Two - Step Polymer Analysis

Applications |  | CDS AnalyticalInstrumentation
GC/MSD, Pyrolysis
Industries
Energy & Chemicals
Manufacturer
Agilent Technologies, CDS Analytical

Summary

Significance of the Topic

The analysis of polymer-based products requires reliable detection of residual monomers, solvents and plasticizers. Conventional solvent extractions are time-consuming, dilute samples and generate chemical waste. Advanced thermal sampling technology enables rapid, automated separation of volatile additives and polymer fragments without solvents, improving throughput and environmental safety.

Objectives and Study Overview

This application note illustrates a two-step polymer analysis workflow. In the first step, a low-temperature desorption isolates plasticizer additives. In the second step, high-temperature pyrolysis decomposes the polymer matrix for identification of monomers and degradation products. A clear packaging plastic containing poly(vinyl chloride), poly(methyl methacrylate) and polystyrene plasticized with dioctyl phthalate (DOP) served as the test sample.

Methodology and Instrumentation

  • Thermal sampler: CDS Model 2500+ Pyrolysis Autosampler
  • Gas chromatograph: HP 6890 GC equipped with HP-5 capillary column (30 m × 0.25 mm)
  • Detector: HP 5973 Mass Selective Detector (MSD)
  • Desorption conditions: 300 °C for 15 s, split ratio 50:1, helium carrier
  • Pyrolysis conditions: 750 °C following GC readiness trigger
  • GC temperature program: initial 40 °C (2 min), ramp 8 °C/min to 300 °C, hold as required

Main Results and Discussion

The 300 °C desorption chromatogram released all DOP without polymer degradation, demonstrating efficient plasticizer extraction. The subsequent 750 °C pyrolysis run produced a pyrogram featuring HCl, benzene, toluene, naphthalene (PVC fragments), methyl methacrylate (PMMA monomer) and styrene (polystyrene fragment). These signals confirm clear separation of additives and polymer decomposition products, facilitating unambiguous identification.

Benefits and Practical Applications

  • Solvent-free workflow eliminates chemical waste and disposal concerns
  • Automated two-step sampling accelerates analysis and reduces hands-on time
  • Comprehensive profiling of plasticizers and polymer fragments supports quality control, material authentication and failure analysis

Future Trends and Applications

Emerging developments may integrate high-resolution mass spectrometry for detailed structural elucidation, implement real-time data processing with AI-driven deconvolution, and extend thermal sampling to complex composites, multilayer materials and environmental microplastics.

Conclusion

The presented two-step thermal sampling approach delivers a fast, automated and solvent-free solution for complete polymer analysis. By separating volatile additives from polymer fragments, it enhances analytical efficiency and accuracy for research, industrial QC and regulatory applications.

Reference

  • T. P. Wampler, Introduction to Pyrolysis-Capillary Gas Chromatography, J. Chromatography A, 842 (1999) 207.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Coating Analysis on Pharmaceutical Tablet Forms
Coating Analysis on Pharmaceutical Tablet Forms
|Agilent Technologies|Applications
#68 olutions APPLICATIONS INFORMATION USING ADVANCED SAMPLE HANDLING TECHNOLOGY Coating Analysis on Pharmaceutical Tablet Forms A large number of solid pharmaceutical dosage forms in the marketplace are coated. The reason for the coating varies from bioavailability control, stability considerations of…
Key words
coating, coatingdispersal, dispersalbundanc, bundancpyrolyzed, pyrolyzedpharmaceutical, pharmaceuticaladhesion, adhesionolutions, olutionsplasticizer, plasticizerhelp, helpdosage, dosageconcerning, concerningtablets, tabletsapplications, applicationsanalyzed, analyzedtablet
Tri-Step Analysis of Food Packaging
Tri-Step Analysis of Food Packaging
|CDS Analytical|Applications
#146a Tri-Step Analysis of Food Packaging Application Note Packaging Thermal sampling provides a simple way to analyze products like food packaging without extractions or complicated sample preparation. In this example, a 1 mm punch of the wrapping used for a…
Key words
ipdi, ipdipackaging, packagingpolyurethane, polyurethaneheptene, heptenediisocyanate, diisocyanatetraces, tracespyrolyzed, pyrolyzedcereal, cerealprinting, printingfood, foodpet, petregenerated, regeneratedareratio, areratioeanbeen, eanbeenesand
Pyrolysis-MSD for Rapid Polymer Analysis
Pyrolysis-MSD for Rapid Polymer Analysis
|Agilent Technologies|Applications
#76 olutions APPLICATIONS INFORMATION USING ADVANCED SAMPLE HANDLING TECHNOLOGY Pyrolysis-MSD for Rapid Polymer Analysis Although most analytical pyrolysis performed on polymers involves gas chromatography, a quick analysis may be achieved by connecting the Pyroprobe directly to the MSD using a…
Key words
pyrolysis, pyrolysispyroprobe, pyroprobepyrolysate, pyrolysatemsd, msdabundance, abundancecontacting, contactingolutions, olutionsdirectly, directlyinsertion, insertioninterfaced, interfacedapplications, applicationsprobes, probesmay, mayearlier, earlierperformed
Biochemical Analysis by Py-GC/MS
Biochemical Analysis by Py-GC/MS
|Agilent Technologies|Applications
#74 olutions APPLICATIONS INFORMATION USING ADVANCED SAMPLE HANDLING TECHNOLOGY Biochemical Analysis by Py-GC/MS Biosystem components, such as DNA, do not normally lend themselves to thermal analysis. Matrix complexity as well as lack of volatility would seemingly rule out pyrolysis-GC/MS as…
Key words
biosystem, biosystemdeoxyribose, deoxyribosethermally, thermallypyrolysis, pyrolysisbundanc, bundancseemingly, seeminglylikewise, likewiseime, imecontacting, contactingolutions, olutionsapplied, appliedrule, rulebiochemical, biochemicaljournal, journalapplications
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike