Analysis of PCBs in Food and Biological Samples Using GC Triple Quadrupole GC-MS/MS
Applications | 2010 | Thermo Fisher ScientificInstrumentation
PCBs are globally distributed persistent organic pollutants with recognized toxicity. Monitoring their levels in food and biological matrices is critical for ensuring public health and compliance with international regulations such as the Stockholm Convention and EU directives.
This study presents a fast and selective screening and quantitation method for PCBs in environmental, food, and biological samples. It adapts USEPA Method 1668A to a triple quadrupole GC-MS/MS platform, aiming to achieve low detection limits and robust confirmation according to EU Commission Directive 96/23/EC.
The analytical workflow is based on isotope dilution quantitation, using 13C-labeled PCB congeners as internal standards. Key instrumental setup:
The method demonstrated high sensitivity, with on-column injections as low as 1 pg yielding clear chromatographic peaks for PCBs from mono- to decachlorinated congeners. In complex matrices (blood, milk, egg yolk, green cabbage), background interference was negligible. The approach achieved 6-8 data points per peak, ensuring reliable integration.
Advancements in triple quadrupole technology may further lower detection limits and streamline workflows. Integration with automated sample preparation and data processing could enhance laboratory efficiency. Expanding the approach to emerging persistent organic pollutants and metabolite profiling offers additional applications in environmental and clinical studies.
The TSQ Quantum XLS GC-MS/MS method provides a rapid, reliable solution for trace-level PCB analysis in complex matrices, fulfilling regulatory requirements and delivering robust quantitation with isotope dilution.
GC/MSD, GC/MS/MS, GC/QQQ
IndustriesFood & Agriculture, Forensics
ManufacturerThermo Fisher Scientific
Summary
Importance of the Topic
PCBs are globally distributed persistent organic pollutants with recognized toxicity. Monitoring their levels in food and biological matrices is critical for ensuring public health and compliance with international regulations such as the Stockholm Convention and EU directives.
Objectives and Study Overview
This study presents a fast and selective screening and quantitation method for PCBs in environmental, food, and biological samples. It adapts USEPA Method 1668A to a triple quadrupole GC-MS/MS platform, aiming to achieve low detection limits and robust confirmation according to EU Commission Directive 96/23/EC.
Methodology and Instrumentation
The analytical workflow is based on isotope dilution quantitation, using 13C-labeled PCB congeners as internal standards. Key instrumental setup:
- GC-MS/MS System: Thermo Scientific TSQ Quantum XLS with TRACE GC Ultra and TriPlus autosampler
- Column: TRACE TR-Dioxin 5MS (30 m × 0.25 mm, 0.10 µm film)
- Ionization: EI at 40 eV, source at 240 °C
- SRM Transitions: Two precursor ions per congener and two product ions (loss of Cl2), providing five identification points
- Chromatography: Split/splitless injection at 260 °C, He carrier at 0.8 mL/min, 32 min temperature program
Main Results and Discussion
The method demonstrated high sensitivity, with on-column injections as low as 1 pg yielding clear chromatographic peaks for PCBs from mono- to decachlorinated congeners. In complex matrices (blood, milk, egg yolk, green cabbage), background interference was negligible. The approach achieved 6-8 data points per peak, ensuring reliable integration.
Benefits and Practical Applications of the Method
- Enhanced selectivity and confirmation through MS/MS with hyperbolic quadrupole
- Compliance with USEPA Method 1668A and EU identification criteria (minimum 4, achieved 5 points)
- Low detection limits in challenging matrices without extensive cleanup
- Increased throughput for routine screening in food safety and environmental monitoring
Future Trends and Potential Applications
Advancements in triple quadrupole technology may further lower detection limits and streamline workflows. Integration with automated sample preparation and data processing could enhance laboratory efficiency. Expanding the approach to emerging persistent organic pollutants and metabolite profiling offers additional applications in environmental and clinical studies.
Conclusion
The TSQ Quantum XLS GC-MS/MS method provides a rapid, reliable solution for trace-level PCB analysis in complex matrices, fulfilling regulatory requirements and delivering robust quantitation with isotope dilution.
References
- USEPA Method 1668A: Chlorinated Biphenyl Congeners by HRGC/HRMS, EPA-821-R-00-002
- EU Commission Directive 96/23/EC on analytical method performance and result interpretation
- Turner W.E. et al., Proceedings of the Dioxin Conference, Oslo 2006
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
PCDD/F Screening at the Maximum Residue Level for Food Safety Analysis using Highly Selective Triple Quadrupole GC/MS/MS
2010|Thermo Fisher Scientific|Applications
Application Note: 51843 PCDD/F Screening at the Maximum Residue Level for Food Safety Analysis using Highly Selective Triple Quadrupole GC/MS/MS Dirk Krumwiede, Hans-Joachim Huebschmann, Thermo Fisher Scientific, Bremen, Germany Abstract Key Words • EPA 1613A • Food Safety Analysis •…
Key words
xls, xlstsq, tsqquantum, quantumscreening, screeningdioxins, dioxinsions, ionsconfirmatory, confirmatorypcbs, pcbshyperbolic, hyperbolicpcdd, pcddlevel, levelhecdd, hecddhecdf, hecdfscientific, scientifictrace
Reproducible trace analysis of PCBs in environmental matrices using triple quadrupole GC-MS/MS
2022|Thermo Fisher Scientific|Applications
Application note | 000561 Mass Spectrometry Reproducible trace analysis of PCBs in environmental matrices using triple quadrupole GC-MS/MS Authors Goal Kjell Hope , Giulia Riccardino , Adam The aim of this application note is to demonstrate the performance of the…
Key words
min, minpcbs, pcbsidls, idlsaei, aeiinjection, injectiontime, timenevervent, neverventallowed, allowedwash, washcounts, countscalculated, calculatedcompound, compoundensured, ensuredsmarttune, smarttuneion
NEMC: Reproducible trace analysis of PCBs in environmental matrices using triple quadrupole GC-MS/MS
2022|Thermo Fisher Scientific|Presentations
Reproducible trace analysis of PCBs in environmental matrices using triple quadrupole GC-MS/MS Andy Fornadel, PhD Product Marketing Manager – Americas 4th August 2022 The world leader in serving science 1 [email protected] | 4-August-2022 Content 2 1 Introduction to analysis of…
Key words
pcbs, pcbspcb, pcbclean, cleanpersistence, persistenceabsolute, absolutecompound, compoundmin, mincongeners, congenersalumina, aluminarsd, rsdcounts, countstoxicity, toxicitytenacity, tenacityanalyis, analyisgel
EAS: Analysis of PAHs and PCBs in multiple matrices using GC-MS and GC-MS/MS
2022|Thermo Fisher Scientific|Presentations
Analysis of PAHs and PCBs in multiple matrices using GC-MS and GC-MS/MS Andy Fornadel, PhD Product Marketing Manager – Americas November 15th, 2022 The world leader in serving science 1 [email protected] | 15-November-2022 Agenda 2 1 Introduction to polycyclic hydrocarbons…
Key words
pcbs, pcbspahs, pahscounts, countspcb, pcbisobaric, isobaricdftpp, dftppavrf, avrfanalysis, analysissoil, soilchromatographic, chromatographicmin, mineluting, elutingbroadening, broadeningpolycyclic, polycyclicinterferences