Ultra Fast Determination of VOCs in Packaging Materials by Headspace GC

Applications | 2007 | Thermo Fisher ScientificInstrumentation
GC, HeadSpace
Industries
Materials Testing
Manufacturer
Thermo Fisher Scientific

Summary

Importance of the Topic


Residual volatile organic compounds (VOCs) in packaging materials can pose health and quality risks to consumer products. Rapid, reliable monitoring of these compounds during high-speed production is crucial to ensure consistent product safety and regulatory compliance. Traditional headspace GC methods require 30–60 minutes per analysis, creating a bottleneck in modern packaging lines.

Study Objectives and Overview


This study aims to develop and validate an ultra-fast headspace GC method for on-line determination of residual solvents in flexible packaging materials. By integrating a high-speed GC module and automated headspace sampling, the goal is to reduce analysis time to 1–2 minutes while maintaining precision and accuracy.

Methodology


Sample preparation involved cutting 100 cm² slices of uncoated, coated or laminated films and placing them into 20 mL headspace vials. Flexographic ink samples (20 µL) were also analyzed. Calibration used evaporated standard mixtures at the 10 µg level. Up to six vials were incubated simultaneously to overlap sample heating and data acquisition, eliminating idle time between runs. Incubation conditions ranged from 5 to 15 minutes at 125 °C. GC temperature programs were optimized for fast ramping (e.g., 35 °C hold, 1.5 °C/s to 200 °C) with split injection ratios of 1:250.

Instrumentation Used


  • Thermo Scientific TRACE GC Ultra with Ultra Fast Module (UFM) enabling resistive column heating at rates up to 20 °C/s and rapid cool-down cycles (~1 min from 350 °C to 50 °C).
  • Capillary columns: 5 m × 0.1 mm i.d., 0.2 µm TR-WAX; and 10 m × 0.1 mm i.d., 0.1 µm TR-1701.
  • Fast flame ionization detector (FID).
  • Thermo Scientific TriPlus static headspace autosampler for precise vial and syringe control.

Key Results and Discussion


The ultra-fast GC method achieved a 10–20× speed increase versus conventional systems, completing VOC separations in 1–2 minutes. Calibration curves for ethanol, acetone, ethyl acetate, MEK and toluene exhibited excellent linearity (R² > 0.998). Repeatability tests (n=5) showed retention time standard deviations below 0.03 s and peak area RSDs under 2.1 %. Detection and quantification limits reached as low as 0.1 mg/m² on film surfaces. Analyses on metallized polypropylene and polyester/polyethylene films, as well as red flexographic inks, demonstrated reliable identification and quantification of multiple solvent species within production-compatible timeframes.

Benefits and Practical Applications


  • Total analysis cycle of approximately five minutes per sample aligns with high-speed packaging workflows.
  • Simultaneous incubation and rapid GC cycles maximize laboratory throughput.
  • High precision and low detection limits support quality control in food, pharmaceutical and industrial packaging sectors.

Future Trends and Potential Applications


Further integration of ultra-fast GC with mass spectrometric detection could enhance compound identification. Miniaturized headspace modules and advanced automation will drive real-time inline monitoring. Expanding the approach to solid and liquid samples beyond packaging materials may broaden its application in environmental and pharmaceutical analytics.

Conclusion


The combination of headspace autosampling and ultra-fast GC provides a powerful, high-throughput solution for monitoring residual solvents in packaging materials. The validated method meets rigorous precision, speed and sensitivity requirements, enabling real-time quality assurance in modern production environments.

References


  1. P. Magni, R. Facchetti, D. Cavagnino and S. Trestianu, Proceedings of the 25th International Symposium on Capillary Chromatography, KNL05, Riva del Garda, Italy, May 13–17, 2002.
  2. F. Munari, S. Pelagatti and P. Mapelli, Enhanced Performances in Liquid and Gas Injections in Capillary GC through a Robotic, Versatile Autosampler, I11, Riva del Garda, Italy, May 31–June 4, 2004.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Ultra Fast Determination of VOCs in Packaging Materials by Headspace GC
Application Note: 10132 Ultra Fast Determination of VOCs in Packaging Materials by Headspace GC Thermo Fisher Scientific Inc., Milan, Italy Introduction Key Words • TRACE GC Ultra • TRACE TR-1701 Column • Volatile Organic Compounds • Ultra Fast GC •…
Key words
packaging, packagingacetate, acetateink, inkethanol, ethanolethyl, ethylacetone, acetonemek, mekflexographic, flexographicultra, ultraethoxypropanol, ethoxypropanolcomponent, componentsolvents, solventspeak, peakheadspace, headspacevocs
An Easy Solution to Monitor Printing Residual Solvents in Packaging Material using an Agilent 7820 GC system and an Agilent 7697A Headspace Sampler
An Easy Solution to Monitor Printing Residual Solvents in Packaging Material using an Agilent 7820 GC system and an Agilent 7697A Headspace Sampler Application Note Author Abstract Suli Zhao Printing inks have been used extensively to enhance the visual appearance…
Key words
printing, printingacetate, acetateidentitfication, identitficationsolvents, solventsethoxypropanol, ethoxypropanolresidual, residualopenlab, openlabpackaging, packagingmek, mekisobutanol, isobutanolcds, cdsheadspace, headspacepropanol, propanolbutanol, butanolpropyl
Rapid qualitative and quantitative analysis of residual solvents in food packaging by static headspace coupled to GC-FID/MS
APPLICATION NOTE 10689 Rapid qualitative and quantitative analysis of residual solvents in food packaging by static headspace coupled to GC-FID/MS Authors Giulia Riccardino and Cristian Cojocariu Thermo Fisher Scientific, Runcorn, UK Keywords Residual solvents, flexible food packaging, food safety, valve…
Key words
mhe, mhepackaging, packagingresidual, residualacetate, acetatefid, fidacetaldehyde, acetaldehydesolvent, solventsolvents, solventsethyl, ethylfood, foodsalami, salamivial, vialmaterials, materialspressure, pressurestandard
Unattended Quantitative Determination of VOCs in Food Packaging Samples Using a Robotic Sampler for Standard Addition and Analysis
Unattended Quantitative Determination of VOCs in Food Packaging Samples Using a Robotic Sampler for Standard Addition and Analysis S. Pelagatti1, S. Gemme1, A. Caruso1, F. Pigozzo1, M. Santoro1, E. Phillips2 Thermo Fisher Scientific, Milan, Italy1; Thermo Fisher Scientific, Austin, Texas,…
Key words
acetate, acetatearea, areaheadspace, headspaceethyl, ethylbeen, beenmeans, meansperformed, performedaddition, additiontoluene, toluenecellosolve, cellosolvepackaging, packagingbutyl, butylethanol, ethanolhas, hasstandard
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike