Volatile Organic Compound Screening in Soil Using SPME-GC/MS

Applications | 2015 | PerkinElmerInstrumentation
GC/MSD, SPME, GC/IT
Industries
Environmental
Manufacturer
PerkinElmer

Summary

Význam tématu


The ability to perform rapid, on-site screening of volatile organic compounds (VOCs) in soils is critical for environmental monitoring, risk assessment and real-time decision making during field investigations.

Cíle a přehled studie


This application note evaluates a portable solid phase microextraction (SPME) coupled to a compact gas chromatograph/mass spectrometer (GC/MS) system for in-field VOC analysis in soil. The main objective was to demonstrate positive identification of multiple target VOCs and estimate their relative concentrations within minutes, following U.S. EPA guidelines for near-real-time screening.

Použitá instrumentace


The field workflow employed a Custodion SPME syringe fitted with a 65 µm PDMS/DVB fiber to extract VOCs from saline water-treated soil headspace. Desorption and separation were achieved with a Torion T-9 portable GC/MS using:
  • MXT-5 capillary column (5 m × 0.1 mm, 0.4 µm film)
  • Helium carrier gas at constant flow (0.2 mL/min)
  • Injector temperature of 250 °C and a temperature program from 40 °C to 280 °C at 2 °C/s (total run under 2.5 min)
  • Toroidal ion trap mass analyzer (mass range 41–500 Da, ∼10 Hz scan rate) with in-trap electron impact ionization and electron multiplier detection

Hlavní výsledky a diskuse


In a spiked reference soil sample containing 42 VOCs (2.4–15 ppm w/w), 37 compounds were positively identified in under two minutes. Co-elution of several analytes was resolved by on-board spectral deconvolution and a custom target library. Five highly volatile compounds were not detected, likely due to losses during extraction or selective fiber discrimination. Overall method precision and speed support effective field screening.

Přínosy a praktické využití metody


This portable SPME-GC/MS approach eliminates the need for solvents and laboratory infrastructure, reducing sampling costs and turnaround times. On-site positive identification of individual VOCs guides further sampling strategies and minimizes expensive re-analysis and instrument downtime due to carryover.

Budoucí trendy a možnosti využití


Future developments may include advanced SPME coatings for broader analyte coverage, integration of automated data processing and cloud-based reporting, and expanded target libraries for emerging contaminants. Continued miniaturization and battery operation will further enhance in-field capabilities.

Závěr


The combination of Custodion SPME sampling and Torion T-9 GC/MS detection provides a robust, rapid and portable solution for VOC screening in soil. Its ease of use and fast turnaround support informed decision making during environmental field investigations.

Reference


  1. EPA SW-846 Method 3815 Screening Solid Samples for Volatile Organics, Revision 0, February 2007.
  2. Zhang Z.; Pawliszyn J. Analysis for organic compounds in environmental samples by headspace solid phase microextraction. Journal of High Resolution Chromatography. 1993, 16(12), 689-692.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Volatile Organic Compound Screening in Soil Using SPME-GC/MS
A P P L I C AT I O N N O T E Gas Chromatography/ Mass Spectrometry Author: Portable GC/MS Team PerkinElmer, Inc. Shelton, CT Volatile Organic Compound Screening in Soil Using SPME-GC/MS Introduction The ability to screen soil…
Key words
vocs, vocsspme, spmevolatile, volatileorganic, organicsoil, soildivinylbenzene, divinylbenzenecompounds, compoundsscreening, screeningfiber, fibercustodion, custodionretracted, retractedmay, mayelectron, electronroughing, roughingtoroidal
Volatile Organic Compound Identification in Water by Portable SPME-GC/MS
A P P L I C AT I O N N O T E Gas Chromatography/ Mass Spectrometry Authors: Portable GC/MS Team Shelton, CT Volatile Organic Compound Identification in Water by Portable SPME-GC/MS Introduction Groundwater water sources are routinely tested…
Key words
spme, spmecustodion, custodionvolatile, volatiletms, tmsorganic, organicportable, portablevocs, vocsethylbenzene, ethylbenzenefiber, fiberwater, waterretracted, retracteddegradation, degradationelectron, electronrefrigeration, refrigerationroughing
Custodion Syringes for Sampling Collection
Custodion Syringes for Sampling Collection
2019|PerkinElmer|Brochures and specifications
IT’S NOT JUST A SYRINGE Custodion® Syringes for Sample Collection Custodion SPME Syringes Custodion CME Syringes Custodion NT Syringes From Collection to Injection Crime scenes, chemical spills, environmental contaminants – our Custodion family of sampling devices enable you to accurately…
Key words
syringe, syringecustodion, custodioncme, cmehazmat, hazmatspme, spmeclick, clickportable, portablelearn, learnsyringes, syringeshere, hereforensic, forensicabout, aboutcontact, contactcustodial, custodialsampling
MTBE and BTEX On-Site Rapid Screening of Contaminated Ground Water by Portable SPME-GC/MS Methodology
A P P L I C AT I O N N O T E Gas Chromatography/ Mass Spectrometry Authors: Portable GC/MS Team PerkinElmer, Inc. Shelton, CT MTBE and BTEX On-Site Rapid Screening of Contaminated Ground Water by Portable SPME-GC/MS Methodology…
Key words
spme, spmemtbe, mtbebtex, btextms, tmstoroidal, toroidalground, groundpetroleum, petroleummass, massportable, portabletrap, trapmicroextraction, microextractioninfield, infieldcontaminated, contaminatedsolid, solidwater