Analysis of Trace Hydrocarbon Impurities in 1,3-Butadiene Using Optimized Rt®-Alumina BOND/MAPD PLOT Columns

Applications | 2011 | RestekInstrumentation
GC, GC columns, Consumables
Industries
Energy & Chemicals
Manufacturer
Agilent Technologies, Restek

Summary

Significance of the Topic


Detecting trace hydrocarbon impurities in 1,3-butadiene is essential for ensuring consistent polymerization performance and high-quality synthetic rubber. Low-level polar contaminants such as methyl acetylene and propadiene can poison catalysts, while heavier by-products like 4-vinylcyclohexene form during storage and require accurate quantification to meet quality specifications.

Objectives and Study Overview


This study evaluates the capability of Rt®-Alumina BOND/MAPD PLOT columns to separate and quantify both light polar C4 isomers and heavier impurities in a single chromatographic run. By applying an optimized thermal program across the full operating range of the column, the work aims to streamline impurity profiling of crude and refined 1,3-butadiene samples.

Methodology and Used Instrumentation


Chromatographic separations were performed on an Agilent 5890 GC equipped with a 50 m × 0.53 mm ID × 10 µm Rt®-Alumina BOND/MAPD PLOT column. Key parameters included:
  • Injection: 10 µL split, 200 °C injector temperature, 100 mL/min split flow
  • Carrier gas: Helium at constant 20 psi
  • Detector: FID at 250 °C with N2 makeup gas at 30 mL/min
  • Oven programs: Crude sample ramp from 70 °C (5 min hold) to 200 °C@10 °C/min; refined sample to 250 °C@10 °C/min (5 min hold)
The alumina adsorbent used a specialized MAPD deactivation to improve inertness toward polar hydrocarbons.

Main Results and Discussion


The optimized column and temperature program delivered high resolution and sensitivity for both low molecular weight C4 impurities and higher molecular weight compounds. In crude 1,3-butadiene, all C4 isomers, propadiene, methyl acetylene, and an additional peak identified as 1,2-butadiene were baseline-resolved. In refined samples, extending the oven to 250 °C allowed detection of 4-vinylcyclohexene alongside native impurities. Instrument parameters such as initial hold time and flow rate were found to further influence critical separations.

Benefits and Practical Applications


  • Consolidated analysis of both light and heavy impurities in one run reduces testing time and resource usage.
  • Enhanced column inertness improves reproducibility and sensitivity for polar analytes.
  • Extended temperature stability widens the range of detectable compounds without changing columns.
This approach streamlines quality control workflows in synthetic rubber production and petrochemical operations.

Future Trends and Potential Applications


Expanding column deactivation chemistries and integrating mass spectrometric detection could further improve trace analysis. Advances in high-temperature stationary phases may enable real-time online monitoring of feed streams. Data-driven optimization of GC parameters promises enhanced method robustness and faster throughput.

Conclusion


The Rt®-Alumina BOND/MAPD PLOT column effectively separates and quantifies a comprehensive range of hydrocarbon impurities in 1,3-butadiene within a single chromatographic method. Its improved inertness and extended thermal tolerance enable reliable detection from C4 isomers to 4-vinylcyclohexene, offering significant efficiency gains in purity testing.

References


  • J. de Zeeuw, R. Morehead, T. Vezza, B. Bromps. Chromatographic Behavior of Activated Alumina Adsorbents for the Analysis of Hydrocarbons. American Laboratory, 2011.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Alumina BOND/MAPD PLOT Columns
Alumina BOND/MAPD PLOT Columns
2013|Restek|Brochures and specifications
Restek Refined ® Analyze Trace Polar Hydrocarbons More Accurately and Reliably With Alumina BOND/MAPD PLOT Columns! • Reproducible, predictable responses for reduced calibration frequency. • Exceptional sample loading capacity, which improves resolution and response. • Highest temperature stability—application range extended…
Key words
mapd, mapdalumina, aluminabond, bondsulfinert, sulfinertrestek, restekmxt, mxtstainless, stainlesssteel, steeltreated, treatedacetylene, acetylenekit, kitpeakarea, peakareafilter, filterplot, plotcolumns
Restek’s PLOT Column Family
Restek’s PLOT Column Family
2019|Agilent Technologies|Applications
Restek Refined Restek’s PLOT Column Family — The Benchmark for Performance! • Innovative bonding process minimizes particle release. • More consistent flow means stable retention times. • Outstanding peak symmetry improves impurity analysis. Visit us at www.restek.com/petro www.restek.com Next Generation…
Key words
bond, bondalumina, aluminaplot, plotmxt, mxtcolumns, columnsrestek, restekmapd, mapdporous, porousfused, fusedcfc, cfcsilica, silicacolumn, columngas, gasphases, phasespeaks
Improve Trace Analysis of Acetylene, Propadiene, and Methyl Acetylene Impurities with Higher Capacity Alumina MAPD Columns
000_PIN_GUIDE_11:Layout 1 2 1/12/11 16:32 Page 2 Analytical Instrumentation Improve Trace Analysis of Acetylene, Propadiene, and Methyl Acetylene Impurities with Higher Capacity Alumina MAPD Columns Rick Morehead, Jan Pijpelink, and Jaap de Zeeuw*, Restek Corporation Restek Corporation, 110 Benner Circle,…
Key words
mapd, mapdalumina, aluminaacetylene, acetylenetailing, tailingpropadiene, propadienedeactivation, deactivationcapacity, capacityimpurities, impuritiescolumns, columnsrestek, restekmanufacturer, manufacturerbond, bondusp, uspdeveloper, developerfactor
Improved Gas Chromatograph Method for the Analysis of Trace Hydrocarbon Impurities in 1, 3-Butadiene
Improved Gas Chromatograph Method for the Analysis of Trace Hydrocarbon Impurities in 1, 3-Butadiene Application Note Hydrocarbon processing industry Author Abstract Chunxiao Wang An improved gas chromatographic method based on an Agilent 7890 Series GC for the Agilent Technologies (Shanghai)…
Key words
acetylene, acetylenepropyne, propynebutane, butaneiso, isopentane, pentanedimer, dimerpropadiene, propadienevinyl, vinylbackflushing, backflushinghpli, hpliethyl, ethylvalve, valvepropylene, propylenepropane, propaneimpurities
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike