US EPA Method 524.4 with the Tekmar Lumin P&T Concentrator and Agilent 7890B GC and 5977A MSD System
Applications | 2018 | Agilent TechnologiesInstrumentation
Analysis of volatile organic compounds in drinking water is critical for public health monitoring and regulatory compliance. US EPA Method 524.4 defines a standard approach for trace-level VOC detection, but transferring water vapor to GC-MS instruments can compromise column integrity and analytical quality. Advanced moisture control in purge and trap concentrators can improve method performance and instrument reliability.
This study evaluates the performance of the Tekmar Lumin purge and trap concentrator with integrated moisture control system compared with a conventional AQUATek 100 setup. Using US EPA Method 524.4, the goal was to demonstrate reduced water transfer to an Agilent 7890B GC and 5977A MSD system while maintaining calibration linearity, low method detection limits, and acceptable accuracy and precision.
Water samples were spiked with a range of VOCs from 0.2 ppb to 50 ppb. Internal standards 1,4-difluorobenzene and chlorobenzene-d5 were used along with surrogate standards methyl-t-butyl ether-d3, 4-bromofluorobenzene and 1,2-dichlorobenzene-d4. Seven replicates of 0.5 ppb and 5 ppb standards were analyzed for method detection limit, accuracy, and precision assessment. A linear calibration with 1 over concentration weighting was applied.
Instrumentation Used
Implementing the moisture control system on the Lumin concentrator significantly reduced water load to the GC inlet and MS source. Calibration curves exhibited coefficients of determination above 0.994 for all analytes. Method detection limits ranged from 0.05 to 0.19 ppb. Accuracy across seven replicates at 0.5 ppb fell within 88 to 113 percent, and precision remained below 12.3 percent relative standard deviation. Chromatograms of a 0.2 ppb standard showed clear characteristic ions for early eluting gases with negligible water interference. A total ion chromatogram of a 5 ppb standard confirmed consistent peak shapes and reliable detection for all compounds.
By minimizing water transfer to the GC-MS, laboratories can extend column lifetime, reduce maintenance frequency, and improve data quality for regulatory compliance testing. The Lumin system’s rapid trap cooling and efficient moisture management also enable higher sample throughput over a standard work shift. Lower operational costs are realized through use of nitrogen purge gas and reduced instrument downtime.
Ongoing development of moisture control technology may further lower detection limits and expand applicability to more complex matrices such as wastewater or ground water. Integration of automated diagnostics and real-time moisture monitoring could enhance method robustness. Advances in low-dead-volume connectors and novel trap materials offer additional potential for improving analytical sensitivity and speed.
The Tekmar Lumin purge and trap concentrator with moisture control system meets and exceeds US EPA Method 524.4 requirements when coupled to an Agilent 7890B GC and 5977A MSD. The method provides reliable calibration, low detection limits, high accuracy, and improved instrument protection. Reduced water transfer translates into better chromatography, lower maintenance, and enhanced laboratory efficiency.
1. Munch J W Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography Mass Spectrometry US EPA Method 524.3 Revision 1.0 June 2009
GC/MSD, Purge and Trap, GC/SQ
IndustriesEnvironmental
ManufacturerAgilent Technologies, Teledyne LABS
Summary
Significance of the Topic
Analysis of volatile organic compounds in drinking water is critical for public health monitoring and regulatory compliance. US EPA Method 524.4 defines a standard approach for trace-level VOC detection, but transferring water vapor to GC-MS instruments can compromise column integrity and analytical quality. Advanced moisture control in purge and trap concentrators can improve method performance and instrument reliability.
Objectives and Study Overview
This study evaluates the performance of the Tekmar Lumin purge and trap concentrator with integrated moisture control system compared with a conventional AQUATek 100 setup. Using US EPA Method 524.4, the goal was to demonstrate reduced water transfer to an Agilent 7890B GC and 5977A MSD system while maintaining calibration linearity, low method detection limits, and acceptable accuracy and precision.
Methodology and Instrumentation
Water samples were spiked with a range of VOCs from 0.2 ppb to 50 ppb. Internal standards 1,4-difluorobenzene and chlorobenzene-d5 were used along with surrogate standards methyl-t-butyl ether-d3, 4-bromofluorobenzene and 1,2-dichlorobenzene-d4. Seven replicates of 0.5 ppb and 5 ppb standards were analyzed for method detection limit, accuracy, and precision assessment. A linear calibration with 1 over concentration weighting was applied.
Instrumentation Used
- Tekmar Lumin purge and trap concentrator with moisture control system
- Agilent 7890B gas chromatograph equipped with a DB-624 UI column (20 m x 0.18 mm id, 1 µm film) and helium carrier at 0.8 mL per minute
- Agilent 5977A mass selective detector operated in scan mode from 35 to 260 atomic mass units
Main Results and Discussion
Implementing the moisture control system on the Lumin concentrator significantly reduced water load to the GC inlet and MS source. Calibration curves exhibited coefficients of determination above 0.994 for all analytes. Method detection limits ranged from 0.05 to 0.19 ppb. Accuracy across seven replicates at 0.5 ppb fell within 88 to 113 percent, and precision remained below 12.3 percent relative standard deviation. Chromatograms of a 0.2 ppb standard showed clear characteristic ions for early eluting gases with negligible water interference. A total ion chromatogram of a 5 ppb standard confirmed consistent peak shapes and reliable detection for all compounds.
Benefits and Practical Applications
By minimizing water transfer to the GC-MS, laboratories can extend column lifetime, reduce maintenance frequency, and improve data quality for regulatory compliance testing. The Lumin system’s rapid trap cooling and efficient moisture management also enable higher sample throughput over a standard work shift. Lower operational costs are realized through use of nitrogen purge gas and reduced instrument downtime.
Future Trends and Opportunities
Ongoing development of moisture control technology may further lower detection limits and expand applicability to more complex matrices such as wastewater or ground water. Integration of automated diagnostics and real-time moisture monitoring could enhance method robustness. Advances in low-dead-volume connectors and novel trap materials offer additional potential for improving analytical sensitivity and speed.
Conclusion
The Tekmar Lumin purge and trap concentrator with moisture control system meets and exceeds US EPA Method 524.4 requirements when coupled to an Agilent 7890B GC and 5977A MSD. The method provides reliable calibration, low detection limits, high accuracy, and improved instrument protection. Reduced water transfer translates into better chromatography, lower maintenance, and enhanced laboratory efficiency.
Reference
1. Munch J W Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography Mass Spectrometry US EPA Method 524.3 Revision 1.0 June 2009
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
US EPA Method 524.4 with the Tekmar Lumin P & T Concentrator and Agilent 7890B GC and 5977A MSD System
2018|Agilent Technologies|Applications
US EPA Method 524.4 with the Tekmar Lumin P & T Concentrator and Agilent 7890B GC and 5977A MSD System Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA Method 524.4 is used…
Key words
temp, tempbake, bakepurge, purgeether, etherlpir, lpirppb, ppbupir, upirlumin, lumindesorb, desorbvariable, variablemcs, mcsprecision, precisionmdl, mdltable, tablebutylbenzene
US EPA Method 524.4 with the Tekmar Lumin P & T Concentrator and Thermo Scientific™ TRACE™ 1310 GC and ISQ™ LT MS System
2017|Thermo Fisher Scientific|Applications
US EPA Method 524.4 with the Tekmar Lumin P & T Concentrator and Thermo Scientific™ TRACE™ 1310 GC and ISQ™ LT MS System Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA Method…
Key words
temp, temppurge, purgebake, bakeether, etherdesorb, desorblpir, lpirppb, ppbupir, upirlumin, luminvariable, variablemcs, mcsprecision, precisionmdl, mdltime, timetable
US EPA Method 524.3 with the Tekmar Lumin P&T Concentrator and Agilent 7890B GC and 5977A MSD System
2018|Agilent Technologies|Applications
Application Note US EPA Method 524.3 with the Tekmar Lumin P&T Concentrator and Agilent 7890B GC and 5977A MSD System Author Roger Bardsley Applications Chemist, Teledyne Tekmar Abstract US EPA Method 524.3 is used primarily by environmental labs for the…
Key words
temp, temppurge, purgebake, bakelumin, luminwater, waterether, etherdesorb, desorbtekmar, tekmarvariable, variabletime, timemcs, mcsteledyne, teledyneamount, amountlpir, lpirbutylbenzene
US EPA Method 524.3 with the Tekmar Lumin P & T Concentrator and Thermo Scientific™ TRACE™ 1310 GC and ISQ™ LT MS System
2017|Thermo Fisher Scientific|Applications
US EPA Method 524.3 with the Tekmar Lumin P & T Concentrator and Thermo Scientific™ TRACE™ 1310 GC and ISQ™ LT MS System Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA Method…
Key words
temp, temppurge, purgebake, bakeether, etherdesorb, desorblpir, lpirppb, ppbupir, upirlumin, luminvariable, variablemcs, mcsprecision, precisionmdl, mdltime, timetable