Analytical evaluation of the Agilent IDP-10 Dry Scroll Vacuum Pump and Oil Roughing Pump on the Agilent GC Intuvo 9000 coupled to 7010B Triple Quad Mass Spectrometry.
Applications | 2018 | Agilent TechnologiesInstrumentation
The reliable detection of pesticide residues in food matrices is critical for regulatory compliance and consumer safety ensuring trace level monitoring in complex samples such as fruit extract
This study evaluates the Agilent IDP-10 Dry Scroll Vacuum Pump performance against a conventional oil roughing pump within an Agilent GC Intuvo 9000 system coupled to a 7010B Triple Quadrupole MS focusing on multi residue pesticide analysis in lemon matrix
Ten gram homogenized lemon sample extracted with acetonitrile and NaOH followed by QuEChERS cleanup then analyzed by pulsed splitless injection on GC MS MS with tuning performed for each pump configuration Calibration was conducted at eight concentration levels from 0.2 to 50 ug per liter in matrix standards
Vacuum pressure and tune parameters were essentially equivalent for both pumps within recommended limits Calibration curves showed r squared values above 0.99 across all analytes IDP 10 pump delivered improved repeatability with lower RSD values and enhanced signal to noise ratios for the majority of compounds while peak shape remained consistent between configurations
The oil free IDP 10 pump offers reduced maintenance and lower acoustic noise environmental advantages along with reliable vacuum performance yielding better precision and sensitivity ideal for routine QA QC and high throughput pesticide surveillance
Integration of compact scroll pumps into automated workflows Internet of Things enabled monitoring predictive maintenance advances in pump materials may further enhance vacuum stability and broaden application to diverse sample types requiring oil free operation
The Agilent IDP 10 Dry Scroll Pump matches conventional roughing pump in vacuum stability and tuning performance while providing superior analytical precision and signal quality making it a robust choice for multi residue pesticide analysis
GC/MSD, GC/MS/MS, GC/QQQ
IndustriesFood & Agriculture
ManufacturerAgilent Technologies
Summary
Significance of the topic
The reliable detection of pesticide residues in food matrices is critical for regulatory compliance and consumer safety ensuring trace level monitoring in complex samples such as fruit extract
Objectives and overview
This study evaluates the Agilent IDP-10 Dry Scroll Vacuum Pump performance against a conventional oil roughing pump within an Agilent GC Intuvo 9000 system coupled to a 7010B Triple Quadrupole MS focusing on multi residue pesticide analysis in lemon matrix
Used instrumentation
- Agilent Intuvo 9000 Gas Chromatograph
- Agilent 7010B Triple Quadrupole Mass Spectrometer
- Agilent IDP-10 Dry Scroll Vacuum Pump
- Conventional oil roughing pump
- Ceramic homogenizer and QuEChERS dispersive SPE kits
Methodology
Ten gram homogenized lemon sample extracted with acetonitrile and NaOH followed by QuEChERS cleanup then analyzed by pulsed splitless injection on GC MS MS with tuning performed for each pump configuration Calibration was conducted at eight concentration levels from 0.2 to 50 ug per liter in matrix standards
Main results and discussion
Vacuum pressure and tune parameters were essentially equivalent for both pumps within recommended limits Calibration curves showed r squared values above 0.99 across all analytes IDP 10 pump delivered improved repeatability with lower RSD values and enhanced signal to noise ratios for the majority of compounds while peak shape remained consistent between configurations
Benefits and practical applications
The oil free IDP 10 pump offers reduced maintenance and lower acoustic noise environmental advantages along with reliable vacuum performance yielding better precision and sensitivity ideal for routine QA QC and high throughput pesticide surveillance
Future trends and opportunities
Integration of compact scroll pumps into automated workflows Internet of Things enabled monitoring predictive maintenance advances in pump materials may further enhance vacuum stability and broaden application to diverse sample types requiring oil free operation
Conclusion
The Agilent IDP 10 Dry Scroll Pump matches conventional roughing pump in vacuum stability and tuning performance while providing superior analytical precision and signal quality making it a robust choice for multi residue pesticide analysis
Reference
- Agilent Technologies Application Note 5991 9201EN printed March 2018
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Pesticide Residue Analysis with Automated EDGE Extraction System and Multi-Layer GC/MS Sandwich Injection
2021|Agilent Technologies|Posters
Poster Reprint ASMS 2021 Poster number MP076 Pesticide Residue Analysis with Automated EDGE Extraction System and Multi-Layer GC/MS Sandwich Injection Anastasia Andrianova,1 Jessica Westland,1 Bruce Quimby,1 Alicia Stell,2 Benedict Liu,2 Candice Cashman2 Agilent Technologies, Inc. Wilmington DE 2 CEM Corporation,…
Key words
edge, edgequechers, quechersmethyl, methylendosulfan, endosulfanextraction, extractionsandwich, sandwichbhc, bhcethyl, ethylendrin, endrinbromfenvinfos, bromfenvinfosmatched, matchedbromophos, bromophosmatrix, matrixistd, istdpirimiphos
ANALYTICAL EVALUATION OF THE AGILENT IDP-15 OIL-FREE SCROLL PUMP AND EDWARDS ROTARY PUMP ON THE AGILENT 7000 GC/MS-MS
2016|Agilent Technologies|Technical notes
ENVIRONMENTAL ANALYSIS ANALYTICAL EVALUATION OF THE AGILENT IDP-15 OIL-FREE SCROLL PUMP AND EDWARDS ROTARY PUMP ON THE AGILENT 7000 GC/MS-MS Solutions for Your Analytical Business Markets and Applications Programs Authors ABSTRACT Jennifer P. Pascali & Michele Pozzebon dtoLABS, authorized Agilent…
Key words
mcpa, mcpachlorothalonil, chlorothalonildeltamethrin, deltamethrintebuconazole, tebuconazolecompound, compoundmarkets, marketsrsd, rsdpozzebon, pozzebonprograms, programsbusiness, businessmichele, micheleviolent, violentquieter, quieteredwards, edwardsabsolute
Increasing productivity in pesticide residues analysis by GC-MS/MS with on-line automated micro solid phase extraction (μSPE) clean-up
2023|Thermo Fisher Scientific|Applications
Technical note | 002569 Food safety Increasing productivity in pesticide residues analysis by GC-MS/MS with on-line automated micro solid phase extraction (µSPE) clean-up Authors Goal Lukasz Rajski1, Moira Zanaboni2, The aims of this study were to evaluate the overall performance,…
Key words
methyl, methylethyl, ethylbromophos, bromophosrsd, rsdrecovery, recoveryµspe, µspebromfenvinphos, bromfenvinphosendrin, endrinendosulfan, endosulfannonachlor, nonachloronion, onionclean, cleanpirimiphos, pirimiphosmass, massazinphos
Analysis of Multiclass Multiresidue Pesticides in Milk Using Agilent Captiva EMR—Lipid with LC/MS/MS and GC/MS/MS
2020|Agilent Technologies|Applications
Application Note Food Testing and Agriculture Analysis of Multiclass Multiresidue Pesticides in Milk Using Agilent Captiva EMR—Lipid with LC/MS/MS and GC/MS/MS Authors Xia Yang, Youjuan Zhang, and Zhiming Zhang Agilent Technologies, Inc. Abstract This application note describes the analysis of…
Key words
sulfone, sulfonemean, meanfipronil, fipronilemr, emrrsd, rsdrecovery, recoverylipid, lipidcleanup, cleanupcaptiva, captivaphorate, phoratehch, hchfenamiphos, fenamiphosmethyl, methylterbufos, terbufossetting