Operating principle of medium pressure flow controller that allows catalyst screening at varied pressures
Technical notes | | Frontier LabInstrumentation
In catalyst screening, the ability to decouple reaction pressure from chromatographic flow is critical for reliable evaluation of conversion and selectivity. Independent control of these parameters ensures that reactor conditions can be optimized without compromising separation performance or detector sensitivity.
This technical note describes the design and operation of a medium pressure flow controller (MP-3050FC) integrated into a Tandem µ-Reactor system (Rx-3050TR). The study aims to maintain constant GC column flow rates while varying reactor pressures between 0.3 and 0.98 MPa, thereby enabling catalyst tests under diverse pressure conditions without degrading chromatographic resolution.
The modified reactor assembly comprises:
Adjusting the reactor inlet pressure does not alter column head pressure, preserving stable gas flow and detector vacuum.
Introducing the MP-3050FC successfully decouples reactor and column pressures. Experimental data demonstrate that, over the 0.3–0.98 MPa reaction pressure range, the GC column flow remains constant. This stability preserves chromatographic peak shape, resolution, and mass spectrometer vacuum integrity, facilitating more accurate catalyst activity and selectivity assessments.
Further developments may include miniaturized pressure controllers for multi-channel screening, automated feedback loops for real-time pressure adjustment, and broader pressure ranges for simulating industrial conditions. Integration with advanced detectors and data analytics will accelerate catalyst discovery workflows.
The medium pressure flow controller MP-3050FC effectively maintains constant column flow while allowing reaction pressure variation. Its integration into the Tandem µ-Reactor enhances the reliability and throughput of catalyst screening experiments under diverse pressure regimes.
Thermal desorption
IndustriesManufacturerFrontier Lab
Summary
Importance of the Topic
In catalyst screening, the ability to decouple reaction pressure from chromatographic flow is critical for reliable evaluation of conversion and selectivity. Independent control of these parameters ensures that reactor conditions can be optimized without compromising separation performance or detector sensitivity.
Objectives and Study Overview
This technical note describes the design and operation of a medium pressure flow controller (MP-3050FC) integrated into a Tandem µ-Reactor system (Rx-3050TR). The study aims to maintain constant GC column flow rates while varying reactor pressures between 0.3 and 0.98 MPa, thereby enabling catalyst tests under diverse pressure conditions without degrading chromatographic resolution.
Methodology and Instrumentation
The modified reactor assembly comprises:
- An electronic flow controller in the GC pneumatic manifold to set reaction pressure.
- A back-pressure control valve on the MP-3050FC to regulate column head pressure independently.
- A restrictor tube and Open-Split Interface to guide effluent to the separation column or mass spectrometer.
Adjusting the reactor inlet pressure does not alter column head pressure, preserving stable gas flow and detector vacuum.
Used Instrumentation
- Tandem µ-Reactor Rx-3050TR
- Medium Pressure Flow Controller MP-3050FC
- GC pneumatic system with electronic pressure control (EPC/AFC)
- Back-pressure control valve integrated in MP-3050FC
- Restrictor tubing and Open-Split Interface
- Mass spectrometer detector
Key Results and Discussion
Introducing the MP-3050FC successfully decouples reactor and column pressures. Experimental data demonstrate that, over the 0.3–0.98 MPa reaction pressure range, the GC column flow remains constant. This stability preserves chromatographic peak shape, resolution, and mass spectrometer vacuum integrity, facilitating more accurate catalyst activity and selectivity assessments.
Benefits and Practical Applications
- Unbiased catalyst screening under variable pressures without flow‐rate artifacts.
- Improved chromatographic separation and detector stability.
- Enhanced reproducibility in high‐throughput catalyst evaluation.
- Flexible integration with existing GC-based µ-Reactor systems.
Future Trends and Opportunities
Further developments may include miniaturized pressure controllers for multi-channel screening, automated feedback loops for real-time pressure adjustment, and broader pressure ranges for simulating industrial conditions. Integration with advanced detectors and data analytics will accelerate catalyst discovery workflows.
Conclusion
The medium pressure flow controller MP-3050FC effectively maintains constant column flow while allowing reaction pressure variation. Its integration into the Tandem µ-Reactor enhances the reliability and throughput of catalyst screening experiments under diverse pressure regimes.
References
- Frontier Laboratories Ltd. MP-3050FC Medium Pressure Flow Controller Brochure.
- Technical Note RXT-001E, Frontier Laboratories Ltd.
- Technical Note RXA-007E, Frontier Laboratories Ltd.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Operating principle of High Pressure Flow Controller which allows catalyst screening at varied pressures
|Frontier Lab|Technical notes
µ-Reactor Technical Note ( RXT-003E ) Operating principle of High-Pressure Flow Controller which allows catalyst screening at varied pressures [Background] In catalyst screening experiments, varying reaction conditions such as temperature and pressure are essential to investigate the conversion efficiency and…
Key words
controller, controllerpressure, pressurehigh, highreaction, reactionega, egainquiries, inquiriesrestrictor, restrictorflow, flowreactor, reactorcatalyst, catalystsend, sendvaried, variedpressures, pressuresforward, forwardprinciple
Effect of pressure on catalytic conversion efficiency - Catalytic reaction of glycerin with a Pd catalyst
|Frontier Lab|Applications
µ-Reactor Technical Note ( RXA-007E ) Effect of pressure on catalytic conversion efficiency - Catalytic reaction of glycerin with a Pd catalyst [Background] Comprehensive catalyst characterization requires an understanding of how factors such as reaction temperature and pressure impact the…
Key words
reaction, reactionpropane, propanepressure, pressureglycerin, glycerincatalytic, catalyticmpa, mpabutane, butaneethane, ethanecatalyst, catalystpentane, pentanehexane, hexanetic, ticinquiries, inquiriesreactor, reactorsend
Frontier Lab Medium Pressure Flow controller - Optional accessory for Tandem and Single μ-Reactors
|Frontier Lab|Brochures and specifications
NEW ! Optional accessory for Tandem and Single µ-Reactors Medium Pressure Flow controller; MP-3050FC Online analysis of medium pressure catalytic reaction MAJOR FEATURES 1) New flow control system allows online GC analysis of products formed by catalytic reaction under 1…
Key words
controller, controllerpressure, pressurereactors, reactorsethylene, ethylenemedium, mediumunaffected, unaffectedtandem, tandemrestrictor, restrictorreaction, reactionopen, openflow, flowtube, tubesplit, splitega, egainj
Frontier Lab High Pressure Flow controller; HP-3050FC - Optional accessory for Tandem and Single μ-Reactors
|Frontier Lab|Brochures and specifications
NEW ! Optional accessory for Tandem and Single µ-Reactors High Pressure Flow controller; HP-3050FC MAJOR FEATURES Retention times stay unchanged at different catalyst pressures 1) Online High pressure (HP) reaction analysis by innovative pneumatics A new flow system permits ideal…
Key words
pressure, pressurereactor, reactorreactors, reactorsreaction, reactionhigh, highpneumatics, pneumaticsonline, onlinecatalyst, catalystoff, offsolid, solidethylene, ethyleneinnovative, innovativelimiter, limitersampler, samplerflow