Indecisive About Your Column Selection? Let Us Help You Choose the Correct GC Column
Presentations | 2022 | Agilent TechnologiesInstrumentation
The choice of a gas chromatography column is crucial for analytical performance and reliable results in fields such as environmental testing, food analysis, pharmaceuticals and petrochemicals. Differences in stationary phase chemistry, column dimensions and inertness directly impact retention, selectivity, resolution and repeatability of complex mixtures.
This application note by Agilent outlines key considerations for selecting the correct GC column. It reviews the evolution of capillary columns, contrasts stationary phase chemistries, describes performance metrics such as bleed and inertness and provides practical guidance on dimension tuning for specific applications.
The study leverages Agilent J&W fused silica capillary columns of various phase chemistries including nonpolar methyl polysiloxanes, phenyl polysiloxanes, cyanopropyl and polyethylene glycol phases. Column bleed performance was assessed under isothermal conditions at high temperature. Inertness was evaluated using test mixes of active compounds and comparative detector response. Selectivity studies compared standard and Ultra Inert versions of DB-5 and WAX phases. Key instrumentation included an Agilent gas chromatograph configured with flame ionization and mass spectrometric detectors and controlled carrier gas flows in hydrogen or helium.
Bleed tests at 280 °C over 100 hours demonstrated the superior low bleed of DB-HeavyWAX versus other WAX phases. Ultra Inert columns showed minimal peak tailing and loss of active analytes such as acids and bases. Comparative selectivity experiments revealed that DB-5ms offers improved stability and maintained key separations of aromatic isomers versus standard DB-5. Medium polarity phases like DB-35ms provided enhanced response for high molecular weight aromatics. WAX and polar phases showed predictable retention shifts for fatty acids and esters, with UI coatings delivering sharper peaks for polar solutes.
Applying these findings enables analysts to:
Ongoing development focuses on high temperature polymer additive columns capable of operating above 350 °C, novel low bleed chemistries such as DB-XLB, and advanced Ultra Inert coatings to enhance analysis of challenging compounds in metabolomics, petrochemical and environmental research. Integration of advanced selectivity tuning with two dimensional chromatography and fast GC methods represents the next frontier.
Successful GC method development starts with thorough understanding of sample volatility, matrix effects and analyte interactions. By selecting the appropriate stationary phase chemistry and column dimensions, one can achieve optimal separation performance, maintain column stability and meet analytical throughput requirements.
GC columns, Consumables
IndustriesManufacturerAgilent Technologies
Summary
Significance of the Topic
The choice of a gas chromatography column is crucial for analytical performance and reliable results in fields such as environmental testing, food analysis, pharmaceuticals and petrochemicals. Differences in stationary phase chemistry, column dimensions and inertness directly impact retention, selectivity, resolution and repeatability of complex mixtures.
Objectives and Article Overview
This application note by Agilent outlines key considerations for selecting the correct GC column. It reviews the evolution of capillary columns, contrasts stationary phase chemistries, describes performance metrics such as bleed and inertness and provides practical guidance on dimension tuning for specific applications.
Methodology and Instrumentation
The study leverages Agilent J&W fused silica capillary columns of various phase chemistries including nonpolar methyl polysiloxanes, phenyl polysiloxanes, cyanopropyl and polyethylene glycol phases. Column bleed performance was assessed under isothermal conditions at high temperature. Inertness was evaluated using test mixes of active compounds and comparative detector response. Selectivity studies compared standard and Ultra Inert versions of DB-5 and WAX phases. Key instrumentation included an Agilent gas chromatograph configured with flame ionization and mass spectrometric detectors and controlled carrier gas flows in hydrogen or helium.
Main Results and Discussion
Bleed tests at 280 °C over 100 hours demonstrated the superior low bleed of DB-HeavyWAX versus other WAX phases. Ultra Inert columns showed minimal peak tailing and loss of active analytes such as acids and bases. Comparative selectivity experiments revealed that DB-5ms offers improved stability and maintained key separations of aromatic isomers versus standard DB-5. Medium polarity phases like DB-35ms provided enhanced response for high molecular weight aromatics. WAX and polar phases showed predictable retention shifts for fatty acids and esters, with UI coatings delivering sharper peaks for polar solutes.
Benefits and Practical Application
Applying these findings enables analysts to:
- Match sample polarity and target analyte chemistry to stationary phase interactions
- Choose high temperature or low bleed phases to extend column life in demanding matrices
- Tune column length, diameter and film thickness to optimize resolution, throughput and sensitivity
- Select Ultra Inert columns when peak shape and detector response for active compounds is critical
Future Trends and Applications
Ongoing development focuses on high temperature polymer additive columns capable of operating above 350 °C, novel low bleed chemistries such as DB-XLB, and advanced Ultra Inert coatings to enhance analysis of challenging compounds in metabolomics, petrochemical and environmental research. Integration of advanced selectivity tuning with two dimensional chromatography and fast GC methods represents the next frontier.
Conclusion
Successful GC method development starts with thorough understanding of sample volatility, matrix effects and analyte interactions. By selecting the appropriate stationary phase chemistry and column dimensions, one can achieve optimal separation performance, maintain column stability and meet analytical throughput requirements.
References
- Ucci A Indecisive About Your Column Selection Let Us Help You Choose the Correct GC Column Agilent Technologies Application Note DE10230776 October 6 2022
- Agilent Application Note 5991-6709EN Ultra Inert DB-WAX UI
- Agilent Application Note 5991-6683EN DB-WAX UI comparative evaluation
- Agilent Application Note 5991-8706EN Fast FAME Analysis with DB-FastFAME
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Overwhelmed with Too Many GC Column Options? Let Us Help
2023|Agilent Technologies|Presentations
Overwhelmed with Too Many GC Column Options? Let Us Help Mark Sinnott Online Application Engineer August 30th , 2023 1 Indecisive About your Column Selection? DE31969766 JW Column Portfolio- DB, HP, CP, VF Low Polarity Mid Polarity High Polarity CP-Sil…
Key words
indecisive, indecisiveyes, yesselection, selectiondipole, dipoleabout, aboutyour, yourcolumn, columnbonding, bondingstrong, stronghydrogen, hydrogendispersion, dispersioninduced, inducedwax, waxhexanol, hexanolpolarity
Decisions, Decisions: How to Select the Correct GC Column for Your Application
2021|Agilent Technologies|Presentations
Decisions, Decisions: How to Select the Correct GC Column for Your Application Alexander Ucci Online Application Engineer July 21, 2021 1 July 21, 2021 Decisions Decisions: How to Select the Correct GC Column for Your Application DE44396.3167708333 Things to Consider…
Key words
yes, yesdipole, dipolebonding, bondinghydrogen, hydrogenstrong, strongdispersion, dispersionhexanol, hexanolinduced, inducedthickness, thicknessfilm, filmdecane, decanecolumn, columnpolarity, polaritydiameter, diameterinteraction
Finding the Perfect Match: Practical Advice on GC Column Selection
2019|Agilent Technologies|Presentations
Finding the Perfect Match: Practical Advice on GC Column Selection Alexander Ucci Online Application Engineer September 12, 2019 1 September 12, 2019 Finding the Perfect Match: Practical Advice on GC Column Selection Agilent Restricted Things to Consider when Choosing a…
Key words
restricted, restrictedyes, yesagilent, agilentdipole, dipolebonding, bondinghydrogen, hydrogenstrong, strongdispersion, dispersionhexanol, hexanolinduced, inducedthickness, thicknessfilm, filmwax, waxdecane, decanecolumn
Understanding the Capillary GC Column: How to Choose the Correct Type and Dimension
2015|Agilent Technologies|Presentations
Understanding the Capillary GC Column: How to Choose the Correct Type and Dimension Simon Jones Application Engineer Things to Consider… • Is it Volatile enough to chromatograph by GC? • Is it a Gas or a Liquid? • How are…
Key words
title, titlepresentation, presentationrestricted, restrictedgroup, groupagilent, agilentdipole, dipolestrong, strongdispersion, dispersionbonding, bondingthickness, thicknessfilm, filmdiameter, diametermoderate, moderatestationary, stationarycnpr