Importance of the topic
The accurate profiling of diesel hydrocarbons by gas chromatography is vital for quality assurance, regulatory compliance and process optimization in the petroleum industry.Objectives and study overview
This application note demonstrates the capabilities of the SH-I-5HT high-temperature capillary column on a Shimadzu GC-FID system to separate diesel-range hydrocarbons from C10 to C40 within a single analytical run.Methodology and instrumentation
Methodology:- Column: SH-I-5HT (30 m × 0.32 mm ID, 0.25 μm film thickness) with 5 m guard column
- Temperature program: 50 °C initial, ramp 10 °C/min to 300 °C, hold 20 min
- Injection: 1 μL cold on-column at 53 °C ramped with same program
- Carrier gas: H₂ at 40 cm/s constant linear velocity
- Detector: FID at 330 °C with N₂ make-up gas (30 mL/min)
Main results and discussion
The method achieved baseline separation of C10 to C40 hydrocarbons with a total analysis time under 40 minutes. The SH-I-5HT column maintained peak shape integrity at high temperatures, delivering consistent retention times and minimal column bleed.Benefits and practical applications
- High thermal stability for extended temperature ranges
- Reliable quantitation of diesel fuel components
- Enhanced throughput for routine QA/QC laboratories
Future trends and possible applications
Advances may include coupling with mass spectrometry for structural identification, ultra-fast GC for reduced cycle times, and automated data processing for real-time monitoring of fuel quality.Conclusion
The SH-I-5HT column on a GC-FID platform offers a robust, high-temperature solution for comprehensive diesel hydrocarbon analysis, balancing speed, resolution and column longevity.Reference
Shimadzu Corporation. ERAS-1000-0471, First Edition: September 2023.Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Analysis of Kerosene
2023|Shimadzu|Applications
ERAS-1000-0470 GC FID SH Series SH-I-5HT Analysis of Kerosene 470 Keywords: petrochemical, high-temperature resistant column 1. C10 2. C40 1 2 Instrument : GC Column Guard column Column temp. Inj. vol. Inj. temp. program Carrier gas Dead time Detector Make-up…
Key words
kerosene, kerosenedead, deadguard, guardgas, gasfid, fidcolumn, columnmake, makeseries, seriescarrier, carrierdetector, detectorrate, rateinstrument, instrumentdata, datatime, timeanalysis
Analysis of Gasoline
2023|Shimadzu|Applications
ERAS-1000-0469 GC FID SH Series SH-I-5HT Analysis of Gasoline 469 Keywords: Petrochemical, High-temperature resistant column 1. C10 2. C40 1 2 Instrument : GC Column Column temp. Event time MS Injector temp. Inj. vol. Carrier gas Dead time Detector Make-up…
Key words
gasoline, gasolinedead, deadevent, eventgas, gasfid, fidmake, maketime, timeseries, seriescarrier, carrierdetector, detectorrate, rateinstrument, instrumentdata, datacolumn, columnanalysis
Analysis of Primary Amines
2023|Shimadzu|Applications
ERAS-1000-0459 GC FID SH Series SH-35 Amine Analysis of Primary Amines 459 Keywords: Underivatized amines, Basic compounds 1. Methylamine 2. Dimethylamine 3. Trimethylamine 4. Ethylamine 5. Isopropylamine 6. tert-Butylamine 7. n-Propylamine 8. Diethylamine 9. sec-Butylamine Instrument Column Column temp. Inj.…
Key words
amines, aminesdead, deadliner, linerfid, fidprimary, primarysplit, splitinlet, inletseries, seriesratio, ratiocarrier, carrierdetector, detectorinstrument, instrumentgas, gascolumn, columntime
Analysis of Styrene Impurities
2023|Shimadzu|Applications
ERAS-1000-0464 GC FID SH Series SH-PolarWax Analysis of Styrene Impurities 464 Keywords: Polar-deactivated column, Chemistry, Styrene 1. Butene 2. 1,3-Butadiene 3. Diethylhydroxoamine 4. Acrylonitrile 5. Vinylcyclohexene 6. Toluene 7. Ethylbenzene 8. m-Xylene 9. Cumene 10. a-Xylene 11. Styrene Instrument :…
Key words
fid, fidpolarwax, polarwaxstyrene, styrenevent, ventimpurities, impuritiessplit, splitseries, seriescarrier, carrierdetector, detectorinstrument, instrumentgas, gascolumn, columnanalysis