Analysis of PCBs in environmental matrices using triple quadrupole GC-MS/MS
Posters | 2022 | Thermo Fisher Scientific | ASMSInstrumentation
Polychlorinated biphenyls (PCBs) are highly persistent pollutants that accumulate in environmental and biological systems. Their chemical stability and toxicity pose significant risks to ecosystems and human health. Sensitive, selective methods are essential to monitor trace levels of PCBs in water, soil, and biotic samples for regulatory compliance and risk assessment.
This work evaluated a gas chromatography triple quadrupole mass spectrometry method for quantifying PCB congeners in environmental matrices. Key performance metrics included calibration linearity, instrument detection limits, and system robustness under routine operating conditions.
Samples were extracted by separatory funnel for aqueous matrices and Soxhlet extraction for solid and semi solid samples, followed by acid and base silica gel cleanup and alumina treatment. Calibration standards covered 0.10 to 2000 ng/mL using EPA Method 1668C native and C13 labeled PCB congener kits. Target analytes were monitored by timed selected reaction monitoring with two transitions per congener and quantified by isotope dilution.
The method achieved chromatographic separation of critical PCB pairs within a 21 minute run. Calibration curves exhibited coefficients of determination above 0.990 and average response factor RSD below 20 percent across 0.10 to 2000 ng/mL. Instrument detection limits ranged from 3 to 19 fg on column, corresponding to sub pg/L levels in water and sub ng/kg levels in soil. Robustness testing with 100 consecutive matrix injections showed stable peak areas and ion ratios within 15 percent without the need for maintenance.
Advances may include coupling with high resolution accurate mass detection for non targeted screening, miniaturized sampling for on site analysis, and further automation of sample preparation and data processing. Integration with large scale monitoring networks will enhance environmental risk assessment.
Triple quadrupole GC MS MS using the TSQ 9610 platform provides a reliable, sensitive, and robust solution for quantifying PCB congeners in environmental samples. Its performance characteristics support routine monitoring and regulatory compliance.
GC/MSD, GC/MS/MS, GC/QQQ
IndustriesEnvironmental
ManufacturerThermo Fisher Scientific
Summary
Significance of the Topic
Polychlorinated biphenyls (PCBs) are highly persistent pollutants that accumulate in environmental and biological systems. Their chemical stability and toxicity pose significant risks to ecosystems and human health. Sensitive, selective methods are essential to monitor trace levels of PCBs in water, soil, and biotic samples for regulatory compliance and risk assessment.
Objectives and Study Overview
This work evaluated a gas chromatography triple quadrupole mass spectrometry method for quantifying PCB congeners in environmental matrices. Key performance metrics included calibration linearity, instrument detection limits, and system robustness under routine operating conditions.
Methodology
Samples were extracted by separatory funnel for aqueous matrices and Soxhlet extraction for solid and semi solid samples, followed by acid and base silica gel cleanup and alumina treatment. Calibration standards covered 0.10 to 2000 ng/mL using EPA Method 1668C native and C13 labeled PCB congener kits. Target analytes were monitored by timed selected reaction monitoring with two transitions per congener and quantified by isotope dilution.
Instrumentation Used
- Thermo Scientific TSQ 9610 triple quadrupole GC MS MS with NeverVent AEI ion source
- Thermo Scientific TRACE 1610 gas chromatograph with TR PCB column
- Thermo Scientific AI AS 1610 liquid autosampler
- Thermo Scientific Chromeleon CDS software version 7.3
Main Results and Discussion
The method achieved chromatographic separation of critical PCB pairs within a 21 minute run. Calibration curves exhibited coefficients of determination above 0.990 and average response factor RSD below 20 percent across 0.10 to 2000 ng/mL. Instrument detection limits ranged from 3 to 19 fg on column, corresponding to sub pg/L levels in water and sub ng/kg levels in soil. Robustness testing with 100 consecutive matrix injections showed stable peak areas and ion ratios within 15 percent without the need for maintenance.
Benefits and Practical Applications
- High selectivity reduces interference from complex matrices
- Extended linear range supports both trace and higher concentration analysis
- Low detection limits enable compliance monitoring at regulatory thresholds
- Robust performance under high sample throughput conditions
Future Trends and Potential Applications
Advances may include coupling with high resolution accurate mass detection for non targeted screening, miniaturized sampling for on site analysis, and further automation of sample preparation and data processing. Integration with large scale monitoring networks will enhance environmental risk assessment.
Conclusion
Triple quadrupole GC MS MS using the TSQ 9610 platform provides a reliable, sensitive, and robust solution for quantifying PCB congeners in environmental samples. Its performance characteristics support routine monitoring and regulatory compliance.
Reference
- United States Environmental Protection Agency EPA Method 1668C Chlorinated Biphenyl Congeners in Water Soil Sediment Biosolids and Tissue by HRGC HRMS April 2010
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Reproducible trace analysis of PCBs in environmental matrices using triple quadrupole GC-MS/MS
2022|Thermo Fisher Scientific|Applications
Application note | 000561 Mass Spectrometry Reproducible trace analysis of PCBs in environmental matrices using triple quadrupole GC-MS/MS Authors Goal Kjell Hope , Giulia Riccardino , Adam The aim of this application note is to demonstrate the performance of the…
Key words
min, minpcbs, pcbsidls, idlsaei, aeiinjection, injectiontime, timenevervent, neverventallowed, allowedwash, washcounts, countscalculated, calculatedcompound, compoundensured, ensuredsmarttune, smarttuneavcf
Low-level consistent analysis of PBDEs in environmental and food matrices using triple quadrupole GC-MS/MS
2021|Thermo Fisher Scientific|Applications
Application note | 000493 Environmental Low-level consistent analysis of PBDEs in environmental and food matrices using triple quadrupole GC-MS/MS Authors Goal Kjell Hope , Giulia Riccardino , The aim of this application note is to demonstrate the performance of the…
Key words
pbdes, pbdesntr, ntrcalculated, calculatedation, ationpbde, pbdecon, conmin, minloq, loqinjection, injectiontracegold, tracegoldidl, idlensured, ensuredmea, meaoret, oreticonnect
NEMC: Reproducible trace analysis of PCBs in environmental matrices using triple quadrupole GC-MS/MS
2022|Thermo Fisher Scientific|Presentations
Reproducible trace analysis of PCBs in environmental matrices using triple quadrupole GC-MS/MS Andy Fornadel, PhD Product Marketing Manager – Americas 4th August 2022 The world leader in serving science 1 [email protected] | 4-August-2022 Content 2 1 Introduction to analysis of…
Key words
pcbs, pcbspcb, pcbclean, cleanpersistence, persistenceabsolute, absolutecompound, compoundcongeners, congenersmin, minalumina, aluminarsd, rsdcounts, countstoxicity, toxicitytenacity, tenacitysrm, srmgel
Fast, ultra-sensitive analysis of PBDEs in food using advanced GC-MS/MS technology
2018|Thermo Fisher Scientific|Applications
APPLICATION NOTE 10674 Fast, ultra-sensitive analysis of PBDEs in food using advanced GC-MS/MS technology Authors Aaron Lamb1, Cristian Cojocariu1, and Sean Panton2 Thermo Fisher Scientific, Runcorn, UK 1 2 Fera Science, Ltd., York, UK Keywords Food safety, polybrominated diphenyl ethers…
Key words
pbde, pbdedeviation, deviationhrms, hrmsaei, aeipbdes, pbdescounts, countsratio, ratioion, ionpenta, pentatetra, tetrabde, bdehexa, hexapass, passcongeners, congenershepta