Analysis of Volatile Organic Compounds in the Environment Using the Restore Function of TD-GC/MS
Applications | 2019 | ShimadzuInstrumentation
Volatile organic compounds (VOCs) represent a major class of air pollutants across industrial, urban, and indoor environments. Accurate VOC measurement helps assess air quality and regulatory compliance. Thermal desorption–gas chromatography/mass spectrometry (TD-GC/MS) following EPA TO-17 method is standard for VOC analysis. However, lengthy sampling and risk of analysis failure have led to the need for reliable re-analysis (restore) functionality.
This study introduces a novel TD system (TD-30R) with enhanced sample line temperature control to improve the restore capability for VOCs spanning low to high boiling points. Key goals included identifying optimal temperatures and flow conditions for trap tubes, sample tubes, and transfer interfaces to maximize sensitivity and recovery during restoration.
VOC standards and real indoor air samples were analyzed using a Shimadzu GCMS-QP2020 coupled with the TD-30R unit. Experimental variables included:
Recovery and carry-over were evaluated for low (e.g. chloroethene) and high boiling VOCs (e.g. 1,2,4-trichlorobenzene). Calibration covered 1–50 µg, with internal standard toluene-d8.
Lowering trap cooling to −25 °C greatly improved capture of compounds below 0 °C. A 250 °C trap desorption also sufficed for high-boiling analytes. Optimized sample tube desorption reduced carry-over from >5% to <1%. Adjusting joint/valve/transfer temperatures enhanced low-boiler restoration. Method reproducibility was excellent (RSD <10%), and calibration linearity exceeded 0.998. Real indoor air measurements showed comparable initial and restored quantitation, with restoration ratios close to unity.
Future developments may integrate continuous real-time monitoring, expand sorbent options for polar compounds, and automate workflows for high-throughput environmental surveillance.
The TD-30R system’s optimized temperature control at each desorption and trapping stage achieves high-sensitivity, high-recovery VOC analysis across low to high boiling points. Its robust restore function not only mitigates risk of analysis failure but also enhances laboratory efficiency and data reliability.
GC/MSD, Thermal desorption, GC/SQ
IndustriesEnvironmental
ManufacturerShimadzu
Summary
Significance of the Topic
Volatile organic compounds (VOCs) represent a major class of air pollutants across industrial, urban, and indoor environments. Accurate VOC measurement helps assess air quality and regulatory compliance. Thermal desorption–gas chromatography/mass spectrometry (TD-GC/MS) following EPA TO-17 method is standard for VOC analysis. However, lengthy sampling and risk of analysis failure have led to the need for reliable re-analysis (restore) functionality.
Objectives and Study Overview
This study introduces a novel TD system (TD-30R) with enhanced sample line temperature control to improve the restore capability for VOCs spanning low to high boiling points. Key goals included identifying optimal temperatures and flow conditions for trap tubes, sample tubes, and transfer interfaces to maximize sensitivity and recovery during restoration.
Methodology and Instrumentation
VOC standards and real indoor air samples were analyzed using a Shimadzu GCMS-QP2020 coupled with the TD-30R unit. Experimental variables included:
- Trap tube cooling: 0 °C, −20 °C, −25 °C
- Trap desorption: 250 °C vs. 325 °C
- Sample tube desorption: 250 °C for 10 min at 70 mL/min
- Joint, valve, transfer line temperatures: 75 °C, 185 °C, 220 °C
Recovery and carry-over were evaluated for low (e.g. chloroethene) and high boiling VOCs (e.g. 1,2,4-trichlorobenzene). Calibration covered 1–50 µg, with internal standard toluene-d8.
Main Results and Discussion
Lowering trap cooling to −25 °C greatly improved capture of compounds below 0 °C. A 250 °C trap desorption also sufficed for high-boiling analytes. Optimized sample tube desorption reduced carry-over from >5% to <1%. Adjusting joint/valve/transfer temperatures enhanced low-boiler restoration. Method reproducibility was excellent (RSD <10%), and calibration linearity exceeded 0.998. Real indoor air measurements showed comparable initial and restored quantitation, with restoration ratios close to unity.
Benefits and Practical Applications
- Reduces risk of sample loss by enabling reliable re-analysis
- Simultaneously measures VOCs across a wide volatility range
- Delivers high sensitivity, accuracy, and reproducibility
- Applicable in environmental monitoring of industrial and indoor settings
Future Trends and Applications
Future developments may integrate continuous real-time monitoring, expand sorbent options for polar compounds, and automate workflows for high-throughput environmental surveillance.
Conclusion
The TD-30R system’s optimized temperature control at each desorption and trapping stage achieves high-sensitivity, high-recovery VOC analysis across low to high boiling points. Its robust restore function not only mitigates risk of analysis failure but also enhances laboratory efficiency and data reliability.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Shimadzu Thermal Desorption Systems TD-30 Series
2019|Shimadzu|Brochures and specifications
C146-E349B Thermal Desorption Systems TD-30 Series TD-30 T h e r m a l D e s o r p t i o n S y s t e m Revolutionary Thermal Desorption System Provides Excellent Processing Ability and Reliability…
Key words
function, functiondesorption, desorptionretrapping, retrappinganalysis, analysistube, tubepretreatment, pretreatmentthermal, thermaltrap, trapexpandability, expandabilityoverlap, overlapdichlorobenzene, dichlorobenzenehedging, hedginganis, anisboiling, boilingfragrant
Shimadzu Thermal Desorption Systems TD-30 Series
2023|Shimadzu|Brochures and specifications
C146-E349D Thermal Desorption Systems TD-30 Series TD-30 T h e r m a l D e s o r p t i o n S y s t e m Revolutionary Thermal Desorption System Provides Excellent Processing Ability and Reliability…
Key words
function, functionretrapping, retrappingtube, tubeanalysis, analysispretreatment, pretreatmentdesorption, desorptiontrap, trapexpandability, expandabilitydichlorobenzene, dichlorobenzeneoverlap, overlaphedging, hedginganis, anisboiling, boilingbarcode, barcodefragrant
Extractable Study of Pharmaceutical Packaging and Delivery System Used for Ophthalmic Drug Product
2024|Shimadzu|Applications
Gas Chromatograph Mass Spectrometer GCMS-TQ™8040 NX, AOC™ -20i and TD-30R Extractable Study of Pharmaceutical Packaging and Delivery System Used for Ophthalmic Drug Product Application News Hemant Kesarkar, Sanket Chiplunkar, Prashant Hase, Durvesh Sawant, Aseem Wagle, Rahul Dwivedi, Dheeraj Handique, Pratap…
Key words
ccp, ccpphthalate, phthalateextractables, extractablesblq, blqbenzene, benzenefluoranthene, fluorantheneaet, aettetratriacontane, tetratriacontaneextractable, extractableppb, ppbbenzo, benzoreflux, refluxnews, newssct, sctheneicosane
Shimadzu Thermal Desorption System TD-20
2013|Shimadzu|Brochures and specifications
C146-E096B Thermal Desorption System TD-20 T h e rm a l D eso rp tio n Thermal desorption systems trap target substances carried in mobile phase by adsorption in a sample tube packed with adsorbent. Then, by thermal desorption, the…
Key words
desorption, desorptiontube, tubetrap, trapthermal, thermalcarrier, carriersample, samplecoffee, coffeesplit, splitflow, floweso, esoheating, heatingcontroller, controllergases, gasesreplacing, replacingcapillary