High Efficiency, Quantitative Dioxins Screening at the Level of Interest in Feed and Food using Advanced GC-MS/MS
Applications | 2011 | Thermo Fisher ScientificInstrumentation
Dioxins and dioxin-like compounds pose significant health risks through bioaccumulation in the food chain, chiefly from animal-derived products.
Regulatory authorities enforce strict maximum levels and extensive monitoring to protect consumers and avoid economic and political fallout.
This study evaluates the performance of the Thermo Scientific TSQ Quantum XLS Ultra GC-MS/MS platform for high-efficiency screening of PCDD/Fs in feed and food.
It benchmarks screening results against the gold standard GC-HRMS (Thermo Scientific DFS) under EU regulation thresholds.
Extraction and Clean-up:
Sensitivity and Selectivity:
The TSQ Quantum XLS Ultra GC-MS/MS demonstrates robust sensitivity, selectivity, and quantitative accuracy for dioxin screening in compliance with EU regulations.
Its strong agreement with GC-HRMS confirmatory data and predictable ion ratios establish it as a reliable, efficient tool for routine feed and food safety testing.
GC/MSD, GC/MS/MS, GC/QQQ
IndustriesFood & Agriculture
ManufacturerThermo Fisher Scientific
Summary
Význam tématu
Dioxins and dioxin-like compounds pose significant health risks through bioaccumulation in the food chain, chiefly from animal-derived products.
Regulatory authorities enforce strict maximum levels and extensive monitoring to protect consumers and avoid economic and political fallout.
Cíle a přehled studie / článku
This study evaluates the performance of the Thermo Scientific TSQ Quantum XLS Ultra GC-MS/MS platform for high-efficiency screening of PCDD/Fs in feed and food.
It benchmarks screening results against the gold standard GC-HRMS (Thermo Scientific DFS) under EU regulation thresholds.
Použitá metodika a instrumentace
Extraction and Clean-up:
- Sample extraction and multi-step clean-up following validated protocols, targeting a maximum of 3 g fat for low LOQ.
- GC-MS/MS system: TSQ Quantum XLS Ultra with EI source (40 eV, 250 °C, 50 µA).
- MS/MS parameters: Q2 argon collision (1.5 mTorr, 22 V), Q1/Q3 resolution at 0.7 amu.
- PTV injection: 5 µL toluene, 100 °C ramped to 340 °C.
- GC column: DB-5MS (60 m × 0.25 mm, 0.25 µm) with temperature program from 120 °C to 285 °C.
- Confirmatory analysis: Thermo Scientific DFS high resolution MS.
Hlavní výsledky a diskuse
Sensitivity and Selectivity:
- Achieved detection at 0.13 pg/g fat for 2,3,7,8-TCDD with 12% CV across five injections.
- Consistent SRM chromatograms at lowest calibration levels, enabling precise quantification.
- Measured ion ratios matched theoretical values within ±15% tolerance, ensuring reliable identity confirmation.
- Isotope dilution quantification provided real TEQ values and recoveries of 83–110% with RSDs mostly below 20% for 42 food samples.
- WHO-PCDD/F-TEQ results correlated strongly with HRMS data down to ~0.5 pg/g level, indicating low false negatives and high selectivity.
Přínosy a praktické využití metody
- High-throughput screening reduces reliance on time-intensive HRMS confirmatory analyses.
- Retention of isotope dilution and congener profiling supports accurate TEQ calculation and monitoring trends.
- Compliance with EU regulations for screening minimizes non-compliant sample releases.
Budoucí trendy a možnosti využití
- Expansion to other persistent organic pollutants and further multi-residue applications.
- Integration with automated sample preparation and miniaturized platforms to increase throughput.
- Potential coupling with rapid bioassays for tiered screening strategies.
Závěr
The TSQ Quantum XLS Ultra GC-MS/MS demonstrates robust sensitivity, selectivity, and quantitative accuracy for dioxin screening in compliance with EU regulations.
Its strong agreement with GC-HRMS confirmatory data and predictable ion ratios establish it as a reliable, efficient tool for routine feed and food safety testing.
Reference
- European Food Safety Authority; Results of the monitoring of dioxin levels in food and feed. EFSA Journal 2010;8(3):1385.
- Commission Regulation (EC) No 152/2009 and No 1883/2006.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Determination of PCDD/Fs in Environmental Samples using Accelerated Solvent Extraction (ASE) and GC-MS/MS
2016|Thermo Fisher Scientific|Applications
Paul Silcock,1 David Gardner,2 John Fardon,2 Steve Moss2 and Inge de Dobbeleer1 1 Thermo Scientific POPs Centre of Excellence, Bremen, Germany 2 National Laboratory Service - Environment Agency, Leeds, United Kingdom Key Words Dioxins, furans, PCDD, PCDF, environmental, fly ash,…
Key words
xls, xlsdibenzodioxin, dibenzodioxinpcdd, pcddhrms, hrmsquantum, quantumtsq, tsqocdd, ocdddibenzofuran, dibenzofuranocdf, ocdfultra, ultrapecdd, pecddagreement, agreementottawa, ottawahpcdd, hpcddtcdd
Meeting the European Commission Performance Criteria for the Use of Triple Quadrupole GC-MS/MS as a Confirmatory Method for PCDD/Fs in Food and Feed Samples
2016|Thermo Fisher Scientific|Applications
Appli cat i on N ote 1 0 3 8 0 Meeting the European Commission Performance Criteria for the Use of Triple Quadrupole GC-MS/MS as a Confirmatory Method for PCDD/Fs in Food and Feed Samples Cristian Cojocariu,1 Manuela Abalos,2 Esteban…
Key words
qual, qualquant, quantpcdd, pcddhxcdf, hxcdfocdd, ocddpecdf, pecdfhpcdf, hpcdfhpcdd, hpcddhxcdd, hxcddtcdf, tcdfdioxin, dioxintcdd, tcddwere, weresrm, srmdioxins
PCDD/F Screening at the Maximum Residue Level for Food Safety Analysis using Highly Selective Triple Quadrupole GC/MS/MS
2010|Thermo Fisher Scientific|Applications
Application Note: 51843 PCDD/F Screening at the Maximum Residue Level for Food Safety Analysis using Highly Selective Triple Quadrupole GC/MS/MS Dirk Krumwiede, Hans-Joachim Huebschmann, Thermo Fisher Scientific, Bremen, Germany Abstract Key Words • EPA 1613A • Food Safety Analysis •…
Key words
xls, xlstsq, tsqquantum, quantumscreening, screeningdioxins, dioxinsions, ionsconfirmatory, confirmatorypcbs, pcbshyperbolic, hyperbolicpcdd, pcddlevel, levelhecdd, hecddhecdf, hecdfscientific, scientifictrace
Tetra- Through Octa-Chlorinated Dioxins and Furans Analysis in Water by Isotope Dilution GC/MS/MS
2019|Agilent Technologies|Applications
Application Note Environmental, Food Testing and Agriculture Tetra- Through Octa-Chlorinated Dioxins and Furans Analysis in Water by Isotope Dilution GC/MS/MS Authors Hui Lin, Jessica Betz, Jim Dunn, Greg Martin, and David Somerville The DOW Chemical Company Midland, MI, USA Diana…
Key words
cdds, cddscdfs, cdfspecdf, pecdftcdd, tcddtcdf, tcdfdioxin, dioxinhxcdf, hxcdfpecdd, pecddhpcdf, hpcdfhpcdd, hpcddhxcdd, hxcdddioxins, dioxinsabundance, abundancerrt, rrtocdf