GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

How Column Inertness Improves the Chromatography of Basic Compounds

Applications | 2010 | RestekInstrumentation
GC, GC columns, Consumables
Industries
Pharma & Biopharma
Manufacturer
Agilent Technologies, Restek

Summary

Importance of the Topic


The analysis of basic compounds by gas chromatography often suffers from peak tailing, poor sensitivity, and variable quantitation due to interactions between analytes and active sites on the column or inlet surfaces. Improving column inertness is essential for achieving accurate, reproducible, and sensitive measurements in pharmaceutical and environmental applications where amines and other basic analytes are prevalent.

Study Objectives and Overview


This work investigates how different column deactivation strategies affect peak shape, sensitivity, and linearity when analyzing basic compounds. By comparing a highly inert Rxi®-624Sil MS column with a conventional ZB-624 column, the study demonstrates the practical benefits of balanced deactivation for both basic and acidic compounds.

Methodology and Instrumentation


Analyses were performed by capillary GC using an Agilent/HP 6890 system equipped with an FID.
  • Columns compared: Rxi®-624Sil MS (6% cyanopropylphenyl/94% dimethylsiloxane) versus ZB-624
  • Column dimensions: 30 m × 0.32 mm ID, 1.8 µm film
  • Injection: 1 µL split (20:1), injector at 250 °C, 5 mm gooseneck liner with wool
  • Carrier gas: He at 37 cm/s linear velocity
  • Oven program: 50 °C (1 min) to 200 °C at 20 °C/min (5 min)

Main Results and Discussion


Peak tailing increased dramatically on the ZB-624 column at low on-column masses, leading to poor integration below 25 ng of isopropylamine and retention time shifts. In contrast, the Rxi®-624Sil MS column delivered symmetric peaks across 5–100 ng with an R2 of 0.99996. Tailing resulting from silanol interactions was confirmed by mass-dependence tests and by comparing primary, secondary, and tertiary amines. The balanced deactivation in the Rxi® phase minimized adsorption for both acidic and basic analytes, whereas specialized deactivations often favor one class over another.

Benefits and Practical Applications


Choosing a column with balanced inertness offers several advantages:
  • Enhanced sensitivity and lower detection limits without increasing sample load
  • Improved peak symmetry for accurate integration and calibration
  • Greater method linearity and reproducibility
  • Extended column lifetime due to reduced surface reactivity
  • Reduced method development time by avoiding column swaps when analyzing mixtures of varied polarity

Future Trends and Applications


Advances in deactivation chemistries are expected to yield even more universal column phases that combine ultra-low bleed with high inertness for diverse compound classes. Integration with automated inlet liners and sample-preparation systems will further streamline analysis workflows. Additionally, expanding inert column technology into two-dimensional GC and coupling with mass spectrometry will enhance trace-level analyses in complex matrices.

Conclusion


Column inertness is a critical factor in achieving reliable GC analysis of basic compounds. Balanced deactivation, as exemplified by the Rxi® technology, eliminates excessive interaction with silanol sites, delivering symmetric peaks, high sensitivity, and robust quantitation without the need for mobile-phase additives. Selecting a highly inert column simplifies method development and improves overall data quality in pharmaceutical and analytical laboratories.

Instrumentation


  • GC system: Agilent/HP 6890 GC
  • Detector: Flame Ionization Detector at 250 °C
  • Columns: Rxi®-624Sil MS and ZB-624, both 30 m × 0.32 mm ID, 1.8 µm
  • Liner: 5 mm gooseneck with wool

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Guide to GC Column Selection and Optimizing Separations
Guide to GC Column Selection and Optimizing Separations WHICH COLUMN DO I NEED? • Learn how to choose the right column the first time. • Optimize separations for the best balance of resolution and speed. • Troubleshoot quickly and effectively…
Key words
restek, restekfilm, filmcolumn, columnrtx, rtxcauses, causesstationary, stationarycharacteristics, characteristicsloading, loadingproprietary, proprietaryinnovation, innovationthickness, thicknessphase, phasecapacity, capacityapplications, applicationssilarylene
Zebron PLUS GC Columns - Improved Inertness, Better Peak Shape, Low Bleed, and Aqueous Stability
Zebron PLUS GC Columns Improved Inertness, Better Peak Shape, Low Bleed, and Aqueous Stability ™ G C C o lu mns ZB-1plus ZB-5plus ZB-5MSplus ZB-624plus ZB-WAXplus ™ ™ ™ ™ ™ PLUS www.phenomenex.com/GCplus Exceptionally Versatile GC phases PLUS Exceptional Inertness…
Key words
zebron, zebronphenomenex, phenomenexbleed, bleedfilter, filterclick, clickwaxplus, waxplusindicator, indicatorgas, gasmin, mintrap, trapsolvents, solventsplus, pluselectronic, electronicbetter, betterwall
Guide to GC Column Selection and Optimizing Separations
Guide to GC Column Selection and Optimizing Separations • Learn how to choose the right column the first time. • Optimize separations for the best balance of resolution and speed. • Troubleshoot quickly and effectively based on chromatographic symptoms. Length…
Key words
restek, restekcolumn, columnsilarylene, silaryleneproprietary, proprietaryrtx, rtxfilm, filmcauses, causesstationary, stationaryinnovation, innovationcharacteristics, characteristicsloading, loadingphase, phasethickness, thicknessinner, innerstabilwax
Guide to GC Column Selection and Optimizing Separations
Guide to GC Column Selection and Optimizing Separations • Learn how to choose the right column the first time. • Optimize separations for the best balance of resolution and speed. • Troubleshoot quickly and effectively based on chromatographic symptoms. Length…
Key words
restek, restekcolumn, columnsilarylene, silaryleneproprietary, proprietaryfilm, filmrtx, rtxcauses, causesstationary, stationaryinnovation, innovationloading, loadingcharacteristics, characteristicsthickness, thicknessphase, phaseinner, innerstabilwax
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike