GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

GC Integrated Permeation Device

Applications | 2011 | PACInstrumentation
GC
Industries
Energy & Chemicals
Manufacturer
Agilent Technologies, PAC

Summary

Importance of the Topic


Calibrating trace-level reactive gases presents challenges due to adsorption and reactive losses when using pressurized cylinders. Permeation devices enable stable, low-concentration standards without high-pressure gas handling.

Objectives and Study Overview


This note evaluates an integrated GC permeation system for ppb–ppm level sulfur calibration. It aims to assess repeatability, accuracy, system inertness, and operational benefits in a single- or dual-oven configuration.

Used Instrumentation


  • AC GC-integrated permeation device with one or two temperature-controlled ovens
  • Permeation tubes for H2S, COS, DMS, MeSH, EtSH
  • Agilent 7890 Gas Chromatograph coupled to Antek 7090 Sulfur Chemiluminescence Detector
  • Mass flow controllers (20–200 ml/min) and deactivated sample lines


Methodology


A permeation tube containing liquid analyte within a thermally controlled chamber releases vapor at a constant rate. An inert carrier gas flow mixes to yield known ppm/ppb concentrations. Dual ovens allow simultaneous calibration of chemically incompatible compounds at distinct temperatures.

Main Results and Discussion


  • Repeatability: Over 200 analyses across four days showed RSD ≤1.5% for H2S, COS, DMS.
  • Accuracy: Bias remained within ±10% against certified cylinder standards at both 20 and 200 ml/min flows for five sulfur species.
  • Inertness: Deactivated flow paths facilitated reliable multipoint calibration; ~30 ppb levels achieved with RSDs typically <15%.


Benefits and Practical Applications


  • Eliminates need for bulky, high-pressure cylinders, saving lab space and reducing transport risk.
  • Enables continuous on-demand, NIST-traceable gas generation with minimal reactive losses.
  • Sealed tubes contain low chemical quantities, enhancing safety.
  • Flexible configuration supports diverse matrices and multi-component calibration curves.


Future Trends and Potential Applications


  • Expansion of permeation tube libraries for emerging trace analytes.
  • Integration with automated, high-throughput multi-channel GC systems.
  • Refined temperature and flow control to improve detection limits and dynamic range.
  • Deployment in environmental monitoring, process control, and in situ analysis.


Conclusion


The AC integrated permeation device delivers a safe, flexible, and accurate alternative to traditional gas cylinders for trace-level calibration. Its dual-oven capability, inert sample paths, and NIST-traceable rates ensure consistent performance for reactive sulfur species under varied analytical conditions.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
UOP 603-13: Analysis of Trace CO and CO2 in bulk H2 and Light Gaseous Hydrocarbons by GC
APPLICATION NOTE UOP 603-13: Analysis of Trace CO and CO2 in bulk H2 and Light Gaseous Hydrocarbons by GC • • • • Fast Analysis in <5 Minutes Excellent Repeatability & Linearity Robust Solution using AC proprietary Methanizer No Matrix…
Key words
methane, methaneuae, uaecarbon, carbonthailand, thailandrussia, russiappm, ppmmonoxide, monoxidesingapore, singaporekorea, koreasouth, southdioxide, dioxidefrance, francenetherlands, netherlandsindia, indiaarea
FAME analysis in Diesel by Flow Modulated GCxGC FID
APPLICATION NOTE FAME analysis in Diesel by Flow Modulated GCxGC FID. INTRODUCTION • • • • Dedicated PiPNA + FAME For (Bio-)Diesel and Jet Fuels Robust System, Easy to use No Cryogenic coolant Required Keywords: FAME in Diesel, Flow Modulation,…
Key words
fame, famemodulation, modulationgcxgc, gcxgcuae, uaethailand, thailandrussia, russiasingapore, singaporediesel, dieselkorea, koreasouth, southfrance, francenetherlands, netherlandsindia, indiaflow, flowchina
FAME analysis in Diesel by Flow Modulated GCxGC FID
APPLICATION NOTE FAME analysis in Diesel by Flow Modulated GCxGC FID. INTRODUCTION • • • • Dedicated PiPNA + FAME For (Bio-)Diesel and Jet Fuels Robust System, Easy to use No Cryogenic coolant Required Keywords: FAME in Diesel, Flow Modulation,…
Key words
fame, famemodulation, modulationgcxgc, gcxgcuae, uaethailand, thailandrussia, russiasingapore, singaporekorea, koreadiesel, dieselsouth, southfrance, francenetherlands, netherlandsindia, indiaflow, flowchina
Group-Type Analysis (PiPNA) in Diesel and Jet Fuel by Flow Modulated GCxGC FID
APPLICATION NOTE Group-Type Analysis (PiPNA) in Diesel and Jet Fuel by Flow Modulated GCxGC FID. INTRODUCTION • • • • Dedicated PiPNA + FAME For (Bio-)Diesel and Jet Fuels Robust System, Easy to use No Cryogenic coolant Required Recent developments…
Key words
modulation, modulationuae, uaethailand, thailandrussia, russiasingapore, singaporenaphtenes, naphtenesfame, famekorea, koreasouth, southfrance, francenetherlands, netherlandsindia, indiachina, chinagcxgc, gcxgcgermany
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike