Rxi®-5Sil MS Assured Performance for Forensic Applications
Brochures and specifications | 2014 | RestekInstrumentation
Forensic toxicology laboratories require highly inert, low-bleed gas chromatography columns to achieve reliable trace-level detection, reproducible results, and minimal instrument downtime. The choice of column chemistry directly impacts sensitivity, data quality, and operational efficiency when analyzing drugs of abuse.
This application note evaluates Rxi-5Sil MS capillary columns—featuring a 5 % phenyl/95 % dimethyl polysiloxane phase with embedded arylene groups—across multiple forensic drug analyses. Goals include demonstrating enhanced inertness, reduced bleed, extended column lifetime, and versatile selectivity compared to conventional 5 % phenyl phases.
Analytes—amphetamines, opiates, cocaine and metabolites, cannabinoids, benzodiazepines—were derivatized using reagents such as HFBA, BSTFA + 1 % TMCS, and propionic anhydride. GC–MS analyses employed a Shimadzu 2010 GC coupled to a QP2010+ quadrupole MS. Key parameters:
Rxi-5Sil MS columns delivered symmetric peak shapes, stable retention times, and high signal-to-noise ratios across derivatized analytes. Under repeated exposure to harsh derivatization reagents (400 injections of HFBA), no increase in bleed or loss of inertness was observed. Selectivity closely matched traditional 5 % phenyl phases while offering superior robustness.
By combining inertness, low bleed, and broad selectivity, Rxi-5Sil MS columns reduce the need for column changes between methods, lower maintenance frequency, and improve confidence in trace-level quantitation. This translates to streamlined workflows in forensic, clinical, workplace, and sports drug testing environments.
Future developments will focus on stationary phases that further minimize bleed and extend chemical resistance to support more sensitive MS detectors. Integration with automated sample preparation and the emergence of superficially porous particle LC columns may offer complementary speed and resolution benefits in multi-platform laboratories.
Rxi-5Sil MS columns present a unified solution for forensic GC–MS analysis, delivering exceptional inertness, low bleed, and reproducible selectivity. Their robustness under harsh conditions and extended lifetime make them ideal for high-throughput, trace-level drug testing applications.
GC, GC/MSD, GC/SQ, GC columns, Consumables
IndustriesForensics
ManufacturerAgilent Technologies, Shimadzu, Restek
Summary
Importance of the Topic
Forensic toxicology laboratories require highly inert, low-bleed gas chromatography columns to achieve reliable trace-level detection, reproducible results, and minimal instrument downtime. The choice of column chemistry directly impacts sensitivity, data quality, and operational efficiency when analyzing drugs of abuse.
Objectives and Study Overview
This application note evaluates Rxi-5Sil MS capillary columns—featuring a 5 % phenyl/95 % dimethyl polysiloxane phase with embedded arylene groups—across multiple forensic drug analyses. Goals include demonstrating enhanced inertness, reduced bleed, extended column lifetime, and versatile selectivity compared to conventional 5 % phenyl phases.
Methodology and Instrumentation
Analytes—amphetamines, opiates, cocaine and metabolites, cannabinoids, benzodiazepines—were derivatized using reagents such as HFBA, BSTFA + 1 % TMCS, and propionic anhydride. GC–MS analyses employed a Shimadzu 2010 GC coupled to a QP2010+ quadrupole MS. Key parameters:
- Column: 30 m × 0.25 mm ID × 0.25 μm Rxi-5Sil MS
- Carrier Gas: Helium at constant linear velocity
- Injection: Splitless (hold time) with premium taper liners
- Oven Programs: Optimized temperature ramps for each compound class
- Detection: Full-scan and SIM modes, electron ionization at 70 eV
Main Results and Discussion
Rxi-5Sil MS columns delivered symmetric peak shapes, stable retention times, and high signal-to-noise ratios across derivatized analytes. Under repeated exposure to harsh derivatization reagents (400 injections of HFBA), no increase in bleed or loss of inertness was observed. Selectivity closely matched traditional 5 % phenyl phases while offering superior robustness.
- Amphetamines: Sharp, baseline-resolved peaks in under 10 min
- Cocaine Series: Clear separation of ecgonine derivatives and metabolites in SIM mode
- Benzodiazepines: Complete resolution of seven structural isomers in <8 min
- Opiates & Cannabinoids: High sensitivity and low bleed enhance detection limits
Benefits and Practical Applications
By combining inertness, low bleed, and broad selectivity, Rxi-5Sil MS columns reduce the need for column changes between methods, lower maintenance frequency, and improve confidence in trace-level quantitation. This translates to streamlined workflows in forensic, clinical, workplace, and sports drug testing environments.
Future Trends and Opportunities
Future developments will focus on stationary phases that further minimize bleed and extend chemical resistance to support more sensitive MS detectors. Integration with automated sample preparation and the emergence of superficially porous particle LC columns may offer complementary speed and resolution benefits in multi-platform laboratories.
Conclusion
Rxi-5Sil MS columns present a unified solution for forensic GC–MS analysis, delivering exceptional inertness, low bleed, and reproducible selectivity. Their robustness under harsh conditions and extended lifetime make them ideal for high-throughput, trace-level drug testing applications.
Instrumentation
- GC: Shimadzu 2010
- MS: Shimadzu QP2010+
- Column Chemistry: 5 % phenyl/95 % dimethyl polysiloxane with embedded arylene groups
- Inlet Liners: Premium taper and cyclo liners with wool for splitless injection
- Derivatization Reagents: HFBA, BSTFA + 1 % TMCS, propionic anhydride
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Rxi®-5Sil MS Columns
2015|Restek|Brochures and specifications
Rxi®-5Sil MS Columns Rugged, Low-Bleed Performance for Challenging GC-MS Work Part of the Rxi® GC Column Family • Accurate MS data • The right results fast • Maximize instrument uptime 174 201 Time (s) 892 894 896 898 900 902…
Key words
gcs, gcsinlets, inletssplitless, splitlessane, anedibenzodioxins, dibenzodioxinsbenzo, benzoepa, epaequipped, equippedcarcinog, carcinoghxcdfs, hxcdfspecdds, pecddscolumn, columnrestek, restekenic, enicpotentia
GCxGC Columns Your One Source for 2D Gas Chromatography
2012|Agilent Technologies|Technical notes
GCxGC Columns Your One Source for 2D Gas Chromatography Why use GCxGC?.......................................... 2 Meet the Experts.......................................... 4 PAHs in Seafood....................................... 4–5 PCBs and BDEs............................................... 6 Pesticides in Dietary Supplements....... 7 Halogenated POPs in Human Milk....... 8 Aromatics and Nonaromatics in…
Key words
gcxgc, gcxgcrxi, rxipolar, polaristd, istdrestek, restekstabilwax, stabilwaxxlb, xlbnon, noncolumn, columnmarijuana, marijuanaprimary, primarymodulator, modulatordimensional, dimensionalpress, pressgcs
NEW! Rxi®-1301Sil MS GC Columns
2015|Restek|Brochures and specifications
NEW! Rxi®-1301Sil MS GC Columns The Selectivity You Need Without the Bleed Cyano-based 1301 columns are general use GC columns that are well suited for the analysis of solvents across a range of volatilities. The cyano stationary phase provides more…
Key words
gcs, gcsrestek, restekbleed, bleedadaptor, adaptoragilent, agilentinlets, inletsinlet, inletviton, vitoncyano, cyanoliner, linercompetitor, competitorrings, ringssplitless, splitlessrigorous, rigorouspremium
Rxi®-5Sil MS Columns - Exceptionally Inert Columns for GC/MS and Trace-Level Analyses
2011|Restek|Brochures and specifications
3-IN-1 TECHNOLOGY Highest Inertness • Lowest Bleed • Exceptional Reproducibility ® Rxi -5SilColumns MS Exceptionally Inert Columns for GC/MS and Trace-Level Analyses • Low bleed at higher temperatures for better sensitivity and lower detection limits. • Excellent inertness for highest…
Key words
manufacturer, manufacturerexceptionally, exceptionallyrestek, restekbleed, bleeddicyclohexylamine, dicyclohexylaminepatents, patentscolumn, columnbasic, basicinert, inertpyridine, pyridineinertness, inertnessexceptional, exceptionalacidic, acidicguard, guardintegra