Fast Screening, Identification, and Quantification of Pesticide Residues in Baby Food Using GC Orbitrap MS Technology
Applications | 2016 | Thermo Fisher ScientificInstrumentation
Baby food safety and pesticide residue control is critical due to infants’ vulnerability and strict regulatory maximum residue levels (MRLs). High-throughput, accurate screening methods ensure compliance and protect public health.
This application note demonstrates full-scan GC-Orbitrap™ MS technology to screen, identify, and quantify 132 pesticides in baby food using fast GC separation and high-resolution accurate mass detection. The aim is to achieve broad-scope analysis with sensitivity and quantitative performance comparable to triple quadrupole GC-MS/MS under SANCO guidelines.
Sample Preparation
Instrumentation
GC-MS Conditions
Data Processing
Chromatography and Acquisition Speed
Targeted Screening
Sensitivity and Detection Limits
Quantification Performance
Mass Accuracy and Resolution
Full-scan GC-Orbitrap MS enables broad-scope screening and quantification in one run, reduces false negatives, supports EU regulatory compliance, and enhances laboratory throughput.
High-resolution GC-MS may integrate automated workflows, expand non-targeted and retrospective analysis, and extend to diverse food matrices for multi-residue monitoring.
GC-Orbitrap MS delivers sensitive, accurate, and selective pesticide analysis in baby food with throughput and performance comparable to triple quadrupole systems, while offering unlimited screening scope.
GC/MSD, GC/MS/MS, GC/HRMS, GC/Orbitrap
IndustriesFood & Agriculture
ManufacturerThermo Fisher Scientific
Summary
Significance of the Topic
Baby food safety and pesticide residue control is critical due to infants’ vulnerability and strict regulatory maximum residue levels (MRLs). High-throughput, accurate screening methods ensure compliance and protect public health.
Study Objectives and Overview
This application note demonstrates full-scan GC-Orbitrap™ MS technology to screen, identify, and quantify 132 pesticides in baby food using fast GC separation and high-resolution accurate mass detection. The aim is to achieve broad-scope analysis with sensitivity and quantitative performance comparable to triple quadrupole GC-MS/MS under SANCO guidelines.
Methodology and Instrumentation
Sample Preparation
- Baby food extracts prepared by citrate-buffered QuEChERS
- Extracts in acetonitrile (1 g/mL) spiked with pesticides at 0.5–200 ng/g
Instrumentation
- Q Exactive GC hybrid quadrupole-Orbitrap MS
- TRACE 1310 GC with TriPlus RSH autosampler
- TraceGOLD TG-5SilMS, 15 m × 0.25 mm i.d. × 0.25 µm column
GC-MS Conditions
- Oven program: 40 °C (1.5 min) to 180 °C at 25 °C/min, then to 300 °C at 100 °C/min (3 min)
- PTV splitless injection, He carrier gas (1.2 mL/min)
- Mass range 50–500 m/z, 60,000 FWHM at m/z 200, EI mode at 70 eV
- Lockmass m/z 207.03235 for <0.5 ppm mass accuracy
Data Processing
- TraceFinder software for screening, library matching (NIST14), and quantitation
Main Results and Discussion
Chromatography and Acquisition Speed
- Efficient separation: 2.5 s peak widths, >12 points/peak at 60,000 resolution
- Clean extracted ion chromatograms for early and late eluting pesticides
Targeted Screening
- 132 pesticides identified in 10 ng/g matrix by RT (±0.1 min), mass (±2 ppm), isotopic pattern, and fragment ions
Sensitivity and Detection Limits
- 95% of pesticides detected at 0.5–1 ng/g
- IDL99 <2 ng/g for 91% of compounds, comparable to triple quadrupole GC-MS/MS
Quantification Performance
- Matrix-matched calibration (0.5–100 ng/g) with R² >0.99 and residuals <25%
- Peak area repeatability %RSD ~6% at 5–10 ng/g
Mass Accuracy and Resolution
- Sub-ppm mass errors (average 0.4 ppm) across all analytes
- 60,000 FWHM resolution eliminates isobaric interferences
Benefits and Practical Applications
Full-scan GC-Orbitrap MS enables broad-scope screening and quantification in one run, reduces false negatives, supports EU regulatory compliance, and enhances laboratory throughput.
Future Trends and Possibilities
High-resolution GC-MS may integrate automated workflows, expand non-targeted and retrospective analysis, and extend to diverse food matrices for multi-residue monitoring.
Conclusion
GC-Orbitrap MS delivers sensitive, accurate, and selective pesticide analysis in baby food with throughput and performance comparable to triple quadrupole systems, while offering unlimited screening scope.
References
- EU Regulation 396/2005 on pesticide MRLs
- SANCO Guidance 12571/2013 on analytical QC
- EU Directive 2003/13/EC on baby food MRLs
- Thermo Scientific AN10432: QuEChERS triple quadrupole GC-MS/MS
- Thermo Scientific TN10455: High-Resolution Compound Detection
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Fast Screening and Quantification of Pesticide Residues in Baby Food Using GC Orbitrap MS Technology
2015|Thermo Fisher Scientific|Posters
F c m Introduction Pesticides are measured almost exclusively by liquid chromatography (LC) and gas chromatography (GC) M analytical methodologies. GC offers good separation U efficiency and a choice of MS detectors such as single or in triple quadrupoles. However,…
Key words
idl, idlpesticide, pesticidebaby, babyusing, usingfull, fullexactive, exactivepesticides, pesticidesrelative, relativefood, foodabundance, abundanceacquired, acquiredfigure, figuremass, massresidues, residuesquantitative
Fast Screening And Quantification Of Pesticide Residues In Baby Food Using GC Orbitrap MS Technology
2016|Thermo Fisher Scientific|Posters
MS instruments. F c m Introduction Pesticides are measured almost exclusively by liquid chromatography (LC) and gas chromatography (GC) analytical methodologies. GC offers good separation efficiency and a choice of MS detectors such as single or triple quadrupoles. However, targeting…
Key words
baby, babyidl, idlfood, foodpesticides, pesticidespesticide, pesticidefull, fullcitrate, citrateusing, usingmass, massalmost, almostassessed, assessedaccuracy, accuracyscan, scantriple, tripleexactive
ISCC 2016: Fast screening and quantification of pesticide residues in baby food using GC Orbitrap MS technology
2016|Thermo Fisher Scientific|Posters
Overview POSTER NOTE Purpose: To evaluate the utility of Orbitrap based GC-MS technology for fast pesticides screening and quantification. Methods: Baby food samples were extracted following QuEChERS protocol. The final extracts were spiked with a mixture of 132 pesticides at…
Key words
exactive, exactivebaby, babypesticide, pesticideidl, idlorbitrap, orbitrapfull, fullfood, foodchromatography, chromatographypesticides, pesticidesscan, scanmass, masstriple, triplefigure, figurequadrupole, quadrupoleusing
Fast Screening and Quantification of Pesticide Residues in Baby Food Using GC-Orbitrap MS Technology
2016|Thermo Fisher Scientific|Posters
1 2 2 1 1 Poster Note 64424 Fast Screening and Quantification of Pesticide Residues in Baby Food Using GC-Orbitrap MS Technology FastScreening Screening and Quantification ofPesticide Pesticide ResiduesininBaby BabyF Fast and Quantification Cristian Cojocariu, Richard Fussell, Mike Hetmanski, Dominicof…
Key words
andand, andandthethe, thethegcgc, gcgcmass, massexactive, exactivepesticides, pesticidesbaby, babymin, minfigure, figurefood, foodscan, scanpesticide, pesticideareare, arearemsms, msmsfwhm