Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ 644-2013
Applications | 2017 | Agilent TechnologiesInstrumentation
VOC monitoring in ambient air is critical because these compounds contribute to photochemical smog, particulate pollution and pose direct health risks. Reliable measurement of trace level VOCs supports air quality management and regulatory compliance, particularly under Chinese EPA Method HJ 644-2013.
This study evaluates an automated cryogen free thermal desorption GC MS system for analysis of VOCs trapped on sorbent tubes. The aim is to demonstrate compliance with HJ 644 2013, focusing on system linearity, sensitivity, repeatability and the innovative feature of quantitative split flow re collection for method validation.
Ambient air was sampled at flow rate 50 mL min for 20 minutes (1 L) using sorbent tubes and ACTI VOC sampling pump. A 62 component gas standard including the 34 target VOCs was used for calibration. The thermal desorption protocol involved tube desorption at 220 C for 4 minutes, focusing trap held at 25 C then ramped to 220 C at 40 C s, followed by GC separation on an Agilent J W DB 624 column (20 m × 0.18 mm, 1.0 um) with helium carrier at 1.2 mL min. Oven program started at 35 C for 3 minutes then ramped to 190 C at 15 C min. Mass spectrometry was performed in full scan mode m z 35-300.
The system showed excellent chromatographic performance with sharp peaks for all 34 target compounds including polar analytes. Linearity was confirmed over 0.5 to 9 ppbv with R2 values above 0.99. Repeatability for seven replicates at 0.5 ppbv yielded RSDs below 8% and LODs below 0.15 ppbv. Quantitative re collection of split flows was demonstrated over a 60 to 1 split ratio and for repeated analyses of benzene and toluene, matching theoretical decay. Real ambient air samples from office, laboratory and semirural sites revealed elevated VOC levels in the office environment.
The combined sorbent tube TD GC MS approach offers high sensitivity, automation and minimal interference for trace VOC analysis. The automated split flow re collection feature enables on demand repeat analysis and simplifies method development, reducing sample reruns and ensuring data reliability. This workflow is suited for environmental monitoring, industrial hygiene and regulatory testing.
Advances may include integration of reactive sorbents for improved capture of highly volatile or reactive species, coupling with tandem mass spectrometry for enhanced selectivity, and deployment in field portable TD GC MS platforms. Data analytics and machine learning could further optimize method development and air quality interpretation.
The automated cryogen free TD GC MS system meets Chinese EPA Method HJ 644 2013 requirements for ambient VOC monitoring. It delivers robust sensitivity, linearity, repeatability and a unique automated split flow re collection capability, supporting efficient method validation and reliable environmental analysis.
GC/MSD, Thermal desorption
IndustriesEnvironmental
ManufacturerMarkes
Summary
Importance of the Topic
VOC monitoring in ambient air is critical because these compounds contribute to photochemical smog, particulate pollution and pose direct health risks. Reliable measurement of trace level VOCs supports air quality management and regulatory compliance, particularly under Chinese EPA Method HJ 644-2013.
Objectives and Study Overview
This study evaluates an automated cryogen free thermal desorption GC MS system for analysis of VOCs trapped on sorbent tubes. The aim is to demonstrate compliance with HJ 644 2013, focusing on system linearity, sensitivity, repeatability and the innovative feature of quantitative split flow re collection for method validation.
Methodology
Ambient air was sampled at flow rate 50 mL min for 20 minutes (1 L) using sorbent tubes and ACTI VOC sampling pump. A 62 component gas standard including the 34 target VOCs was used for calibration. The thermal desorption protocol involved tube desorption at 220 C for 4 minutes, focusing trap held at 25 C then ramped to 220 C at 40 C s, followed by GC separation on an Agilent J W DB 624 column (20 m × 0.18 mm, 1.0 um) with helium carrier at 1.2 mL min. Oven program started at 35 C for 3 minutes then ramped to 190 C at 15 C min. Mass spectrometry was performed in full scan mode m z 35-300.
