Capillary Flow Technology: Purged Ultimate Union
Brochures and specifications | 2013 | Agilent TechnologiesInstrumentation
Routine maintenance of gas chromatography–mass spectrometry (GC/MS) systems typically requires cooling, venting, and lengthy pump-down procedures, resulting in significant instrument downtime and reduced laboratory productivity. Capillary Flow Technology’s Purged Ultimate Union (also known as the Pressure Controlled Tee) addresses this challenge by enabling in situ backflush operations without shutting down the MS detector, thereby streamlining maintenance workflows and preserving analytical performance.
This application note describes the design, implementation, and performance evaluation of Agilent’s Purged Ultimate Union for GC/MS. The primary goals are to demonstrate rapid column and inlet maintenance without venting the mass spectrometer, to evaluate the effect of backflush on analytical metrics in selected reaction monitoring (SRM) and multiple reaction monitoring (MRM) modes, and to validate retention time reproducibility after column trimming or replacement.
The Purged Ultimate Union is installed between the inlet (split/splitless, multimode inlet, or programmable temperature vaporizer) and the MS detector. Two main backflush configurations are compared:
Pre-column arrangements (coated or uncoated) and post-column wizards facilitate pressure compensation and consistent flow. Mechanical Retention Time Locking ensures stable retention times across column changes.
Comparative data show negligible differences in retention time variation (<0.1 s RSD), normalized peak area, height, and signal-to-noise ratio with and without the Purged Ultimate Union. Table 1 results for the m/z 272→241 transition of octafluoronaphthalene demonstrate equivalent analytical precision. Retention time error for a panel of twenty organophosphorus pesticides remained within ±1 s, confirming reliable performance after column modifications.
Ongoing developments may integrate automated backflush control within laboratory information management systems (LIMS) and extend pressure-controlled unions to other detector types such as time-of-flight and tandem mass spectrometers. Further miniaturization and software-driven flow optimization could enhance throughput in high-density analytical platforms.
Agilent’s Capillary Flow Technology Purged Ultimate Union offers a robust, user-friendly solution for rapid GC/MS maintenance. By enabling backflush without MS shutdown, it minimizes downtime, maintains analytical integrity, and simplifies method management across varied applications.
GC
IndustriesManufacturerAgilent Technologies
Summary
Significance of the Topic
Routine maintenance of gas chromatography–mass spectrometry (GC/MS) systems typically requires cooling, venting, and lengthy pump-down procedures, resulting in significant instrument downtime and reduced laboratory productivity. Capillary Flow Technology’s Purged Ultimate Union (also known as the Pressure Controlled Tee) addresses this challenge by enabling in situ backflush operations without shutting down the MS detector, thereby streamlining maintenance workflows and preserving analytical performance.
Objectives and Study Overview
This application note describes the design, implementation, and performance evaluation of Agilent’s Purged Ultimate Union for GC/MS. The primary goals are to demonstrate rapid column and inlet maintenance without venting the mass spectrometer, to evaluate the effect of backflush on analytical metrics in selected reaction monitoring (SRM) and multiple reaction monitoring (MRM) modes, and to validate retention time reproducibility after column trimming or replacement.
Methodology
The Purged Ultimate Union is installed between the inlet (split/splitless, multimode inlet, or programmable temperature vaporizer) and the MS detector. Two main backflush configurations are compared:
- Post-run backflush: activated after target analytes elute, preventing late-eluting compounds from reaching the detector.
- Concurrent backflush: executed during data acquisition to further reduce cycle time.
Pre-column arrangements (coated or uncoated) and post-column wizards facilitate pressure compensation and consistent flow. Mechanical Retention Time Locking ensures stable retention times across column changes.
Main Results and Discussion
Comparative data show negligible differences in retention time variation (<0.1 s RSD), normalized peak area, height, and signal-to-noise ratio with and without the Purged Ultimate Union. Table 1 results for the m/z 272→241 transition of octafluoronaphthalene demonstrate equivalent analytical precision. Retention time error for a panel of twenty organophosphorus pesticides remained within ±1 s, confirming reliable performance after column modifications.
