Reliable Analysis of the Light Ends of Petroleum Fractions and Crude Oil Using Capillary Flow Backflush Technology

Presentations | 2010 | Agilent TechnologiesInstrumentation
GC, GCxGC
Industries
Energy & Chemicals
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


The front‐end analysis of light ends in petroleum fractions and crude oil is critical for quality control, process optimization and regulatory compliance in petrochemical and refining industries. Advanced capillary flow backflush technology allows efficient removal of high‐boiling contaminants and late‐eluting matrix components, reducing downtime and protecting high‐resolution analytical columns. This approach enhances throughput and lowers operating costs by eliminating lengthy bake‐out procedures and improving system robustness.

Objectives and Study Overview


This work evaluates a capillary flow backflush (CFT) pre‐column purged union configuration for gas chromatography (GC) analysis of petroleum hydrocarbon fractions. Main goals include:
  • Demonstrate elimination of bake‐out times and carryover.
  • Optimize pre‐column and analytical column flows for concurrent analysis and backflush.
  • Assess retention repeatability and peak shape for C5–C40 mixtures and crude oil samples.
  • Explore compatibility with GC×GC and chemometric classification of crude sources.

Methodology


A deactivated capillary pre‐column (1–2 m retention gap) is installed upstream of a long analytical column (100 m × 0.20–0.25 mm ID, 0.50 µm film). The multimode inlet (MMI) and an auxiliary EPC module control pressure to maintain forward flow in the analytical column while reversing flow in the pre‐column at a programmed backflush time. Inlet temperature is ramped to 425 °C at backflush to purge heavy residues. Both coated (0.10 µm film) and uncoated pre‐columns are evaluated for tuning backflush carbon number cut‐offs.

Instrumentation


  • Agilent GC system with multimode inlet (MMI) and AUX EPC module
  • Purged union (G3186B) configured for capillary flow backflush
  • Flame ionization detector (FID)
  • Pre‐column: 1–2 m deactivated or thin‐film (0.10 µm DB‐1) capillary
  • Analytical column: 100 m × 0.20–0.25 mm × 0.50 µm DB‐PETRO or HP-1

Main Results and Discussion


Backflush after specific carbon numbers (C11–C14) yielded stable baselines and negligible carryover across up to 12 consecutive runs of C5–C40 standards or crude oil samples. Retention time standard deviations were below 0.002 min (<0.3 %), confirming excellent reproducibility. Using coated pre‐columns provided finer control of backflush cut‐off at desired carbon numbers but increased purge duration. Integration with GC×GC using a flow modulator demonstrated robust first‐dimension performance and allowed multivariate analysis to discriminate four crude oil sources via principal component analysis.

Benefits and Practical Applications


  • Substantial reduction of bake‐out and maintenance cycles
  • Enhanced protection and lifetime of high‐resolution analytical columns
  • Improved throughput for routine QC of refinery streams and crude blends
  • Flexible tuning of backflush cut‐off by selecting pre‐column type and length
  • Compatibility with GC×GC for complex mixture profiling and chemometric source identification

Future Trends and Potential Applications


Ongoing developments may include automated dynamic control of backflush thresholds based on real‐time detector feedback, expansion to mass spectrometry detectors for trace heavy‐end analysis, and integration with machine learning for predictive maintenance. Further research can extend the technique to biofuels, environmental samples, and polymer additive profiling where high‐boiling residues pose analytical challenges.

Conclusion


The capillary flow backflush pre‐column configuration offers a simple, high‐performance, in‐oven solution for the reliable analysis of light ends and heavy fractions in petroleum streams. By coupling precise pressure control and programmable temperature ramps, it eliminates column fouling, maximizes uptime and ensures consistent chromatographic performance. Compatibility with GC×GC and chemometric tools further broadens its application for detailed fingerprinting and source attribution.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Prefractionator for Reliable Analysis of the Light Ends of Crude Oil and other Petroleum Fractions
Prefractionator for Reliable Analysis of the Light Ends of Crude Oil and other Petroleum Fractions Application Note Hydrocarbon Processing Author Abstract Roger L Firor A precolumn backflush system based on capillary columns using midpoint pressure Agilent Technologies, Inc. control is…
Key words
backflush, backflushprecolumn, precolumncrude, crudeaux, auxpetroleum, petroleummmi, mmipressure, pressureoil, oilanalytical, analyticalmidpoint, midpointcarbon, carbondesired, desiredtime, timecft, cftcolumn
A New Approach to ASTM D3606 for the Analysis of Benzene and Toluene in Gasoline using Mid-Column Backflush and Hydrogen Carrier with the Agilent 7890B GC
A New Approach to ASTM D3606 for the Analysis of Benzene and Toluene in Gasoline using Mid-Column Backflush and Hydrogen Carrier with the Agilent 7890B GC Application Note Hydrocarbon processing industry Author Abstract Roger L. Firor An Agilent 7890B Series…
Key words
mmi, mmibenzene, benzenetoluene, toluenegasoline, gasolineaviation, aviationbackflush, backflushcft, cftmotor, motorrestrictor, restrictorsplitter, splitteramt, amtethanol, ethanolchannel, channelvol, voltiming
Agilent 7890 Series GC with Prefractionator for Reliable Analysis of Crude Butadiene and Mixed C4 Streams
Agilent 7890 Series GC with Prefractionator for Reliable Analysis of Crude Butadiene and Mixed C4 Streams Application Note Hydrocarbon processing industry Author Abstract Chunxiao Wang A previous application note [1] shows that an Agilent 7890 Series GC configured with Agilent…
Key words
iso, isobutane, butanepentane, pentanepropyne, propynebutylene, butylenepropadiene, propadienebackflushing, backflushingpropane, propanepropylene, propylenehpli, hplialumina, aluminadimer, dimerstdev, stdevcft, cftbackflush
Simple Heart Cutting with Deans Switch and Backflushing with Switch and Backflushing with Capillary Flow Technolgy
Simple Heart Cutting with Deans Switch and Backflushing with Capillary Flow Technolgy Simon Jones Applications pp Engineer g GC or GC/MS Analysis in Complex Matrices • In complex sample matrices, there are often too many overlapping pp g compounds p…
Key words
backflush, backflushrestrictor, restrictormidrofluidic, midrofluidicinnowax, innowaxbenzene, benzenepcm, pcmfid, fidcolumn, columndeans, deansinlet, inletdevices, devicesrestricted, restrictedsolenoid, solenoidpuu, puuethanol
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike