Extended Dynamic Range, High Precision Analysis of Polynuclear Aromatic Hydrocarbon Compounds by GC-MS
Posters | 2012 | Thermo Fisher Scientific | PittconInstrumentation
The reliable detection and quantitation of polynuclear aromatic hydrocarbons (PAHs) at trace levels is crucial for assessing environmental contamination and human exposure risks. These persistent organic pollutants, arising from petroleum and combustion processes, pose carcinogenic and ecological hazards. Advanced analytical methods that extend dynamic range while maintaining high precision enable effective monitoring and remediation efforts.
This work describes the development and validation of a gas chromatography–mass spectrometry (GC-MS) method capable of analyzing 16 priority PAH compounds over a concentration range from 10.0 to 0.01 ng/µL. Key goals included achieving linearity across four orders of magnitude, precise quantitation for high molecular weight PAHs, and accurate measurement in soil extract matrices at sub-nanogram levels.
Standards and internal standards were prepared by automated serial dilution from 10.0 to 0.01 ng/µL. Soil samples were extracted and analyzed at 0.01 ng/µL to assess method precision and accuracy in a complex matrix. The GC oven program spanned 250 °C initial hold to 300 °C final hold, with helium carrier flow at 2.0 mL/min. The MS was run in electron ionization mode at 350 °C source temperature, monitoring three characteristic ions per compound.
All 16 PAHs demonstrated excellent linearity with R² values ≥ 0.996 over the full concentration range. Internal standard precision showed RSD ≤ 10% across calibration levels. Analysis of soil extracts at 0.01 ng/µL achieved RSD ≤ 5% for all target analytes. Accuracy was confirmed by measuring measured vs. target concentrations at low, mid, and high levels, with deviations within ±15%. High molecular weight PAHs such as benzo[a]pyrene and indeno[1,2,3-cd]pyrene exhibited consistent quantitation performance.
This method offers:
The ongoing integration of high-resolution mass spectrometry could enhance selectivity for isomeric PAHs and reduce interference. Emerging sample introduction techniques, such as multi-dimensional GC and direct-injection interfaces, may further lower detection limits. Advances in data processing and machine learning are expected to streamline quantitation workflows and automate anomaly detection in large environmental data sets.
The described GC-MS protocol demonstrates reliable extended dynamic range and high-precision analysis of priority PAHs. Its combination of automated calibration, robust instrumentation, and selective SIM acquisition renders it well suited for routine environmental and regulatory monitoring.
GC/MSD, GC/SQ
IndustriesEnvironmental
ManufacturerThermo Fisher Scientific
Summary
Significance of the Topic
The reliable detection and quantitation of polynuclear aromatic hydrocarbons (PAHs) at trace levels is crucial for assessing environmental contamination and human exposure risks. These persistent organic pollutants, arising from petroleum and combustion processes, pose carcinogenic and ecological hazards. Advanced analytical methods that extend dynamic range while maintaining high precision enable effective monitoring and remediation efforts.
Study Objectives and Overview
This work describes the development and validation of a gas chromatography–mass spectrometry (GC-MS) method capable of analyzing 16 priority PAH compounds over a concentration range from 10.0 to 0.01 ng/µL. Key goals included achieving linearity across four orders of magnitude, precise quantitation for high molecular weight PAHs, and accurate measurement in soil extract matrices at sub-nanogram levels.
Instrumentation
- TriPlus RSH autosampler with automated calibration dilution functionality
- TRACE 1310 gas chromatograph with splitless/surge injection port
- ISQ single quadrupole mass spectrometer operated in selected ion monitoring (SIM) mode
- Thermo Scientific TraceGOLD TG-5SilMS analytical column (30 m × 0.25 mm i.d. × 0.50 µm film)
- Thermo Scientific TraceFinder software for data acquisition and processing
Methodology
Standards and internal standards were prepared by automated serial dilution from 10.0 to 0.01 ng/µL. Soil samples were extracted and analyzed at 0.01 ng/µL to assess method precision and accuracy in a complex matrix. The GC oven program spanned 250 °C initial hold to 300 °C final hold, with helium carrier flow at 2.0 mL/min. The MS was run in electron ionization mode at 350 °C source temperature, monitoring three characteristic ions per compound.
Main Results and Discussion
All 16 PAHs demonstrated excellent linearity with R² values ≥ 0.996 over the full concentration range. Internal standard precision showed RSD ≤ 10% across calibration levels. Analysis of soil extracts at 0.01 ng/µL achieved RSD ≤ 5% for all target analytes. Accuracy was confirmed by measuring measured vs. target concentrations at low, mid, and high levels, with deviations within ±15%. High molecular weight PAHs such as benzo[a]pyrene and indeno[1,2,3-cd]pyrene exhibited consistent quantitation performance.
Benefits and Practical Applications
This method offers:
- A broad dynamic range suitable for diverse environmental monitoring scenarios
- High precision at trace levels, essential for regulatory compliance and risk assessment
- Time and resource savings through automated dilution and sample handling
- Robust performance in complex soil matrices
Future Trends and Opportunities
The ongoing integration of high-resolution mass spectrometry could enhance selectivity for isomeric PAHs and reduce interference. Emerging sample introduction techniques, such as multi-dimensional GC and direct-injection interfaces, may further lower detection limits. Advances in data processing and machine learning are expected to streamline quantitation workflows and automate anomaly detection in large environmental data sets.
Conclusion
The described GC-MS protocol demonstrates reliable extended dynamic range and high-precision analysis of priority PAHs. Its combination of automated calibration, robust instrumentation, and selective SIM acquisition renders it well suited for routine environmental and regulatory monitoring.
Reference
- Khan Z.; Troquet J.; Vachelard C. Sample Preparation and Analytical Techniques for Determination of Poly Aromatic Hydrocarbons in Soil. Int. J. Environ. Sci. Tech. 2005, 2, 275–286.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
A fast, robust method for routine determination of polycyclic aromatic hydrocarbons (PAH) in drinking water by single quadrupole GC-MS
2019|Thermo Fisher Scientific|Applications
APPLICATION NOTE 10718 A fast, robust method for routine determination of polycyclic aromatic hydrocarbons (PAH) in drinking water by single quadrupole GC-MS Authors Michela Balsamo1, Giulia Riccardino2, and Cristian Cojocariu2 Laboratorio interno “Francesco Scognamiglio” della società GORI S.p.A gestore del…
Key words
benzo, benzofluoranthene, fluoranthenepyrene, pyrenepah, pahpahs, pahsghi, ghiperylene, peryleneanthracene, anthracenefluoranathene, fluoranathenephenanthrene, phenanthrenechrysene, chrysenebyphenyl, byphenylpicene, piceneanthanthrene, anthanthrenetriphenylene
Analysis of multiple matrices with a single calibration curve for polycyclic aromatic hydrocarbons (PAHs) with the ISQ 7610 GC-MS system following EPA Method 8270E
2021|Thermo Fisher Scientific|Applications
Application note | 000455 Environmental analysis Analysis of multiple matrices with a single calibration curve for polycyclic aromatic hydrocarbons (PAHs) with the ISQ 7610 GC-MS system following EPA Method 8270E Authors Goal Chiara Calaprice , Brian Pike , The aim…
Key words
benzo, benzofluoranthene, fluorantheneanthracene, anthracenepyrene, pyrenethra, thranaphthalene, naphthalenedibenzofuran, dibenzofuranacenaphthene, acenaphthenechrysene, chrysenepahs, pahsphenanthrene, phenanthreneaph, aphbenz, benzacenaphthylene, acenaphthylenefluorene
Determination of polycyclic aromatic hydrocarbons in drinking water at ppt levels by Solid Phase Micro Extraction Arrow coupled with GC-MS
2017|Thermo Fisher Scientific|Applications
APPLICATION NOTE 10558 Determination of polycyclic aromatic hydrocarbons in drinking water at ppt levels by Solid Phase Micro Extraction Arrow coupled with GC-MS Authors Manuela Bergna, Giulia Riccardino Thermo Fisher Scientific, Rodano, Italy Goal To demonstrate a fully automatized method…
Key words
arrow, arrowbenzo, benzospme, spmeextraction, extractionpahs, pahsfiber, fiberfluoranthene, fluorantheneanthracene, anthracenepyrene, pyrenersh, rshisq, isqtriplus, triplusthermo, thermophase, phasescientific
Automated liquid-liquid extraction workflow for direct ultra-trace analysis of pesticides and PAHs in water matrices using GC-MS/MS
2025|Thermo Fisher Scientific|Applications
Application note | 003943 Environmental Automated liquid-liquid extraction workflow for direct ultra-trace analysis of pesticides and PAHs in water matrices using GC-MS/MS Authors Goal Nicholas A. Warner1, Sinéad McDonald2, The goal of this application note is to demonstrate the capabilities…
Key words
hch, hchbenzo, benzofluoranthene, fluoranthenetrans, transcis, cisnonachlor, nonachlorendrin, endrinpyrene, pyreneheptachlor, heptachlorchlordane, chlordaneanthracene, anthracenegamma, gammaendosulfan, endosulfanovex, ovexpermethrin