Instrumentation
- TD100 xr thermal desorption system (Markes International)
- ACTI VOC sampling pump (Markes International)
- Calibration Solution Loading Rig (Markes International)
- DB 624 GC column (Agilent Technologies)
Main Results and Discussion
The system showed excellent chromatographic performance with sharp peaks for all 34 target compounds including polar analytes. Linearity was confirmed over 0.5 to 9 ppbv with R2 values above 0.99. Repeatability for seven replicates at 0.5 ppbv yielded RSDs below 8% and LODs below 0.15 ppbv. Quantitative re collection of split flows was demonstrated over a 60 to 1 split ratio and for repeated analyses of benzene and toluene, matching theoretical decay. Real ambient air samples from office, laboratory and semirural sites revealed elevated VOC levels in the office environment.
Benefits and Practical Applications
The combined sorbent tube TD GC MS approach offers high sensitivity, automation and minimal interference for trace VOC analysis. The automated split flow re collection feature enables on demand repeat analysis and simplifies method development, reducing sample reruns and ensuring data reliability. This workflow is suited for environmental monitoring, industrial hygiene and regulatory testing.
Future Trends and Opportunities
Advances may include integration of reactive sorbents for improved capture of highly volatile or reactive species, coupling with tandem mass spectrometry for enhanced selectivity, and deployment in field portable TD GC MS platforms. Data analytics and machine learning could further optimize method development and air quality interpretation.
Conclusion
The automated cryogen free TD GC MS system meets Chinese EPA Method HJ 644 2013 requirements for ambient VOC monitoring. It delivers robust sensitivity, linearity, repeatability and a unique automated split flow re collection capability, supporting efficient method validation and reliable environmental analysis.
References
- Chinese Ministry of Environmental Protection HJ 644 2013 Ambient air Determination of volatile organic compounds Sorbent adsorption and thermal desorption GC MS method
- Markes International Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD GC MS
- Agilent Technologies DB 624 GC Column Datasheet
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Monitoring VOCs in Stationary Source Emissions Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ 734-2014
2017|Markes|Applications
Application Note 117 Monitoring VOCs in Stationary Source Emissions Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ 734-2014 Application Note Abstract This application note demonstrates the excellent performance offered by Markes International’s automated cryogen-free…
Key words
markes, markessorbent, sorbentdesorption, desorptionsplit, splitanisole, anisolexylene, xylenethermal, thermalabundance, abundanceinternational, internationaltube, tubeethylbenzene, ethylbenzenestyrene, styrenecollection, collectiontrifluoromethyl, trifluoromethylbenzene
Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15
2017|Markes|Applications
Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15 Application Note Environmental, Canister, TO-15, Air Monitoring, Air Toxics Abstract This application note describes the GC/MS analysis of humidified canister air toxics samples at various…
Key words
kori, koricanister, canisterabundance, abundanceair, airrrf, rrfcia, ciaretention, retentioncanisters, canisterssampling, samplingmin, minxylene, xylenevinyl, vinylnaphthalene, naphthaleneppbv, ppbvwater
Markes TD100-xr
|Markes|Brochures and specifications
TD100-xr Enabling confident, high‑throughput, automated thermal desorption analysis TD100-xr ™ Introducing the TD100-xr automated thermal desorber for GC and GC–MS – an unrivalled platform for the analysis of trace-level volatile and semi-volatile organic compounds (VOCs and SVOCs) in air and…
Key words
markes, markestube, tubesplit, splitvoc, voccollection, collectiontubes, tubescaps, capsextended, extendedcircled, circledthermal, thermaltrap, trapdifflok, diffloksample, sampledesorption, desorptionair
Markes UNITY-xr
|Markes|Brochures and specifications
UNITY-xr The most versatile and powerful thermal desorption unit UNITY-xr ™ Introducing the UNITY-xr thermal desorber for GC and GC–MS – the most versatile instrument for the analysis of trace‑level volatile and semi-volatile organic compounds (VOCs and SVOCs) in air…
Key words
unity, unitycanister, canisterdesorption, desorptiontube, tubevoc, voccollection, collectionmarkes, markescia, ciasplit, splitthermal, thermaleic, eicair, airtubes, tubesautomated, automatedoutlet