concentration targets. - Inertness and Peak Symmetry
Problematic phenolic analytes such as dinitrophenols and pentachlorophenol showed improved peak symmetry and five to thirty fold higher signal to noise ratios on the inert DB 5Q phase versus active columns. - Thermal and Matrix Robustness
After two hundred injections of heavy soil extract and repeated thermal cycling retention times and peak shapes remained consistent demonstrating the column’s durability for high throughput environmental analyses. - Acquisition Flexibility
The HES 2.0 ion source enabled simultaneous dMRM and full scan acquisition at ten pg on column capturing both targeted transitions and full spectra for late eluting PAHs with high signal to noise. - Selectivity Matching
The DB 5Q phase showed equivalent selectivity to the existing DB 5ms UI column allowing seamless method transfer without revalidation of retention times.
Benefits and Practical Applications
The combination of low bleed inert column chemistry and advanced ion source delivers lower limits of detection enhanced quantitation accuracy extended maintenance intervals and simplified method upgrades. This approach supports high throughput laboratories engaged in environmental monitoring industrial QC and regulatory compliance.
Future Trends and Opportunities
Continued innovation in stationary phase design and ion source technology will further lower detection limits and expand capabilities for simultaneous targeted and non targeted workflows. Integration of self tuning maintenance feedback and real time diagnostic features will streamline method development. These advancements will benefit emerging contaminant analysis and complex matrix applications.
Conclusion
The study demonstrates that the Agilent J W DB 5Q GC column paired with the HES 2 0 source on a triple quadrupole instrument markedly improves sensitivity data accuracy and robustness for semivolatile organic compound analysis meeting and exceeding EPA method performance criteria.
Reference
- Smith Henry A Comparison of Fritted and Wool Liners for Analysis of Semivolatile Organic Compounds by Gas Chromatography Mass Spectrometry Agilent Technologies application note publication number 5994 2179EN 2020
- How Does Bleed Impact GC MS Data and How Can It Be Controlled Agilent Technologies technical overview publication number 5994 7586EN 2024
- Andrianova A Zhao L Brewing Excellence Quantitating Over 200 Pesticides in Black Tea with Steady Performance and Maximized Uptime by GC MS MS Agilent Technologies application note publication number 5994 7436EN 2024
- Reaser BC Enhanced Longevity and Revolutionized Robustness for the Sensitive Detection of 190 Pesticides over 800 Injections Agilent Technologies application note publication number 5994 7385EN 2024
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Get the Most Out of Your MS
2024|Agilent Technologies|Presentations
Get the Most Out of Your MS Novel Column Chemistry Raises the Bar on Sensitivity and Data Accuracy Vanessa Abercrombie DE-000108 The Evolution of Gas Chromatography 1950 1960 Packed columns 1970 Glass columns 2010 8890 GC HydroInert ion source 5830…
Key words
bleed, bleedbenzo, benzophthalate, phthalateyour, youruber, uber𝑚𝑒𝑡ℎ𝑎𝑛𝑒, 𝑚𝑒𝑡ℎ𝑎𝑛𝑒trimethylphosphate, trimethylphosphatemin, minlog, logghi, ghicolumn, columnintelligence, intelligenceperylene, perylenemaot, maotfluoranthene
GC/MS solutions for environmental and food testing
2024|Agilent Technologies|Guides
GC/MS solutions for environmental and food testing Ensure a Healthier World from the Ground Up Application Compendium Table of Contents Introduction 3 Environmental Analysis of PFAS and Environmental Contaminants in Soil and Oat Plants with GC/Q-TOF 4 Analysis of Phthalate…
Key words
quadratic, quadraticlinear, linearcounts, countspfas, pfashvst, hvstacquisition, acquisitionhydrogen, hydrogensource, sourcemass, massinlet, inlethydroinert, hydroinerttime, timedetected, detectedphthalate, phthalatehelium
A Fast Method for EPA 8270 in MRM Mode Using the 7000 Series Triple Quadrupole GC/MS
2019|Agilent Technologies|Applications
Application Note Environmental A Fast Method for EPA 8270 in MRM Mode Using the 7000 Series Triple Quadrupole GC/MS Authors Melissa Churley, Bruce Quimby, and Anastasia Andrianova Agilent Technologies, Inc. Introduction A fast method for EPA 8270D/E has been developed…
Key words
phthalate, phthalatefluoranthene, fluoranthenedelta, deltabenzo, benzoquadratic, quadraticbenzyl, benzylanthracene, anthracenepyrene, pyreneaverage, averagecalibration, calibrationion, iongain, gaincriteria, criteriabutyl, butyldwell
Staying Ahead in a Rapidly Changing World - Application Compendium
2023|Agilent Technologies|Guides
GC/MS solutions that help analytical labs see more, further, and faster Staying Ahead in a Rapidly Changing World Application Compendium Table of Contents Introduction 3 Analysis of Semivolatile Organic Compounds Using Hydrogen Carrier Gas and HydroInert Ion Source by Gas…
Key words
wide, widefalse, falselinear, linearavg, avgdmrm, dmrmphthalate, phthalatescan, scanacquisition, acquisitiontime, timemin, minbenzo, benzoinlet, inletmethyl, methylcounts, countsfluoranthene