Optimized GC/MS/MS Analysis for PAHs in Challenging Matrices
Applications | 2019 | Agilent TechnologiesInstrumentation
Polycyclic aromatic hydrocarbons (PAHs) are toxic, ubiquitous contaminants in food, soil, and the environment. Sensitive, selective analysis of PAHs at trace levels is crucial for environmental monitoring, food safety, and industrial quality control. High organic content and complex matrices challenge conventional GC/MS methods, often leading to spectral interferences, peak tailing, and frequent maintenance.
This study demonstrates an optimized GC/MS/MS workflow using the Agilent 8890 GC coupled to a 7000D triple quadrupole MS. Key goals include achieving robust, linear quantitation of 27 PAHs over a 1–1 000 pg range in challenging matrices, reducing manual maintenance via continuous source cleaning, and simplifying data review through enhanced selectivity.
The combination of JetClean, 9 mm extractor lens, and midcolumn backflush greatly reduces source fouling and column trimming. This optimized GC/MS/MS workflow enhances sensitivity, precision, and throughput for environmental, food, and agricultural laboratories conducting trace PAH analyses.
Adoption of advanced source‐cleaning technologies and automated backflush routines will further improve instrument uptime. Integration with high‐throughput sample preparation, real-time data processing, and AI‐driven anomaly detection promises to extend the approach to broader contaminant classes and regulatory environments.
Optimized GC/MS/MS with continuous hydrogen cleaning, midcolumn backflush, and MRM detection provides a robust, sensitive, and selective platform for PAH analysis in challenging matrices, offering high linearity, stability, and reduced maintenance burden.
GC/MSD, GC/MS/MS, GC/QQQ
IndustriesEnvironmental, Food & Agriculture
ManufacturerAgilent Technologies
Summary
Importance of the Topic
Polycyclic aromatic hydrocarbons (PAHs) are toxic, ubiquitous contaminants in food, soil, and the environment. Sensitive, selective analysis of PAHs at trace levels is crucial for environmental monitoring, food safety, and industrial quality control. High organic content and complex matrices challenge conventional GC/MS methods, often leading to spectral interferences, peak tailing, and frequent maintenance.
Objectives and Study Overview
This study demonstrates an optimized GC/MS/MS workflow using the Agilent 8890 GC coupled to a 7000D triple quadrupole MS. Key goals include achieving robust, linear quantitation of 27 PAHs over a 1–1 000 pg range in challenging matrices, reducing manual maintenance via continuous source cleaning, and simplifying data review through enhanced selectivity.
Methodology and Instrumentation
- GC Configuration: Agilent 8890 with fast oven, pulsed splitless injection (50 psi for 0.7 min), 4 mm straight-bore inlet liner with glass wool, inlet and transfer line at 320 °C.
- MS Setup: Agilent 7000D TQ MS in MRM mode; 9 mm extractor lens; JetClean continuous hydrogen source cleaning (0.33 mL/min); EM gain mode: 10; quadrupoles at 150 °C.
- Backflush Strategy: Midcolumn backflush using a Purged Ultimate Union (PUU) and pneumatic switching device (PSD) to purge high boilers post-run for 1.5 min.
- Columns: Two 15 m DB-EUPAH columns (0.25 mm × 0.25 μm) in series with constant flow (0.96–1.16 mL/min).
- Calibration: Seven levels (1, 2, 10, 20, 100, 200, 1 000 pg/μL) with 500 pg/μL internal standards; R² > 0.999; ISTD area reproducibility ±20 %.
- Robustness Test: Sedge peat extract (high organic soil) spiked with 100 pg PAHs and 500 pg ISTDs; extraction by DCM/acetone; 60 replicate injections.
Main Results and Discussion
- Linearity: All PAHs exhibited R² > 0.999 across the full calibration range, maintained after 60 runs.
- Stability: Over 120 injections of a 100 pg standard, average RSD for calculated concentrations was 1.5 %; ISTD area RSDs were 2.9–5.1 %.
- Matrix Robustness: 60 injections of spiked soil extract yielded average analyte RSD of 4.1 %; most measured concentrations remained within ±20 % of nominal.
- Maintenance Impact: After inlet and column head maintenance, calibration check showed all analytes within ±6 % and R² = 1.000.
- Selectivity: MRM mode eliminated SIM interferences seen in single quadrupole data, reducing manual integration and speeding data review.
Benefits and Practical Applications
The combination of JetClean, 9 mm extractor lens, and midcolumn backflush greatly reduces source fouling and column trimming. This optimized GC/MS/MS workflow enhances sensitivity, precision, and throughput for environmental, food, and agricultural laboratories conducting trace PAH analyses.
Future Trends and Opportunities
Adoption of advanced source‐cleaning technologies and automated backflush routines will further improve instrument uptime. Integration with high‐throughput sample preparation, real-time data processing, and AI‐driven anomaly detection promises to extend the approach to broader contaminant classes and regulatory environments.
Conclusion
Optimized GC/MS/MS with continuous hydrogen cleaning, midcolumn backflush, and MRM detection provides a robust, sensitive, and selective platform for PAH analysis in challenging matrices, offering high linearity, stability, and reduced maintenance burden.
References
- Szelewski M.; Quimby B. D. Optimized PAH Analysis Using the Agilent Self-Cleaning Ion Source; Agilent Technologies Application Note, 2013.
- Anderson K. A.; Szelewski M. J.; Quimby B. D.; Hoffman P. D. Modified Ion Source Triple Quadrupole Mass Spectrometer for Polycyclic Aromatic Hydrocarbons; J. Chromatogr. A, 2015, 1419, 89–99.
- Quimby B. D.; Prest H. F.; Szelewski M. J.; Freed M. K. In-situ Conditioning in Mass Spectrometer Systems; US Patent 8,378,293, 2013.
- Andrianova A. A.; Quimby B. D. Optimized GC/MS Analysis for PAHs in Challenging Matrices; Agilent Technologies Application Note, 2019.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
GC/MS/MS Analysis of PAHs with Hydrogen Carrier Gas
2023|Agilent Technologies|Applications
Application Note Environmental GC/MS/MS Analysis of PAHs with Hydrogen Carrier Gas Using the Agilent HydroInert source in a challenging soil matrix Authors Abstract Samuel P. Haddad, Bruce D. Quimby, and Anastasia A. Andrianova, Agilent Technologies, Inc. The Agilent 8890 GC…
Key words
benzo, benzofluoranthene, fluoranthenepyrene, pyreneperylene, perylenepahs, pahsanthracene, anthracenehydrogen, hydrogeninlet, inlethydroinert, hydroinertsoil, soildibenzothiophene, dibenzothiophenesource, sourceliner, linergas, gasbackflush
Optimized PAH Analysis Using Triple Quadrupole GC/MS with Hydrogen Carrier
2020|Agilent Technologies|Applications
Application Note Optimized PAH Analysis Using Triple Quadrupole GC/MS with Hydrogen Carrier Authors Anastasia A. Andrianova and Bruce D. Quimby Agilent Technologies, Inc. Abstract The Agilent 8890/7000D triple quadrupole GC/MS system with hydrogen carrier gas was used for the analysis…
Key words
hydrogen, hydrogenbenzo, benzocarrier, carrierpah, pahpahs, pahspyrene, pyrenegas, gasfluoranthrene, fluoranthrenefluoranthene, fluoranthenerelative, relativeresponses, responsesinlet, inletcalibration, calibrationenergies, energiesquantifier
Optimized GC/MS Analysis for PAHs in Challenging Matrices
2019|Agilent Technologies|Applications
Application Note Environmental, Food Testing and Agriculture Optimized GC/MS Analysis for PAHs in Challenging Matrices Using the Agilent 5977 Series GC/MSD with JetClean and midcolumn backflush Authors Anastasia A. Andrianova and Bruce D. Quimby Agilent Technologies, Inc. Abstract The Agilent…
Key words
benzo, benzofluoranthene, fluorantheneperylene, perylenepyrene, pyreneghi, ghijetclean, jetcleanpsd, psdacenaphthylene, acenaphthyleneanthracene, anthracenenaphthalene, naphthalenecalculated, calculatedflow, flowsoil, soilsource, sourcebiphenyl
Analysis of PAHs Using Hydrogen Carrier Gas and the Hydrogen-Optimized Source with GC/MS and GC/MS/MS in Challenging Soil Matrix
2023|Agilent Technologies|Posters
Poster Reprint ASMS 2023 Poster number TP 234 Analysis of PAHs Using Hydrogen Carrier Gas and the Hydrogen-Optimized Source with GC/MS and GC/MS/MS in Challenging Soil Matrix Samuel Haddad, Bruce Quimby, Anastasia Andrianova, Erinn Oneill, and Eric Fausett Agilent Technologies,…
Key words
benzo, benzoinlet, inletpahs, pahsfluoranthene, fluorantheneanthracene, anthracenepyrene, pyrenebackflushing, backflushinghydrogen, hydrogenconnection, connectionpuu, puudibenzothiophene, dibenzothiophenemode, modesoil, soilpsd, psdexcellent