Residual NMP and Solvent Analysis in Lithium-ion Battery Cathode
Applications | 2024 | ShimadzuInstrumentation
Residual organic solvents such as N-Methyl-2-pyrrolidone (NMP) remaining in lithium-ion battery cathode materials can negatively impact electrode performance, safety, and long-term stability. Accurate detection and quantification of these residues are essential for manufacturing quality control and process optimization.
This study introduces a simple headspace GC-FID method for quantifying residual NMP in nickel-cobalt-manganese (NCM) cathodes and a headspace GC-MS approach for qualitative identification of multiple residual solvents. The work also compares solvent levels across different drying conditions to guide process improvements.
A two-step analytical workflow was employed:
Headspace GC-FID combined with compact GC-MS provides a reliable, efficient platform for both quantitative and qualitative analysis of residual solvents in lithium-ion battery cathodes. This methodology supports stringent quality control, process optimization, and ongoing research in battery materials manufacturing.
GC, HeadSpace
IndustriesEnergy & Chemicals
ManufacturerShimadzu
Summary
Importance of the Topic
Residual organic solvents such as N-Methyl-2-pyrrolidone (NMP) remaining in lithium-ion battery cathode materials can negatively impact electrode performance, safety, and long-term stability. Accurate detection and quantification of these residues are essential for manufacturing quality control and process optimization.
Objectives and Study Overview
This study introduces a simple headspace GC-FID method for quantifying residual NMP in nickel-cobalt-manganese (NCM) cathodes and a headspace GC-MS approach for qualitative identification of multiple residual solvents. The work also compares solvent levels across different drying conditions to guide process improvements.
Methodology and Instrumentation
A two-step analytical workflow was employed:
- Quantitative analysis using Brevis GC-2050 coupled with HS-20 NX headspace sampler and flame ionization detection.
- Qualitative profiling of residual solvents using Nexis GC-2030 front end, GCMS-QP2050 mass spectrometer, and HS-20 NX (trap) headspace sampler.
Used Instrumentation
- Brevis GC-2050 with FID detector and SH-I-624Sil MS column (0.32 mm×30 m, d.f. 1.8 µm).
- Nexis GC-2030 with SH-I-624Sil MS column (0.32 mm×60 m, d.f. 1.8 µm).
- HS-20 NX headspace sampler (USTL and trap models).
- GCMS-QP2050 mass spectrometer (scan 35–500 m/z).
Main Results and Discussion
- Calibration linearity for NMP showed R² > 0.9999, with RSDs < 3.2% across three concentration levels.
- Average residual NMP in cathode samples was 2.76 ppm (w/w), as determined from headspace GC-FID measurements.
- GC-MS analysis revealed five major residual peaks; library matching identified likely compounds including N-methylmaleimide, 3-pyrrolin-2-one derivatives, and caprolactam.
- Comparison of five cathodes dried under varying temperature and time conditions demonstrated significant differences in residual solvent levels, highlighting the impact of drying protocols on solvent removal efficiency.
Benefits and Practical Applications
- No sample pretreatment beyond cutting and vial sealing simplifies workflow and reduces potential contamination.
- Compact GC and GC-MS systems minimize laboratory footprint while providing robust quantitative and qualitative data.
- Approach supports rapid quality control screening and process development in battery cathode manufacturing.
Future Trends and Potential Applications
- Integration of automated headspace sampling for high-throughput residual solvent screening.
- Expansion to alternative battery chemistries and solvent systems.
- Coupling with advanced detectors or tandem MS for improved sensitivity and specificity.
- Implementation of real-time monitoring solutions inline with coating and drying equipment.
Conclusion
Headspace GC-FID combined with compact GC-MS provides a reliable, efficient platform for both quantitative and qualitative analysis of residual solvents in lithium-ion battery cathodes. This methodology supports stringent quality control, process optimization, and ongoing research in battery materials manufacturing.
Reference
- Analysis of battery electrolytes and N-methyl-2-pyrrolidone (NMP) via headspace GC-FID – Application News No.05-SCA-180-049-EN
- Analysis of Carbonic Esters and Additives in Lithium Ion Battery Electrolytes – Application News No.01-00708-EN
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Analysis of Residual N-Methyl-2-Pyrrolidone (NMP) in Lithium-Ion Battery Electrodes
2024|Biotage|Applications
Application Note Energy & Chemicals Analysis of Residual N-Methyl-2-Pyrrolidone (NMP) in Lithium-Ion Battery Electrodes An analysis approach using the Agilent 8697 headspace sampler with the Agilent 8860 GC system and flame ionization detector Authors Jie Zhang and Hongtao Shang Agilent…
Key words
nmp, nmpelectrodes, electrodesanode, anoderesponse, responseelectrode, electrodeheadspace, headspacecathode, cathodecathodes, cathodesresidual, residualvial, vialblank, blankanodes, anodesspiked, spikedloop, loopconcentration
Efficient Analysis of Residual Solvents in Pharmaceuticals Using the Compact Model, Brevis GC-2050 (1) —JP18 and USP467, Water-Soluble Samples—
2023|Shimadzu|Applications
Gas Chromatograph HS-20 NX USTL/Brevis™ GC-2050 Application News Efficient Analysis of Residual Solvents in Pharmaceuticals Using the Compact Model, Brevis GC-2050 (1) —JP18 and USP467, Water-Soluble Samples— Yui Higashi, Eisuke Kobayashi User Benefits The slim and compact design of…
Key words
mibk, mibkcpme, cpmevial, vialdegree, degreeprocedure, procedureresidual, residualmin, minnews, newsflowrate, flowratesolvents, solventsseparation, separationpharmaceuticals, pharmaceuticalspolarwax, polarwaxfid, fidmodel
Rapid determination of residual N-methyl-2-pyrrolidone (NMP) in lithium battery electrodes by headspace gas chromatography
2022|Thermo Fisher Scientific|Applications
Application note | 000712 Industrial Rapid determination of residual N-methyl-2-pyrrolidone (NMP) in lithium battery electrodes by headspace gas chromatography Authors Goal Jinshui Che , Feng Bu , Giulia Riccardino , The aim of this application note is to demonstrate the…
Key words
nmp, nmpmin, minbattery, batteryheadspace, headspacelithium, lithiumtracegold, tracegoldvial, vialsample, sampleresidual, residualtime, timepath, pathloop, looprepeatability, repeatabilityelectrode, electrodeiconnect
Efficient Analysis of Residual Solvents in Pharmaceuticals Using the Compact Model, Brevis GC-2050 (2) —JP18 and USP467, Water Insoluble Samples—
2023|Shimadzu|Applications
Gas Chromatograph HS-20 NX USTL/Brevis™ GC-2050 Application News Efficient Analysis of Residual Solvents in Pharmaceuticals Using the Compact Model, Brevis GC-2050 (2) —JP18 and USP467, Water Insoluble Samples— Yui Higashi, Eisuke Kobayashi User Benefits The slim and compact design…
Key words
mibk, mibkcpme, cpmeresidual, residualpharmaceuticals, pharmaceuticalssolvents, solventsustl, ustlflowrate, flowratevial, vialprocedure, proceduredegree, degreemin, minpolarwax, polarwaxfid, fidshort, shortnews