Enable Hydrogen Carrier Gas Selections without Compromising GC/MS Performance
Guides | 2023 | Agilent TechnologiesInstrumentation
Ultralow-level monitoring of volatile organic compounds (VOCs) in drinking water is critical for public health and regulatory compliance. Helium shortages and rising costs have driven the search for alternative GC carrier gases. Hydrogen offers cost savings and faster separations, but its reactivity can cause in-source chemical transformations, spectral anomalies, and peak tailing. The Agilent HydroInert source was developed to enable routine use of hydrogen carrier gas in GC/MSD without compromising spectral fidelity or sensitivity.
This work evaluates a static headspace GC/MSD method using hydrogen carrier gas and the Agilent HydroInert extractor source for the quantitative analysis of 80 priority VOCs in drinking water. The method was optimized for chromatographic speed (<7 min run time), mass spectrum integrity, and quantitative performance over a calibration range of 0.05–25 µg/L.
Static headspace GC/MSD using hydrogen carrier gas and the Agilent HydroInert source provides a reliable, high-throughput alternative to helium for VOC analysis in drinking water. The method delivers fast separations, excellent spectral fidelity, and robust quantitation at regulatory levels, with reduced maintenance demands.
GCxGC, GC/MSD, GC/MS/MS, HeadSpace, Thermal desorption, Sample Preparation, GC/SQ, GC/QQQ, Consumables
IndustriesEnvironmental, Food & Agriculture, Energy & Chemicals
ManufacturerAgilent Technologies, Markes
Summary
Význam tématu
Ultralow-level monitoring of volatile organic compounds (VOCs) in drinking water is critical for public health and regulatory compliance. Helium shortages and rising costs have driven the search for alternative GC carrier gases. Hydrogen offers cost savings and faster separations, but its reactivity can cause in-source chemical transformations, spectral anomalies, and peak tailing. The Agilent HydroInert source was developed to enable routine use of hydrogen carrier gas in GC/MSD without compromising spectral fidelity or sensitivity.
Cíle a přehled studie
This work evaluates a static headspace GC/MSD method using hydrogen carrier gas and the Agilent HydroInert extractor source for the quantitative analysis of 80 priority VOCs in drinking water. The method was optimized for chromatographic speed (<7 min run time), mass spectrum integrity, and quantitative performance over a calibration range of 0.05–25 µg/L.
Použitá metodika a instrumentace
- GC–MS System: Agilent 8890 GC with multimode (MMI) inlet and 8697 headspace sampler; Agilent 5977C Inert Plus MSD equipped with HydroInert source and 9 mm extractor lens.
- GC Column: Agilent J&W DB-624 Ultra Inert, 20 m × 0.18 mm id, 1 µm film.
- Carrier Gas: Hydrogen (99.9999% purity) in constant flow mode (0.95 mL/min).
- Injection: Static headspace, 20 mL vials with 5 g Na₂SO₄; vial equilibration 75 °C for 12 min; pulsed split injection (21:1 split ratio, 26 psi pulse).
- MS Conditions: HydroInert source (325 °C), transfer line 250 °C, quadrupole 200 °C, scan range 35–260 m/z or SIM. Unknowns Analysis deconvolution with NIST 20 library.
- Calibration: Eight levels (0.05, 0.1, 0.25, 0.5, 1, 2.5, 10, 25 µg/L) spiked with internal standards.
Hlavní výsledky a diskuse
- Chromatography: Baseline or near-baseline resolution of 80 VOCs in <7 min, including challenging isomers such as cis-/trans-dichloroethylenes.
- Spectral Fidelity: Deconvoluted scan spectra yielded NIST match scores >90 for >90% of analytes. HydroInert source prevented hydrogenation and dechlorination artifacts.
- Quantitative Performance: Linear calibration (R² ≥0.995) over 0.05–25 µg/L for 78 of 80 compounds. Limits of quantitation (LOQs) of 0.05 µg/L achieved for most targets. Relative standard deviations <12% across the range.
- SIM Mode: Selected-ion monitoring further improved signal-to-noise by factors of 5–20, enabling reliable quantification at regulatory levels.
- Robustness: Hydrogen carrier gas with HydroInert source required no additional source cleaning over hundreds of injections, matching helium-based maintenance intervals.
Přínosy a praktické využití metody
- Helium Replacement: Ensures uninterrupted laboratory operations with lower carrier gas costs.
- Faster Analysis: Hydrogen provides higher linear velocity, reducing run times and boosting sample throughput.
- Spectral Integrity: HydroInert source retains library match confidence, allowing continued use of existing helium-based libraries.
- Maintenance Savings: Reduced in-source reactions minimize downtime and cleaning frequency.
Budoucí trendy a možnosti využití
- Extension to GC/MS/MS quantitation for even lower detection limits and greater matrix selectivity.
- Adaptation to other environmental and food safety analytes where helium alternatives are desired.
- Integration with automated sample preparation and advanced deconvolution algorithms for non-target screening.
Závěr
Static headspace GC/MSD using hydrogen carrier gas and the Agilent HydroInert source provides a reliable, high-throughput alternative to helium for VOC analysis in drinking water. The method delivers fast separations, excellent spectral fidelity, and robust quantitation at regulatory levels, with reduced maintenance demands.
Reference
- U.S. EPA, Method 524.2: Purge-and-Trap Capillary GC/MS for VOCs in Water.
- Agilent Technologies, HydroInert Source Application Note, publication 5994-4889EN.
- Agilent MassHunter Unknowns Analysis Software User Guide.
- Quimby B. et al., In-Situ Conditioning in Mass Spectrometer Systems, US Patent 8,378,293, 2013.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Analysis of Semivolatile Organic Compounds Using Hydrogen Carrier Gas and the Agilent HydroInert Source by Gas Chromatography/Mass Spectrometry
2022|Agilent Technologies|Applications
Application Note Environmental Analysis of Semivolatile Organic Compounds Using Hydrogen Carrier Gas and the Agilent HydroInert Source by Gas Chromatography/Mass Spectrometry Author Angela Smith Henry, PhD Agilent Technologies, Inc. Abstract Gas chromatography/mass spectrometry (GC/MS) is integral to the analysis of…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluoranthenepass, passnitrobenzene, nitrobenzeneanthracene, anthracenearamite, aramitelinear, linearmethanesulfonate, methanesulfonatediallate, diallatepyrene, pyreneazobenzene, azobenzenebenzyl, benzylnitroso, nitrosohydroinert
Agilent Inert Plus GC/MS System with HydroInert Source
2022|Agilent Technologies|Technical notes
Technical Overview Agilent Inert Plus GC/MS System with HydroInert Source Applying H2 carrier gas to real-world GC/MS analyses Introduction With increased price and pressure on the helium (He) market, laboratories are looking for a more sustainable alternative to helium and…
Key words
hydroinert, hydroinertextractor, extractorsource, sourcecompounds, compoundspass, passnitrobenzene, nitrobenzenemass, masslens, lenscarrier, carrierlms, lmsgas, gasdifferentiating, differentiatingrepeller, repellerbenzo, benzohydrogen
Intelligent Agilent GC/MS/MS (HydroInert Application Compendium)
2023|Agilent Technologies|Guides
Intelligent Agilent GC/MS/MS Conquering Analytical Challenges Application Compendium Table of Contents Introduction 3 Application Notes Pesticides Analysis Hydrogen Carrier Gas for Analyzing Pesticides in Pigmented Foods with GC/MS/MS 4 A Fast and Robust GC/MS/MS Analysis of 203 Pesticides in…
Key words
quadratic, quadraticavg, avglinear, linearhydrogen, hydrogendmrm, dmrmhes, hesbenzo, benzoscan, scanhydroinert, hydroinertinlet, inletacquisition, acquisitionsource, sourcetime, timefluoranthene, fluoranthenematrix
Analysis of Semivolatile Organic Compounds with Hydrogen Carrier Gas and HydroInert Source by Gas Chromatography/Triple Quadrupole Mass Spectrometry (GC/MS/MS)
2022|Agilent Technologies|Applications
Application Note Environmental Analysis of Semivolatile Organic Compounds with Hydrogen Carrier Gas and HydroInert Source by Gas Chromatography/Triple Quadrupole Mass Spectrometry (GC/MS/MS) Author Abstract Angela Smith Henry, PhD Agilent Technologies, Inc. Gas chromatography/mass spectrometry (GC/MS) is integral to the analysis…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluoranthenequadratic, quadraticpass, passaramite, aramitemethanesulfonate, methanesulfonatenitrosodi, nitrosodianthracene, anthracenepyrene, pyrenebenzyl, benzylbutyl, butylparathion, parathionnitroso, nitrosohydroinert