Ask the Agilent GC and GC/MS Experts
Presentations | 2022 | Agilent TechnologiesInstrumentation
GC, GC/MSD
IndustriesManufacturerAgilent Technologies
Summary
Importance of the Topic
- Gas chromatography (GC) and GC–mass spectrometry (GC/MS) are critical tools for trace analysis in environmental, pharmaceutical, food, and industrial laboratories.
- Column lifetime, phase stability, and system integrity directly affect data quality, reproducibility, and laboratory uptime.
- Identifying common sources of retention shifts, peak distortion, and leaks allows proactive maintenance and optimized workflows.
Study Objectives and Overview
- Compile expert answers to frequently asked questions from GC and GC/MS users.
- Evaluate factors influencing column lifespan under thermal stress and sample loading.
- Demonstrate leak detection strategies and the impact of gas contaminants on peak shape.
- Present approaches to improve injection performance, analysis speed, and instrument maintenance planning.
Methodology and Instrumentation
- Columns: fused-silica capillaries with various stationary phases (DB-5ht, DB-5MS Ultra Inert, VF-624ms) and dimensions (lengths 10–60 m, IDs 0.18–0.53 mm, film thicknesses 0.1–5 µm).
- Thermal stress tests at temperatures up to 400 °C to monitor phase degradation and retention time shifts using standard mixtures (BTEX, hydrocarbons, phenols).
- Leak detection: Agilent CrossLab CS handheld leak detector with interchangeable cartridges (ADM flow meter) and electronic duster profiling via manual tune in MSD.
- Injection techniques: standard splitless, pulsed splitless optimization (pressure pulse, pulse time), backflushing to remove high-boiler components.
- Carrier gases: comparisons between helium and hydrogen for speed, efficiency, and MS compatibility; hydrogen generator and gas clean filters.
- MS source maintenance workflow: QC standards, tune evaluation, source cleaning intervals, and criteria for cleaning or tuning.
Main Results and Discussion
- Unused column longevity is virtually unlimited; thermal aging at 280 °C for 50 h or 400 °C for 120 h shows retention shifts indicating phase degradation.
- Competitor fused-silica columns exhibited visible damage and retention changes after prolonged exposure to 400 °C, whereas Agilent DB-5ht maintained performance.
- Oxygen contamination (<0.3 ppm to 10 ppm) in hydrogen carrier gas progressively worsens peak asymmetry for acid analytes, highlighting the need for gas purification.
- Leaks reduce peak response, elevate background, impair electron multiplier function, and shorten component lifetimes; systematic leak checking with CrossLab CS and electronic duster techniques enables rapid localization.
- Peak fronting indicates column overload; tailing arises from active sites in liners, septa, seals, or non-deactivated surfaces; Ultra Inert deactivation and glass frit liners improve peak shape for acids and bases.
- Pulsed splitless injections can double signal intensity in trace-level pesticide analyses by minimizing sample dispersion in the inlet.
- Speed optimization depends on carrier gas type and velocity, column ID/length/film thickness, and temperature programming; hydrogen provides higher linear velocities at lower temperatures.
- Hydrogen GC/MS offers faster runs and higher efficiency columns but requires careful tuning, safety measures, and avoidance of chlorinated or nitro compounds prone to hydrogenation.
- Backflush functionality removes late-eluting high-boiler contaminants in minutes, preventing extended high-temperature bakeouts.
Benefits and Practical Applications
- Extended column life and optimized conditioning reduce consumable costs and instrument downtime.
- Improved peak shapes and retention stability enhance quantitative accuracy in QA/QC and trace analysis.
- Effective leak management preserves detector performance and extends MS source and filament lifetimes.
- Advanced injection strategies and faster carrier gases increase sample throughput and laboratory efficiency.
- Backflush integration protects columns from contamination and simplifies method development for complex matrices.
Future Trends and Potential Uses
- Adoption of high-efficiency microbore and short-column formats paired with hydrogen to further decrease runtimes and solvent usage.
- Integration of predictive maintenance algorithms and AI-driven leak detection for self-diagnosing GC/MS systems.
- Development of novel deactivated liner materials and ultra inert seals to accommodate challenging analytes like acids, bases, and reactive volatiles.
- Expansion of pulsed and pressure-assist injection modes for automated optimization in multi-analyte methods.
- Continued evolution of gas purification technologies to support hydrogen carrier gas adoption in regulated environments.
Conclusion
- A thorough understanding of column conditioning, phase stability, and system integrity underpins reliable GC and GC/MS performance.
- Regular maintenance guided by expert-recommended QC checks, leak detection, and tune evaluations can prevent unexpected failures.
- Innovations in columns, injection techniques, and carrier gas management deliver faster, more sensitive analyses while reducing operational costs.
- Proactive application of these best practices ensures consistent data quality and maximizes instrument uptime in demanding analytical laboratories.
References
- Agilent Application Note 5991-9035EN: Thermal Stability and Retention Time Shifting.
- Agilent Publication 5994-1013EN: High-Temperature Column Performance.
- Agilent Application Note 5991-4215EN: GC Column Selector.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Ask the Agilent Experts: GC and GC/MS
2024|Agilent Technologies|Presentations
Ask the Agilent Experts GC and GC/MS scientists discuss and answer your most frequently asked questions Vanessa Abercrombie Mark Sinnott DE29544233 How Long Should My Liner Last? DE29544233 That’s a Tough Question... What is the sample preparation? What is the…
Key words
wool, wooltune, tunebleed, bleedwhat, whatsepta, septafrit, fritgas, gasleaks, leaksseptum, septumhow, howagilent, agilentclean, cleanbarrier, barriereval, evalhydrogen
GC and GC/MS Frequently Asked Questions
2024|Agilent Technologies|Guides
GC and GC/MS Frequently Asked Questions A History of GC and GC/MS Chromatography was invented by the Russian botanist, Mikhail Tswett, in 1903 to separate pigments in plants. It comes from the Greek words chroma, meaning color, and graphein, meaning…
Key words
troubleshooting, troubleshootingknowledge, knowledgecontents, contentsmaintenance, maintenancegeneral, generaloptimization, optimizationtable, tabledevelopment, developmenthow, howmethod, methodagilent, agilentyour, yourwhat, whatlearn, learnyou
Mass Spectrometer Optimization
2020|Agilent Technologies|Presentations
Mass Spectrometer Optimization Angela Smith Henry, Ph.D. Applications Chemist CSD Supplies Division 1 January 20, 2020 Mass Spectrometer Optimization Agilent Restricted Agilent Inert Flow Solution Agilent UltiMetal Plus – TCD, FPD, NPD/FID jets Agilent UltiMetal Plus inlet weldment, shell and…
Key words
restricted, restrictedoptimization, optimizationspectrometer, spectrometeragilent, agilentmass, massleak, leakduster, dustertune, tunecheck, checknut, nuttightening, tighteningtransfer, transferwhat, whatyou, youuse
Sealing the Deal: Strategies for Leak-Free GC
2024|Agilent Technologies|Presentations
Sealing the Deal: Strategies for Leak-Free GC Alexander Ucci Application Engineer October 15, 2024 1 October 15, 2024 Sealing the Deal DE-001646 What Will We Cover Today? • Signs and symptoms of a leak • Instrument diagnostics • Finding the…
Key words
deal, dealsealing, sealingleak, leaknut, nuttightening, tighteningleaks, leakscolumn, columnnuts, nutsagilent, agilentferrule, ferruleferrules, ferrulesyour, yourself, selfwinged, wingedbleed