Benefits and Practical Applications
- Reduced instrument downtime by eliminating cooling and venting cycles.
- Risk-free inlet and column maintenance without air ingress into the MS source.
- Increased sample throughput through rapid removal of high-molecular-weight matrix compounds.
- Unchanged detection limits and chromatographic reproducibility in MRM workflows.
Future Trends and Potential Applications
Ongoing developments may integrate automated backflush control within laboratory information management systems (LIMS) and extend pressure-controlled unions to other detector types such as time-of-flight and tandem mass spectrometers. Further miniaturization and software-driven flow optimization could enhance throughput in high-density analytical platforms.
Conclusion
Agilent’s Capillary Flow Technology Purged Ultimate Union offers a robust, user-friendly solution for rapid GC/MS maintenance. By enabling backflush without MS shutdown, it minimizes downtime, maintains analytical integrity, and simplifies method management across varied applications.
Used Instrumentation
- Agilent 7890 or 6890N GC with 5977A Series GC/MSD
- Purged Ultimate Union accessory (Pressure Controlled Tee)
- Auxiliary Electronic Pneumatic Control (EPC) module with purge regulator kit
- High-temperature Siltite ferrules and fittings
- 100- and 110 µm inner diameter restrictors
- Flow Calculator or Method Translation Software
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Capillary Flow Technology: Backflush
2013|Agilent Technologies|Brochures and specifications
Capillary Flow Technology: Backflush REDUCE RUN TIME AND INCREASE LABORATORY THROUGHPUT Increase productivity by dramatically reducing analytical run time Many GC and GC/MS analyses are performed on complex samples that contain high-boiling compounds. Other methods may require detection and quantitation…
Key words
backflush, backflushbackflushing, backflushingepc, epcwizard, wizardcapillary, capillarycft, cftagilent, agilentaux, auxconfiguration, configurationdevice, devicefirst, firstflow, flowtechnology, technologybackflush, backflushflow
Maintaining Compound Retention Times with the Backflush Enabled Pressure Controlled Tee Configuration for Agilent 7890A GCs with Agilent 5975 Series MSD and Agilent 7000 Series Triple Quadrupole MS Systems
2009|Agilent Technologies|Technical notes
Maintaining Compound Retention Times with the Backflush Enabled Pressure Controlled Tee Configuration for Agilent 7890A GCs with Agilent 5975 Series MSD and Agilent 7000 Series Triple Quadrupole MS Systems Application Note Environmental Authors Abstract Harry Prest The Pressure Controlled Tee…
Key words
retention, retentionlocking, lockingtee, teecompound, compoundchlorpyrifos, chlorpyrifosmethyl, methyltimes, timespct, pctazinphos, azinphosbackflushing, backflushingmaintaining, maintainingconfiguration, configurationrapid, rapidisomer, isomermechanical
Capillary Flow Technology: Splitters
2013|Agilent Technologies|Brochures and specifications
Capillary Flow Technology: Splitters GET MORE INFORMATION IN LESS TIME A time-saving alternative to running samples on multiple GCs Analyzing complex samples may require different types of specific GC detectors in order to meet detection limits, overcome matrix interferences, or…
Key words
splitter, splitterepc, epcaux, auxmakeup, makeupbackflush, backflushway, waydetectors, detectorscolumn, columnsplitters, splittersgas, gaspcm, pcmtwo, twoagilent, agilentdfpd, dfpdeffluent
Implementation of the Pressure Controlled Tee for Backflushing for the 7000 Series Triple Quadrupole Mass Spectrometer: Implications for Sensitivity
2009|Agilent Technologies|Technical notes
Implementation of the Pressure Controlled Tee for Backflushing for the 7000 Series Triple Quadrupole Mass Spectrometer: Implications for Sensitivity Application Note Environmental, Drug testing, Forensics and Metabolomics Author Abstract Harry Prest The 7000 Series Triple Quadrupole Mass Spectrometer is designed…
Key words
backflushing, backflushingpct, pcttee, teemrm, mrmventing, ventingwithout, withoutofn, ofnnormalized, normalizedconfiguration, configurationsignal, signalruntime, runtimereaction, reactioncfts, cftscarryover, carryoverrelaxed