GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Reverse Flow Split/Splitless Injector: Simplification of ASTM D3606 and D4815 Methods

Posters |  | Thermo Fisher ScientificInstrumentation
GC
Industries
Energy & Chemicals
Manufacturer
Thermo Fisher Scientific

Summary

Importance of the Topic


The accurate quantification of benzene, toluene and oxygenates such as MTBE in gasoline is vital for environmental compliance, health protection and fuel quality control. Regulatory limits on toxic aromatics and the need for consistent octane enhancement drive the demand for robust, high-throughput analytical methods in refinery, QA/QC and research laboratories.

Objectives and Study Overview


This study introduces two modified GC methods based on ASTM D3606 and D4815, employing the Thermo Scientific TRACE 1310 GC with an Instant Connect Split/Splitless Injector featuring a backflush module. The goal is to simplify hardware requirements, maintain or improve repeatability and throughput, and enable seamless method switching on a single platform.

Methodology and Instrumentation


The analytical configuration comprises:
  • Thermo Scientific TRACE 1310 GC equipped with Instant Connect SSL Backflush Injector
  • Capillary precolumns (e.g., TR-WAX or TG-5MS) and analytical columns (e.g., TR-1 or TG-TCEP)
  • Flame Ionization Detector (FID)
  • Automated Autosampler for consistent injections
  • Integrated backflush manifold for rapid module swapping and minimal leak risk

Key procedural steps include split/splitless injection, programmable reverse-flow backflush to remove heavy hydrocarbons, and temperature programming tailored to separate benzene, toluene, MTBE and other oxygenates. Cycle times range from 10 to 16 minutes depending on the method.

Main Results and Discussion


Calibration and linearity:
  • Benzene: 0.06–5 % v/v, R² = 0.9998
  • Toluene: 0.5–20 % v/v, R² = 0.9998
  • MTBE: 0.095–19 % w/w, R² = 0.9993

Precision and system stability:
  • Repeatability (150 injections): benzene RSD 1.21 %, toluene RSD 1.26 %
  • MTBE repeatability (70 injections): RSD 1.85 %
  • Blank runs confirmed effective backflush with no carryover

The reverse-flow injector architecture simplifies cut-point control compared to balanced-pressure valves, protecting columns and detectors by venting high-boilers through the split line.

Benefits and Practical Application of the Method


This approach offers:
  • Reduced capital and maintenance costs versus packed-column, multi-valve systems
  • High sample throughput with cycle times as low as 13 minutes
  • Modular plug-and-play injector and detector options for rapid method changeover
  • Compatibility with existing FID-based laboratories for routine gasoline analysis

Future Trends and Opportunities


Potential developments include:
  • Adaptation of backflush timing for complex renewable fuel blends
  • Coupling with mass spectrometry for enhanced selectivity and identification
  • Automated method optimization using AI-driven software
  • Portable GC modules for field-based fuel testing and environmental monitoring

Conclusion


The TRACE 1310 GC with Instant Connect SSL Backflush Injector delivers a streamlined, reliable alternative to ASTM D3606 and D4815 for benzene, toluene and oxygenate analysis in gasoline. Its combination of precision, throughput and modular flexibility supports diverse laboratory workflows without dedicated packed-column equipment.

References


Pelagatti S., Magni P., Santoro M., Pigozzo F., Caruso A., Phillips E. Reverse Flow Split/Splitless Injector: Simplification of ASTM D3606 and D4815 Methods.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Simplification of the ASTM D3606 Method for the Determination of Benzene and Toluene in Gasoline
Application Note: 52284 Simplification of the ASTM D3606 Method for the Determination of Benzene and Toluene in Gasoline Andrea Caruso, Massimo Santoro, Thermo Fisher Scientific, Milan, Italy Key Words • TRACE 1310 GC • Chromeleon Data System • ASTM D3606…
Key words
benzene, benzenetoluene, toluenemicrovolts, microvoltsgasoline, gasolinessl, sslcontent, contentbackflush, backflushaviation, aviationaverage, averagemotor, motorrepeatability, repeatabilitymodule, moduleafrica, africaminutes, minutescompact
Running ASTM Methods D4815 and D5580 on a Single Agilent 6890N Gas Chromatograph with Nitrogen Carrier Gas
Running ASTM Methods D4815 and D5580 on a Single Agilent 6890N Gas Chromatograph with Nitrogen Carrier Gas Application Petrochemical Author James D. McCurry Agilent Technologies, Inc. 2850 Centerville Rd. Wilmington DE 19808 USA Abstract An Agilent 6890N GC system is…
Key words
benzene, benzeneepc, epctcep, tcepgasoline, gasolinecapillary, capillaryvalve, valvecolumn, columnpcm, pcmmtbe, mtbesplit, splitprecision, precisionnonaromatic, nonaromaticfid, fidtoluene, toluenehydrocarbons
Determining Aromatics in Finished Gasoline Using the Agilent 6820 GC System
Determining Aromatics in Finished Gasoline Using the Agilent 6820 GC System Application Gasoline Analysis and Environmental Protection Authors ChunXiao Wang Agilent Technologies Co., Ltd. (Shanghai) 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai 200131, P.R.C. Zhenxi Guan Agilent Technologies Co.,…
Key words
tcep, tceparomatics, aromaticsfid, fidvalve, valvebenzene, benzenefinished, finishedswitch, switchnds, ndsgasoline, gasolinerepeatability, repeatabilityxylene, xyleneinlet, inletcapillary, capillaryintercept, intercepttoluene
Equivalent GC systems performance for regulatory method compliance and validation
WHITE PAPER 74062 Equivalent GC systems performance for regulatory method compliance and validation Author: Giulia Riccardino, David Lee, and Cristian Cojocariu, Thermo Fisher Scientific, Runcorn, UK Keywords: Gas Chromatography, TRACE 1300 Series, TRACE 1600 Series, method transfer, method portability, system…
Key words
methyl, methylinstant, instantequivalency, equivalencyfid, fidgro, grochromatographic, chromatographicprocedure, procedureconnect, connectsuitability, suitabilityresidual, residualwisconsin, wisconsinmethod, methodssl, sslrequirements, requirementsheadspace
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike