Optimization of Analytical Conditions for Improved Sensitivity in the Analysis of Deca-BDE on a DFS High Resolution GC/HRMS system
Presentations | | Thermo Fisher ScientificInstrumentation
Brominated diphenyl ethers (PBDE) are widely used flame retardants with environmental persistence and diverse bromination degrees. The analysis of high-brominated congeners such as deca-BDE presents special challenges: thermal lability, high boiling points, and a broad mass range requiring sensitive detection and robust chromatographic separation.
The optimized analytical conditions on a high-resolution GC/HRMS platform, combining PTV on-column injection, pressure pulsing, and a dual-column configuration, significantly improved the sensitivity, stability, and separation of PBDE congeners. This approach enables reliable quantification of deca-BDE at sub-picogram levels and high throughput for routine environmental analysis.
Thermo Scientific application note Optimization of Analytical Conditions for Improved Sensitivity in the Analysis of Deca-BDE on a DFS High Resolution GC/HRMS System, Dirk Krumwiede.
GC/MSD, GC/HRMS
IndustriesManufacturerThermo Fisher Scientific
Summary
Importance of the Topic
Brominated diphenyl ethers (PBDE) are widely used flame retardants with environmental persistence and diverse bromination degrees. The analysis of high-brominated congeners such as deca-BDE presents special challenges: thermal lability, high boiling points, and a broad mass range requiring sensitive detection and robust chromatographic separation.
Objectives and Study Overview
- Investigate key analytical factors affecting deca-BDE sensitivity on a high-resolution GC/HRMS system.
- Compare splitless (SSL) and programmable temperature vaporization (PTV) injectors for thermal degradation and sensitivity.
- Develop a dual-column configuration to combine optimized sensitivity for deca-BDE with efficient separation of lower-brominated congeners.
Methodology and Instrumentation
- Instrumentation:
- Thermo Scientific DFS high-resolution sector field mass spectrometer
- Trace GC Ultra with dual injectors (SSL and PTV)
- DB-5MS columns: 15 m × 0.25 mm i.d. (0.1 µm film), 30 m × 0.25 mm (0.1 µm), and 6 m × 0.20 mm (0.1 µm)
- Y-shaped dual column transfer line adaptor
- Key parameters optimized:
- Injector type and liner dimensions
- Splitless time and pressure pulse settings
- Column length, carrier gas flow, and oven temperature program
- Ion source temperature (270–280 °C) and ionization energy (35–40 eV)
- Mass detection focused on molecular ions (M+) and reductive isotope peaks ([M–2Br]+, [M–4Br]+) to maximize signal intensity across congeners.
Main Results and Discussion
- Molecular ion behavior shifts with bromination degree; high-brominated congeners show more intense [M–2Br]+ or [M–4Br]+ peaks than M+.
- PTV injection outperformed SSL for deca-BDE, providing stable sensitivity over extended runs and reducing thermal decomposition.
- Optimal splitless times (≈1.6 min) and 20 psi pressure pulses enhanced analyte transfer for SSL; PTV on-column injection further minimized degradation.
- Column length had a key effect on sensitivity: 6 m column achieved detection limits of 0.1–0.5 pg for deca-BDE, compared to 1–5 pg on 15 m and >10 pg on 30 m.
- Dual-column setup enabled sequential runs from the same vial: 6 m PTV column for deca-BDE sensitivity and 30 m SSL column for separation of lower congeners.
- Linearity and repeatability on the 6 m column demonstrated excellent quantification across 0.25 to 1 pg levels (R² ≈ 0.9998).
Benefits and Practical Applications
- Sub-picogram detection of deca-BDE and other PBDE congeners for environmental monitoring and regulatory compliance.
- Reduced sample degradation and improved method robustness using PTV on-column injection.
- Enhanced laboratory throughput with dual GC setup and staggered injections.
- Streamlined data processing using advanced quantification tools like TargetQuan.
Future Trends and Opportunities
- Extension of dual-column, dual-injector strategies to other persistent organic pollutants (POPs) and high-boiling analytes.
- Integration of large-volume injection techniques (PTV LVI) and automated sample preparation for trace analysis in complex matrices.
- Adoption of high-throughput quantification software and machine learning for real-time method optimization.
- Advancements in GC/HRMS interface design to further minimize thermal decomposition and maximize sensitivity.
Conclusion
The optimized analytical conditions on a high-resolution GC/HRMS platform, combining PTV on-column injection, pressure pulsing, and a dual-column configuration, significantly improved the sensitivity, stability, and separation of PBDE congeners. This approach enables reliable quantification of deca-BDE at sub-picogram levels and high throughput for routine environmental analysis.
Reference
Thermo Scientific application note Optimization of Analytical Conditions for Improved Sensitivity in the Analysis of Deca-BDE on a DFS High Resolution GC/HRMS System, Dirk Krumwiede.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Modernizing Persistent Organic Pollutants (POPs) Analysis
2019|Thermo Fisher Scientific|Presentations
Modernizing Persistent Organic Pollutants (POPs) Analysis The world leader in serving science Agenda • Introduction to POPs • Dioxin analysis with the DFS • Recent regulatory changes: • Dioxin analysis in food with the Thermo Scientific™ TSQ™ 9000 triple quadrupole…
Key words
abundance, abundancepops, popsdfs, dfsrelative, relativeicis, icisorbitrap, orbitrapbase, basedioxin, dioxinpeak, peaksector, sectoranalysis, analysismin, minthermo, thermosccps, sccpscalculated
Magnetic Sector GC-HRMS with DualData XL Option: helium savings and flow optimization for maximized productivity and cost savings for PBDE, dioxin and PCB analysis
2019|Thermo Fisher Scientific|Posters
Magnetic Sector GC-HRMS with DualData XL Option: helium savings and flow optimization for maximized productivity and cost savings for PBDE, dioxin and PCB analysis Heinz Mehlmann, Dirk Krumwiede, Thermo Fisher Scientific, Thermo Fisher Scientific, Hanna-Kunath Str. 11, Bremen, Germany, 20199…
Key words
dfs, dfsdualdata, dualdataicis, icismodule, moduleabundance, abundancesaver, saverhelium, heliuminstant, instantrelative, relativestaggered, staggeredinjector, injectormcd, mcdoption, optiontime, timedead
DIOXIN: Magnetic Sector GC-HRMS with DualData XL Option: helium savings and flow optimization for maximized productivity and cost savings for PBDE, dioxin and PCB analysis
2019|Thermo Fisher Scientific|Posters
Magnetic Sector GC-HRMS with DualData XL Option: helium savings and flow optimization for maximized productivity and cost savings for PBDE, dioxin and PCB analysis Heinz Mehlmann, Dirk Krumwiede, Thermo Fisher Scientific, Thermo Fisher Scientific, Hanna-Kunath Str. 11, Bremen, Germany, 20199…
Key words
dfs, dfsdualdata, dualdataicis, icismodule, moduleabundance, abundancesaver, saverhelium, heliuminstant, instantrelative, relativemcd, mcdstaggered, staggeredinjector, injectoroption, optiontime, timedead
Determination of Polybrominated Diphenyl Ethers in Polymeric Materials Using the 6890 GC/5973N inert MSD with Electron Impact Ionization
2005|Agilent Technologies|Applications
Determination of Polybrominated Diphenyl Ethers in Polymeric Materials Using the 6890 GC/5973N inert MSD with Electron Impact Ionization Application Environmental, Component Testing Authors Introduction Chuanhong Tu Agilent Technologies, PuDong Shanghai CHINA Polybrominated diphenyl ethers (PBDEs) are a major issue in…
Key words
pbde, pbdeabundance, abundancepbdes, pbdeshexabdes, hexabdesoctabdes, octabdeshexa, hexapolymer, polymerdeca, decasim, simheptabdes, heptabdesnonabdes, nonabdespolybrominated, polybrominatedpolymeric, polymericcongeners, congenersdiphenyl