GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Reducing Analysis Time of 8270D with the Intuvo 9000 GC

Applications | 2018 | Agilent TechnologiesInstrumentation
GC/MSD, GC/SQ
Industries
Environmental
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Gas chromatography–mass spectrometry (GC–MS) methods for semivolatile organic compounds (SVOCs) under EPA Method 8270D are critical in environmental monitoring of soil, waste, water and air. Accelerating analysis while preserving regulatory compliance enhances laboratory throughput and cost-effectiveness in routine and high-volume testing.

Objectives and Overview


This study describes the translation of a conventional EPA 8270D GC–MS method using a 30 m×0.25 mm column into a faster protocol on an Agilent Intuvo 9000 GC equipped with a conductively heated 20 m×0.18 mm column. Key aims included reducing overall run time by at least 30 % and maintaining required resolution of critical isomer pairs.

Methodology and Instrumentation


A 110-analyte standard mix (acids, bases, neutrals) at 50 µg/mL plus six internal standards at 40 µg/mL in dichloromethane was used for method translation. Temperature and pressure programming from the original four-ramp method were scaled for the narrower bore column using Agilent method translation software. The optimized flow rate was increased to 1.2 mL/min with a split ratio of 10:1 to accommodate the reduced column capacity and prevent overload.

Used Instrumentation


  • Agilent Intuvo 9000 GC with split/splitless Guard Chip
  • Agilent 5977A MSD with EI inert source and 6 mm drawout plate
  • Agilent J&W DB-UI 8270D column, 30 m×0.25 mm, 0.25 µm film (original)
  • Agilent J&W DB-5ms UI Intuvo column, 20 m×0.18 mm, 0.18 µm film (translated)
  • Ultra Inert liners: splitless single-taper (p/n 5190-2293) and split low-pressure drop (p/n 5190-2295)

Results and Discussion


The fast method reduced total run time by a factor of approximately 1.4 compared to the original 30 m column. Resolution of the SVOC critical isomer pair benzo[b]fluoranthene/benzo[k]fluoranthene met the EPA 8270D requirement with a valley-height ratio of 33.5 % (limit 50 %). The extracted-ion chromatogram for indenopyrene (m/z 276) and dibenzoanthracene (m/z 278) achieved 34.3 % separation. Adjusting the Guard Chip starting temperature from 70 °C (track-oven mode) to 200 °C (programmed) improved peak shapes for low-volatility compounds (1,4-dioxane, N-nitrosodimethylamine, pyridine), reducing tailing factors to below 2.5 at 5 % peak height.

Benefits and Practical Applications


  • Throughput increase via shorter analysis times, supporting high-volume environmental testing
  • Maintained compliance with EPA method resolution and quantitation requirements
  • Reduced solvent and carrier gas consumption per analysis
  • Improved peak shapes and quantitation reliability for early-eluting SVOCs through Guard Chip temperature programming

Future Trends and Potential Applications


Emerging developments in fast GC–MS include further miniaturized column formats, advanced microfluidic heating systems, automated method translation with AI-driven optimization, and integration with high-resolution MS detectors. Such innovations will broaden applications in environmental forensics, industrial QA/QC and regulatory screening.

Conclusion


The translated method on the Intuvo 9000 GC reduced run time by ~30 % while preserving critical isomer resolution and quantitation performance per EPA 8270D. Instrumental adjustments, including optimized flow and Guard Chip programming, enabled robust analysis of 110 SVOCs, demonstrating a viable fast-GC approach for routine environmental laboratories.

Reference


Semivolatile Organic Compounds by Gas Chromatography/Mass Spectrometry (GC/MS); Method 8270D; United States Environmental Protection Agency, Revision 4, February 2007.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9000 Gas Chromatograph
Application Note Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9000 Gas Chromatograph Author Matthew Giardina, Ph.D., Applications Chemist, Gas Phase Separations Division, Agilent Technologies, Inc. Abstract Gas chromatography/mass spectrometry is a workhorse for the analysis of semivolatile organic…
Key words
intuvo, intuvochip, chipreplacement, replacementguard, guardpass, passccv, ccvbenzidine, benzidinebenzo, benzomatrix, matrixresponse, responsecalibration, calibrationpentachlorophenol, pentachlorophenolafter, afternumber, numberinjections
GC/MS Analysis of Semivolatile Organic Compounds Using an Agilent J&W VF-5ms Intuvo GC Column
Application Note Environmental GC/MS Analysis of Semivolatile Organic Compounds Using an Agilent J&W VF-5ms Intuvo GC Column Authors Yun Zou Agilent Technologies (Shanghai) Co. Ltd, Shanghai 200131, P. R. China John Oostdijk Agilent Technologies, Inc. Middelburg, The Netherlands Abstract To…
Key words
endrin, endrinsemivolatile, semivolatilecounts, countsanthracene, anthracenebenzo, benzocolumn, columnchlordane, chlordaneheptachlor, heptachlorphthalate, phthalatedftpp, dftppintuvo, intuvocompounds, compoundsbhc, bhcphenanthrene, phenanthrenechrysene
Increased Reproducibility in the Analysis of EU and EPA PAHs with the Agilent J&W Select PAH GC Column and Agilent Intuvo 9000 GC
Application Note Environmental Increased Reproducibility in the Analysis of EU and EPA PAHs with the Agilent J&W Select PAH GC Column and Agilent Intuvo 9000 GC Author Vanessa Abercrombie Agilent Technologies, Inc. Abstract This Application Note demonstrates the use of…
Key words
pah, pahsalmon, salmonchip, chipmicrofluidic, microfluidicbenzo, benzointuvo, intuvocounts, countsselect, selectacquisition, acquisitionepa, epaguard, guardsim, simpahs, pahsoil, oilmsd
Characterization of a Temperature and Flow-Programmable Microfluidic Precolumn for Gas Chromatography
Application Note Environmental Characterization of a Temperature and Flow-Programmable Microfluidic Precolumn for Gas Chromatography Author Matthew Giardina, Ph.D. Agilent Technologies, Inc. Wilmington, USA Abstract This Application Note investigates the temperature control of Agilent Intuvo Guard Chips as a means of…
Key words
chip, chipguard, guardpulsing, pulsingbackflush, backflushtemperature, temperatureprogram, programmin, mintime, timeretention, retentionprograms, programsmicrofluidic, microfluidicoven, ovenexperiments, experimentsflow, flowprograming
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike