Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Applications | 2018 | Thermo Fisher ScientificInstrumentation
GC/MSD, Purge and Trap, GC/SQ
Industries
Environmental
Manufacturer
Thermo Fisher Scientific, Teledyne LABS

Summary

Importance of the Topic


The analysis of volatile organic compounds (VOC) is a cornerstone of environmental and industrial monitoring. Rapid and sensitive detection of trace-level contaminants supports public health, regulatory compliance, and process control. Traditional gas chromatography–mass spectrometry (GC-MS) methods rely on helium as a carrier gas, but global helium supply constraints and rising costs have spurred development of strategies to minimize helium consumption while maintaining analytical performance.

Objectives and Study Overview


This study assesses a novel combination of the Thermo Fisher Instant Connect Helium Saver Module and the Teledyne Tekmar Atomx automated purge-and-trap system using nitrogen as the purge gas. The goal is to demonstrate that coupling nitrogen purge with helium conservation technology delivers robust VOC analysis in accordance with US EPA Method 8260C, without compromising calibration quality, sensitivity, or method detection limits.

Methodology and Instrumentation


The analytical platform comprised a Thermo Scientific TRACE 1310 gas chromatograph fitted with an Instant Connect Helium Saver Module and an ISQ LT single-quadrupole mass spectrometer. Key conditions included:
  • Splitless injection (1 µL) with glass wool liner
  • TraceGOLD TG-VMS column (20 m × 0.18 mm × 1 µm)
  • Oven program: 35 °C (3 min), 14 °C/min to 100 °C, 25 °C/min to 210 °C (2 min)
  • Helium Saver Module: 200 °C, split flow 12 mL/min, helium delay 0.1 min
  • Atomx purge-and-trap: K-trap, nitrogen purge gas, 11 min purge at 40 mL/min, dry purge and desorb at 250 °C
  • MS parameters: EI 70 eV, source 325 °C, transfer line 230 °C, full scan 35–260 m/z

A nine-point calibration (0.5–200 ppb) plus internal/surrogate standards was performed. Method detection limits (MDLs) for 91 target analytes were determined by seven replicate measurements at 0.5 ppb, following EPA protocols.

Main Results and Discussion


Calibration results met US EPA Method 8260C criteria for all but two compounds, yielding a validation failure rate of 0.02%. Average response factors showed acceptable linearity, often requiring only a 1/x weighted quadratic fit for the few low-responding analytes. MDLs ranged from 0.03 to 0.28 ppb (with xylene isomers combined at 1 ppb), and relative standard deviations were generally below 10%. A 2 ppb standard chromatogram demonstrated clear separation and sensitivity for early-eluting gases and higher-boiling VOCs.

Benefits and Practical Applications


This integrated system provides:
  • Substantial reduction in helium usage due to the Helium Saver Module
  • Reliable purge-and-trap performance using lab-generated nitrogen
  • Full compliance with EPA Method 8260C for a broad VOC panel
  • Maintained sensitivity and precision at sub-ppb levels
  • Operational cost savings and enhanced sustainability

The approach is well suited for environmental laboratories, remediation projects, and industrial QA/QC workflows demanding high throughput and low operating expenses.

Future Trends and Potential Applications


Emerging directions include hydrogen carrier gas systems, further automation of sample preparation, and real-time VOC monitoring. Integration with software-driven workflows and miniaturized GC-MS platforms may broaden applications in field screening, on-line process analysis, and high-density monitoring networks.

Conclusion


The combination of the Thermo Fisher Helium Saver Module with the Teledyne Tekmar Atomx nitrogen purge feature delivers robust VOC analysis under US EPA Method 8260C. The system reduces helium consumption substantially while maintaining excellent calibration performance, sensitivity, and method detection limits, offering a sustainable and cost-effective solution for modern GC-MS laboratories.

Reference


Thermo Fisher Scientific and Teledyne Tekmar. Application Note 10441: Helium Conservation in Volatile Organic Compound Analysis Using US EPA Method 8260C. 2018.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
US EPA Method 8260 with the Atomx XYZ and the Thermo Scientific™ TRACE™ 1310 GC and ISQ™ MS
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1803 P+T VOC ENV US EPA Method 8260 with the Atomx XYZ and the Thermo Scientific™ TRACE™ 1310 GC and ISQ™ MS Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA…
Key words
temp, temppurge, purgebake, bakedesorb, desorbacetate, acetaterinse, rinseppb, ppbtert, tertsweep, sweepether, ethertime, timecondensate, condensateaverage, averagemethyl, methylbutyl
Validation of EPA Method 8260C on a Thermo Scientific TRACE 1300 Series GC with New Helium Saver Inlet
Validation of EPA Method 8260C on a Thermo Scientific TRACE 1300 Series GC with New Helium Saver Inlet Application Note P a g e |1 Abstract As helium continues to become more expensive, many labs are searching for more cost…
Key words
rsd, rsdavg, avgmdl, mdlacetate, acetateppb, ppbtert, terttable, tablecalibration, calibrationether, etherbutyl, butylbutylbenzene, butylbenzenecompound, compoundmethyl, methylsurr, surratomx
Using Hydrogen as An Alternative Carrier Gas for US EPA 8260
Using Hydrogen as An Alternative Carrier Gas for US EPA 8260
2013|Thermo Fisher Scientific|Applications
Using Hydrogen as An Alternative Carrier Gas for US EPA 8260 Application Note Abstract Due to regional shortages and increasing costs of helium, the preferred carrier gas in gas chromatography, alternative carrier gas options are increasing in demand. This study…
Key words
temp, tempbake, bakerinse, rinsepurge, purgeacetate, acetatetert, tertsweep, sweepdesorb, desorbcondensate, condensatetime, timetekmar, tekmarteledyne, teledyneether, etherbutyl, butylbutylbenzene
US EPA Method 524.4 using the Teledyne Tekmar Atomx XYZ and Thermo Scientific™ ISQ™ 7000 Mass Spectrometry (MS) System Coupled with a Thermo Scientific™ TRACE™ 1310 Gas Chromatograph (GC)
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN2002 P+T VOC ENV US EPA Method 524.4 using the Teledyne Tekmar Atomx XYZ and Thermo Scientific™ ISQ™ 7000 Mass Spectrometry (MS) System Coupled with a Thermo Scientific™ TRACE™ 1310 Gas Chromatograph (GC) Amy Nutter, Applications…
Key words
temp, tempbake, bakepurge, purgerinse, rinseavg, avgsweep, sweeptert, tertatomx, atomxether, etherdesorb, desorblpir, lpirupir, upirppb, ppbtime, timesurr
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike