Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler
Applications | | Agilent TechnologiesInstrumentation
The analysis of volatile organic compounds (VOCs) in environmental and industrial matrices is critical for ensuring water and soil quality, complying with regulatory standards, and protecting public health. Advanced purge-and-trap techniques combined with sensitive gas chromatography–mass spectrometry (GC–MS) provide reliable detection of trace VOCs.
This application note aims to optimize operating conditions for the Markes Eclipse 4760 purge-and-trap concentrator and the 4100 autosampler following USEPA Method 8260. A multi-point calibration spanning 2 to 200 ppb was established for both liquid and solid matrices to verify method performance.
Calibration produced consistent response factors for over 80 analytes in both liquid and solid preparations, with relative standard deviations generally below 10% and rarely exceeding 15%. Representative chromatograms at 50 ppb demonstrated clear separation, symmetric peak shapes, and stable retention times across the analyte panel. Both volatile and semi-volatile species met USEPA Method 8260 criteria.
The optimized configuration of the 4760 purge-and-trap concentrator and 4100 autosampler under USEPA Method 8260 delivered robust and reproducible results for a broad VOC target list. Method performance met all regulatory benchmarks, supporting reliable environmental analysis.
No literature citations were provided in the original document.
GC/MSD, Purge and Trap, GC/SQ
IndustriesEnvironmental
ManufacturerAgilent Technologies, OI Analytical
Summary
Importance of the Topic
The analysis of volatile organic compounds (VOCs) in environmental and industrial matrices is critical for ensuring water and soil quality, complying with regulatory standards, and protecting public health. Advanced purge-and-trap techniques combined with sensitive gas chromatography–mass spectrometry (GC–MS) provide reliable detection of trace VOCs.
Objectives and Study Overview
This application note aims to optimize operating conditions for the Markes Eclipse 4760 purge-and-trap concentrator and the 4100 autosampler following USEPA Method 8260. A multi-point calibration spanning 2 to 200 ppb was established for both liquid and solid matrices to verify method performance.
Methodology
- Calibration range: 2–200 ppb using standard mixtures.
- Purge-and-trap settings:
- Trap: Tenax®/silica gel/carbosieve; purge gas: helium at 40 mL/min; purge time: 11 min; desorb time: 0.5 min; bake time: 4 min; trap temperatures ramped from ambient to 210 °C.
- Autosampler protocols:
- Water samples: blue vial caps, one rinse, 5 µL injection, purge and desorb as above.
- Soil samples: yellow caps, addition of water, 5 µL injection, pre-heat and stir at 45 °C during purge.
- GC–MS conditions:
- Gas chromatograph: Agilent 7890A with Restek Rxi-624Sil MS column (30 m × 0.25 mm × 1.4 µm), helium carrier at 0.8 mL/min, split ratio 150:1.
- Oven program: 40 °C hold for 1.5 min, ramp to 180 °C at 16 °C/min, then to 220 °C at 40 °C/min, hold 1.75 min (total run 13 min).
- Mass spectrometer: Agilent 5975C, scan 35–300 amu at 5.19 scans/s, transfer line 250 °C, source 300 °C, quadrupole 200 °C.
Instrumentation Used
- Eclipse 4760 purge-and-trap sample concentrator
- 4100 Water/Soil Autosampler
- Gas chromatograph: Agilent 7890A
- Mass spectrometer: Agilent 5975C
- Column: Restek Rxi-624Sil MS
Results and Discussion
Calibration produced consistent response factors for over 80 analytes in both liquid and solid preparations, with relative standard deviations generally below 10% and rarely exceeding 15%. Representative chromatograms at 50 ppb demonstrated clear separation, symmetric peak shapes, and stable retention times across the analyte panel. Both volatile and semi-volatile species met USEPA Method 8260 criteria.
Benefits and Practical Applications
- Sensitive detection of trace VOCs in environmental monitoring (water, soil).
- High throughput enabled by automated sampling and a rapid 13 min GC run.
- Adaptable workflow for regulatory compliance and quality control.
- Improved water management and reduced matrix interferences through controlled purge-and-trap temperatures.
Future Trends and Potential Applications
- Integration with high-resolution mass spectrometry for enhanced selectivity.
- Miniaturized or field-deployable purge-and-trap modules for in-situ analysis.
- Advanced data processing and machine learning to automate peak identification and quantification.
- Expansion to emerging contaminants such as halogenated flame retardants and novel industrial solvents.
Conclusion
The optimized configuration of the 4760 purge-and-trap concentrator and 4100 autosampler under USEPA Method 8260 delivered robust and reproducible results for a broad VOC target list. Method performance met all regulatory benchmarks, supporting reliable environmental analysis.
References
No literature citations were provided in the original document.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Comparison of Helium and Nitrogen Purge Gases for the Analysis of Volatile Organic Compounds (VOCs) by Purge and Trap
2019|Agilent Technologies|Applications
Application Note Comparison of Helium and Nitrogen Purge Gases for the Analysis of Volatile Organic Compounds (VOCs) by Purge and Trap Introduction Although analysis of VOCs by purge and trap (P&T) is considered a mature technique, advances in methodology and…
Key words
helium, heliumnitrogen, nitrogenbutylbenzene, butylbenzenersd, rsdmethyl, methylchloride, chloridepurge, purgemethacrylate, methacrylatetert, tertisopropytoluene, isopropytolueneether, ethersolid, solidsoil, soilanalyte, analyteethyl
Analysis of Volatile Organic Compounds Using USEPA Method 524.3
|Agilent Technologies|Applications
Analysis of Volatile Organic Compounds Using USEPA Method 524.3 Eclipse 4760 Purge and Trap and the 4100 Autosampler Introduction Although analysis of VOCs by purge and trap is considered a mature technique, advances in methodology and refinement of instrumentation are…
Key words
tert, tertstd, stdether, etherbutylbenzene, butylbenzeneethyl, ethylmethyl, methylbutyl, butylpurge, purgeamyl, amylppb, ppbsoil, soilchloride, chloridesam, samchlorodibromomethane, chlorodibromomethanechlorodifluoromethane
Analysis of EDB, 123-TCP, and DBCP By USEPA Method 524.3
|Agilent Technologies|Applications
Analysis of EDB, 123-TCP, and DBCP By USEPA Method 524.3 Eclipse 4760 Purge and Trap and the 4100 Autosampler Introduction Many laboratories analyze 1,2-Dibromoethane (EDB), 1,2,3-Trichloropropane (123-TCP), and 1,2-Dibromo3-chloropropane (DBCP) in drinking water by USEPA Method 504.1 or USEPA 8011.…
Key words
rec, recstd, stdpurge, purgesoil, soilsam, samstir, stirtemperature, temperaturepir, pirdesorb, desorbwater, watergas, gaszero, zerotime, timerinses, rinsesgrade
Analysis of 1,4-Dioxane by Purge and Trap and Gas Chromatogaphy/Mass Spectrometry
|Agilent Technologies|Applications
Analysis of 1,4-Dioxane by Purge and Trap and Gas Chromatogaphy/Mass Spectrometry Introduction 1,4-Dioxane is a heterocyclic molecule that belongs to a class known as ethers. This compound has been used in the past as a stabilizer for different types of…
Key words
chromatogaphy, chromatogaphypurge, purgesam, samtemperature, temperatureidoc, idocgas, gastime, timewater, waterdesorb, desorbvolatiles, volatilestrap, trapsoil, soilmdl, mdlzero, zerospectrometry