GCC: ASTM D3606 Capillary Options: Products and tips for getting started

Posters | 2022 | RestekInstrumentation
GC columns, Consumables
Industries
Environmental, Energy & Chemicals
Manufacturer
Restek

Summary

Significance of the topic


Spark ignition fuels require precise quantification of aromatic hydrocarbons such as benzene and toluene for quality control and regulatory compliance. The ASTM D3606 capillary option enhances separation efficiency through a dual-column gas chromatography setup with backflush capability, offering rapid and reliable analysis of fuel samples.

Objectives and Study Overview

  • Assess the performance of Restek Rxi-1ms and Stabilwax capillary columns in series with backflush for the separation of benzene, toluene, and other volatile components.
  • Demonstrate method setup, optimization strategies, and troubleshooting tips for effective implementation of ASTM D3606 Option A.
  • Compare analyte separation in a certified D3606 standard and real gasoline samples to validate method applicability.

Used Instrumentation

  • Gas chromatograph equipped with dual flame ionization detectors (FID A and FID B).
  • Restek Rxi-1ms capillary column (30 m × 0.25 mm × 0.5 µm; cat# 13338).
  • Restek Stabilwax capillary column (60 m × 0.32 mm × 1 µm; cat# 10657).
  • 0.52 m × 0.1 µm fused-silica restrictor for column connection and backflush management.
  • Auxiliary electronic pressure controller (EPC) for backflush switching.
  • Restek Electronic Leak Detector (cat# 28500) and ProFLOW 6000 electronic flowmeter (cat# 22656).

Methodology and Instrumentation

  • Columns were arranged in series: nonpolar Rxi-1ms first, followed by polar Stabilwax, connected via a restrictor to control backflush timing.
  • Carrier gas pressures: inlet pressure programmed from 35 psi down to 5 psi at backflush initiation (~4.36 min).
  • Column flows: ~2.0 mL/min through Rxi-1ms and ~2.6 mL/min through Stabilwax, adjusted slightly during backflush.
  • Backflush procedure: After toluene elution from the first column, the auxiliary EPC redirects the carrier gas to purge heavy components from Rxi-1ms while maintaining elution through Stabilwax.
  • Detector placement: FID B monitors effluent from the first column; FID A captures separated analytes from the second column.

Main Results and Discussion

  • Resolved Separation: Benzene and toluene peaks were baseline-separated in both the standard and gasoline matrix, with additional volatile organics eluting in defined windows.
  • Reduced Analysis Time: Backflushing of the first column eliminated late-eluting high-boiling compounds, keeping total run time under 20 minutes.
  • Troubleshooting Insights: Regular leak checks, flow verification during backflush using a flowmeter, and inter-sample blanks are critical to ensure complete toluene removal and prevent analyte carryover.

Benefits and Practical Applications

  • Dual-phase separation enhances selectivity and peak resolution for aromatic and polar volatile compounds.
  • Backflush strategy significantly increases sample throughput by reducing post-analysis cleanup time.
  • Compliance: Supports accurate determination of benzene and toluene levels in spark ignition fuels, meeting regulatory requirements.
  • Guidance: Provides laboratories with detailed setup recommendations to minimize method development time.

Future Trends and Opportunities

  • Full Method Validation: Extend performance evaluation across various fuel grades and broader analyte lists to establish method robustness.
  • Automation Enhancements: Integrate EPC calculators and software-driven flow controls to simplify method setup and reduce manual adjustments.
  • Column Innovations: Explore shorter or novel stationary phases to further decrease analysis time without sacrificing resolution.
  • Application Expansion: Adapt the dual-column backflush approach for biofuel blends and alternative fuel matrices.

Conclusion


The dual-column backflush configuration described in ASTM D3606 Option A, utilizing Restek Rxi-1ms and Stabilwax capillary columns, achieves high-resolution separation of benzene and toluene in under 20 minutes. Practical tips for leak control, flow monitoring, and backflush management support reliable method adoption in analytical and quality control laboratories.

Reference


Erica Pack, Jan Pijpelink, Barry Burger, Chris English. ASTM D3606 Capillary Options: Products and tips for getting started. Restek, USA.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Simple Heart Cutting with Deans Switch and Backflushing with Switch and Backflushing with Capillary Flow Technolgy
Simple Heart Cutting with Deans Switch and Backflushing with Capillary Flow Technolgy Simon Jones Applications pp Engineer g GC or GC/MS Analysis in Complex Matrices • In complex sample matrices, there are often too many overlapping pp g compounds p…
Key words
backflush, backflushrestrictor, restrictormidrofluidic, midrofluidicinnowax, innowaxbenzene, benzenepcm, pcmfid, fidcolumn, columndeans, deansinlet, inletdevices, devicesrestricted, restrictedsolenoid, solenoidpuu, puuethanol
A New Approach to ASTM D3606 for the Analysis of Benzene and Toluene in Gasoline using Mid-Column Backflush and Hydrogen Carrier with the Agilent 7890B GC
A New Approach to ASTM D3606 for the Analysis of Benzene and Toluene in Gasoline using Mid-Column Backflush and Hydrogen Carrier with the Agilent 7890B GC Application Note Hydrocarbon processing industry Author Abstract Roger L. Firor An Agilent 7890B Series…
Key words
mmi, mmibenzene, benzenetoluene, toluenegasoline, gasolineaviation, aviationbackflush, backflushcft, cftmotor, motorrestrictor, restrictorsplitter, splitteramt, amtethanol, ethanolchannel, channelvol, voltiming
Determination of Benzene and Toluene in Gasoline by ASTM D3606 on an Agilent 8890 GC with Capillary Columns
Application Note Energy and Chemicals Determination of Benzene and Toluene in Gasoline by ASTM D3606 on an Agilent 8890 GC with Capillary Columns Author Ian Eisele Agilent Technologies, Inc. Abstract In accordance with ASTM D3606, the concentration of benzene and…
Key words
benzene, benzenetoluene, toluenebackflush, backflushgasoline, gasolinemethod, methodhelium, heliumprecolumn, precolumnhydrogen, hydrogentranslation, translationcolumns, columnscarrier, carriermonitor, monitordetermination, determinationtiming, timingpass
Analysis of Benzene and Toluene by Brevis GC-2050 Using ASTM D3606
Brevis GC-2050 Gas Chromatograph Application News Analysis of Benzene and Toluene by Brevis GC-2050 Using ASTM D3606 Misato Ishimoto, Toyohito Wada, Kei Usui User Benefits  Quantitation of benzene and toluene in fuels based on ASTM D3606 is possible. …
Key words
benzene, benzenetoluene, toluenebackflush, backflushbutanol, butanolmibk, mibkapc, apcvalue, valueinquiry, inquiryaux, auxpermissible, permissiblequantitation, quantitationratio, ratiodifference, differencegas, gasspl
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike