Introducing the EDXIR-Holder: Sample Holder/Stocker for Contaminant Measurement

Technical notes | 2017 | ShimadzuInstrumentation
FTIR Spectroscopy, X-ray
Industries
Manufacturer
Shimadzu

Summary

Significance of the Topic


Contaminant analysis in production lines and laboratory environments requires rapid and nondestructive identification of both elemental and molecular residues. Combining energy dispersive X-ray spectrometry with Fourier transform infrared spectroscopy enables a comprehensive assessment of metal inorganic polymeric and organic substances with minimal sample preparation and damage.

Objectives and Overview


This article presents the design and application of the EDXIR-Holder a compact openable case that streamlines sample handling for contaminant measurement. The holder allows samples to be adhered to an acrylic adhesive film for analysis and then closed in the same chamber for elemental analysis without additional manipulation.

Methodology


The EDXIR-Holder comprises a 35 μm thick acrylic adhesive film bonded to a 5 μm polypropylene layer transparent to fluorescent X-rays. For FTIR analysis the case is opened and the ATR prism is pressed directly onto the adhered sample. For EDX analysis the case is closed and positioned with the polypropylene film facing the X-ray source. The recommended sample size ranges from 1.5 mm to 5.0 mm with thicknesses of 0.5 mm to 1.5 mm for EDX and 0.1 mm to 4.0 mm for FTIR and weights under 1 g. Care must be taken with small or protruding samples to avoid film puncture and spectral overlap from the adhesive layer. In such cases difference spectra between the sample and adhesive layer can be used.

Used Instrumentation


  • EDX System: EDX-8000 with Rh X-ray tube 15 kV for light elements auto 50 kV for heavy elements vacuum atmosphere 1 mm measurement diameter up to 60 second integration time
  • FTIR System: IRAffinity-1S with MIRacle10 diamond ATR prism 4 cm-1 resolution 40 scans accumulation Happ-Genzel apodization DLATGS detector

Main Results and Discussion


An unknown contaminant from a production line was analyzed using the EDXIR-Holder. FTIR ATR spectra indicated a mixture of polyvinyl chloride phthalate ester and calcium carbonate. EDX quantification confirmed chlorine at 25.14% calcium 11.43% and trace lead 0.60% along with titanium chromium strontium and copper. Lead detection highlighted the advantage of combined techniques since it was not observed by FTIR alone. A separate test with a small polyethylene film fragment first yielded an incorrect match to ethylene acrylic acid due to the adhesive layer. Using the difference spectrum approach correctly identified polyethylene demonstrating the importance of proper sample coverage or spectral subtraction for minute samples.

Benefits and Practical Applications


  • Minimized sample handling reduces risk of loss and contamination
  • Rapid switch between FTIR and EDX without transfer improves laboratory efficiency
  • Nondestructive analysis preserves sample integrity for further testing or archiving
  • Compact design allows sample labeling and storage within the holder
  • Integration with EDXIR-Analysis software supports streamlined data processing

Future Trends and Potential Applications


Advancements may include development of alternative adhesive materials to reduce spectral interference automated sample exchange systems and integration with machine learning algorithms for real-time contaminant identification. Expansion to other spectroscopic methods such as Raman or X-ray imaging could further broaden forensic quality control and environmental monitoring capabilities.

Conclusion


The EDXIR-Holder represents an innovative tool for efficient combined FTIR and EDX analysis of contaminants. Its openable film design simplifies workflows reduces manual steps and enhances the reliability of trace elemental and molecular identification tasks. The holder is particularly recommended for use alongside dedicated analysis software to maximize laboratory productivity.

References


  • Shimadzu Application News No. X255
  • Shimadzu Application News No. A522A
  • Shimadzu Application News No. A527

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Solutions for Contaminant Analysis
Solutions for Contaminant Analysis
2023|Shimadzu|Brochures and specifications
C10G-E099 Solutions for Contaminant Analysis Total Support for Contaminant Analysis and Failure Identification The adulteration of products by contaminants causes a wide range of problems in every sector of industry. These problems can only be resolved by analyzing the contaminating…
Key words
contaminant, contaminantedx, edxftir, ftirmicroscope, microscopeedxir, edxirinfrared, infraredaimsight, aimsightraman, ramananalysis, analysisinorganic, inorganicarea, areaairsight, airsightmaterials, materialsyellowed, yellowedatr
Combined Analysis of a Contaminant Using a Compact FTIR and EDX
LAAN-A-FT-E096 Application News No. A567 Spectrophotometric Analysis Combined Analysis of a Contaminant Using a Compact FTIR and EDX Demands regarding the analysis of contaminants that are mixed in or adhered to products are increasing for food and chemical manufacturers and…
Key words
edx, edxedxir, edxircontaminant, contaminantftir, ftirdata, dataanalysis, analysishit, hitray, raystocker, stockermeasurement, measurementcontaminants, contaminantsholder, holderfluorescence, fluorescencefood, foodacquired
Fourier Transform Infrared Spectrophotometer IRSpirit-X Series
Fourier Transform Infrared Spectrophotometer IRSpirit-X Series
2024|Shimadzu|Brochures and specifications
C103-E418A Fourier Transform Infrared Spectrophotometer IRSpirit-X Series IRSpirit , Ready to Run ™ Space-Saving, Expandable • Great for Small Lab Spaces • Standard-Sized Sample Compartment in a Compact Bench FTIR Made Easier • IR Pilot ™ Pre-built Macro Program •…
Key words
irspirit, irspiritdehumidifier, dehumidifierabs, absedx, edxlabsolutions, labsolutionsinfrared, infraredftir, ftirfourier, fourierspectrophotometer, spectrophotometertransform, transformcontaminant, contaminantfunction, functiondata, datainterferometer, interferometerhumidity
Fourier Transform Infrared Spectrophotometer IRSpirit-X Series - C103-E418A
C103-E418A Fourier Transform Infrared Spectrophotometer IRSpirit-X Series IRSpirit , Ready to Run ™ Space-Saving, Expandable • Great for Small Lab Spaces • Standard-Sized Sample Compartment in a Compact Bench FTIR Made Easier • IR Pilot ™ Pre-built Macro Program •…
Key words
irspirit, irspiritdehumidifier, dehumidifierabs, absedx, edxlabsolutions, labsolutionsinfrared, infraredftir, ftirfourier, fourierspectrophotometer, spectrophotometertransform, transformcontaminant, contaminantfunction, functiondata, datahumidity, humidityinterferometer
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike