Unambiguous Pesticide Identification in Cigarette Tobacco
Applications | 2014 | LECOInstrumentation
Identification of pesticide residues in cigarette tobacco is critical for ensuring product safety and regulatory compliance. Tobacco presents a challenging matrix due to its complex mixture of natural compounds and additives. High-resolution time-of-flight mass spectrometry (HR-TOF MS) offers the resolution and mass accuracy required to distinguish isobaric species and coeluting analytes, minimizing false positives and negatives.
The primary aim was to demonstrate unambiguous detection and identification of multiple pesticide residues in cigarette tobacco using limited sample cleanup. Specific goals included:
Cigarette tobacco from a commercial brand was extracted using a QuEChERS protocol. The extract was split into two portions: one for incurred pesticide screening and one fortified with a standard pesticide mixture. Deconvolution software processed full-scan HR-TOF data to detect, deconvolve, and match accurate masses against a spectral library with high confidence.
Resolution at 50 000 FWHM was required to separate the cyanazine molecular ion (m/z 240.0885) from a metolachlor fragment (m/z 240.0964), achieving mass accuracies of 0.083 ppm and 0.30 ppm, respectively. At 25 000 resolution, these ions remained convoluted, risking false negatives. Dichlorvos base peak (m/z 109.0049) could only be extracted cleanly at resolutions above 10 000. Deconvolution software successfully differentiated lindane from abundant matrix components such as neophytadiene, yielding library match scores above 800. Comprehensive full-scan acquisition also allowed simultaneous identification of 37 EPA-approved pesticides and multiple tobacco-derived compounds in a single analysis.
The integration of HR-TOF MS into routine QA/QC workflows will expand to other complex food, environmental, and biological matrices. Advances in software deconvolution and spectral libraries will enhance non-target screening. Coupling these instruments with automated sample handling and artificial intelligence–driven data interpretation promises faster, more reliable multi-residue analysis.
This application demonstrates that HR-TOF MS with accurate mass deconvolution provides unmatched confidence in identifying pesticide residues in complex tobacco matrices. Its high resolution and mass accuracy deliver clear separation of isobaric interferences, efficient screening of non-target compounds, and robust confirmation of targets with minimal preparation requirements.
GC/MSD, GC/HRMS, GC/TOF
IndustriesFood & Agriculture
ManufacturerAgilent Technologies, LECO
Summary
Importance of the Topic
Identification of pesticide residues in cigarette tobacco is critical for ensuring product safety and regulatory compliance. Tobacco presents a challenging matrix due to its complex mixture of natural compounds and additives. High-resolution time-of-flight mass spectrometry (HR-TOF MS) offers the resolution and mass accuracy required to distinguish isobaric species and coeluting analytes, minimizing false positives and negatives.
Objectives and Overview of the Article
The primary aim was to demonstrate unambiguous detection and identification of multiple pesticide residues in cigarette tobacco using limited sample cleanup. Specific goals included:
- Evaluating the capability of HR-TOF MS to resolve isobaric interferences such as cyanazine versus metolachlor.
- Assessing the isolation of low-mass diagnostic ions (e.g., dichlorvos) from matrix background.
- Illustrating non-target screening alongside targeted analysis in a single run.
Methodology
Cigarette tobacco from a commercial brand was extracted using a QuEChERS protocol. The extract was split into two portions: one for incurred pesticide screening and one fortified with a standard pesticide mixture. Deconvolution software processed full-scan HR-TOF data to detect, deconvolve, and match accurate masses against a spectral library with high confidence.
Used Instrumentation
- Gas Chromatograph: Agilent 7890 with 7693 Autosampler; injection 1 µL splitless at 250 °C; helium carrier gas at 1.0 mL/min.
- Column: Rxi-5Sil MS, 30 m × 0.25 mm ID × 0.25 µm film.
- Temperature Program: 60 °C (1 min) → 160 °C @ 25 °C/min → 190 °C @ 2.5 °C/min → 205 °C @ 3.0 °C/min → 220 °C @ 5.0 °C/min → 285 °C @ 17 °C/min (10 min).
- Mass Spectrometer: LECO Pegasus GC-HRT; transfer line 275 °C; EI source at 250 °C; mass range m/z 45–550; acquisition rate 8 spectra/s.
- Resolution Modes: R=25 000 (standard) and R=50 000 (ultra-high).
Main Results and Discussion
Resolution at 50 000 FWHM was required to separate the cyanazine molecular ion (m/z 240.0885) from a metolachlor fragment (m/z 240.0964), achieving mass accuracies of 0.083 ppm and 0.30 ppm, respectively. At 25 000 resolution, these ions remained convoluted, risking false negatives. Dichlorvos base peak (m/z 109.0049) could only be extracted cleanly at resolutions above 10 000. Deconvolution software successfully differentiated lindane from abundant matrix components such as neophytadiene, yielding library match scores above 800. Comprehensive full-scan acquisition also allowed simultaneous identification of 37 EPA-approved pesticides and multiple tobacco-derived compounds in a single analysis.
Benefits and Practical Applications of the Method
- High selectivity through ultra-high resolution prevents isobaric overlap and coelution issues.
- Accurate mass measurements enable definitive formula determination and compound confirmation.
- Comprehensive screening allows non-targeted discovery of additives and natural constituents alongside targeted pesticides.
- Minimal sample cleanup reduces preparation time and potential analyte loss.
Future Trends and Opportunities
The integration of HR-TOF MS into routine QA/QC workflows will expand to other complex food, environmental, and biological matrices. Advances in software deconvolution and spectral libraries will enhance non-target screening. Coupling these instruments with automated sample handling and artificial intelligence–driven data interpretation promises faster, more reliable multi-residue analysis.
Conclusion
This application demonstrates that HR-TOF MS with accurate mass deconvolution provides unmatched confidence in identifying pesticide residues in complex tobacco matrices. Its high resolution and mass accuracy deliver clear separation of isobaric interferences, efficient screening of non-target compounds, and robust confirmation of targets with minimal preparation requirements.
Reference
- Misselwitz, M.; Cochran, J.; Kowalski, J. Evaluation of dispersive and cartridge solid phase extraction (SPE) cleanup for multiresidue pesticides in QuEChERS extracts of finished tobacco using GC-TOFMS. Restek Technical Library FFAN-1823.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
LECO PEGASUS® GC-HRT+
2020|LECO|Brochures and specifications
® PEGASUS GC-HRT EMPOWERING RESULTS + LECO’s PEGASUS GC-HRT + The ultimate analytical instrument for increased confidence in the identification of components in your most complex samples. For complete patent information, see specification sheet 209-212-001. Build Your Confidence Further Industry…
Key words
pegasus, pegasushrt, hrthrd, hrddeconvolution, deconvolutionefp, efpffp, ffpion, ionyour, yourkadas, kadashigh, highdefect, defectspectral, spectralgridless, gridlessgcxgc, gcxgcsensitivity
LECO PEGASUS® GC-HRT+
2019|LECO|Brochures and specifications
® PEGASUS GC-HRT EMPOWERING RESULTS + LECO’s PEGASUS GC-HRT + The ultimate analytical instrument for increased confidence in the identification of components in your most complex samples. For complete patent information, see specification sheet 209-212-001. Build Your Confidence Further Industry…
Key words
pegasus, pegasushrt, hrthrd, hrddeconvolution, deconvolutionefp, efpffp, ffpion, ionyour, yourkadas, kadashigh, highdefect, defectspectral, spectralgridless, gridlesspatented, patentedgcxgc
Comprehensive Characterization of Green Leaf Tobacco Extracts Using GC-HRT
2014|Agilent Technologies|Applications
Comprehensive Characterization of Green Leaf Tobacco Extracts Using GC-HRT LECO Corporation; Saint Joseph, Michigan USA 1. Introduction This application note highlights the attributes of High Resolution Time-of-Flight MS (HRT) which provide an extra level of confidence for challenging plant extract…
Key words
hrt, hrthrd, hrdsimilarity, similarityscience, sciencecoeluting, coelutinglife, lifesubstantiates, substantiatesethylmethylmaleimide, ethylmethylmaleimidemass, masscembrene, cembrenedelivering, deliveringamyrin, amyrinneophytadiene, neophytadienecommercially, commerciallychemical
Identifying Disinfection Byproducts in Treated Water
2015|Agilent Technologies|Applications
Identifying Disinfection Byproducts in Treated Water Jonathan D. Byer, LECO Corporation; Saint Joseph, Michigan USA Susan D. Richardson, University of South Carolina; Columbia, South Carolina USA 1. Introduction ® LECO's Pegasus GC-HRT was used for the identification of "known unknowns"…
Key words
hrt, hrtpegasus, pegasushrd, hrdleco, lecoionization, ionizationchemical, chemicalmass, massscience, sciencelife, lifeidentification, identificationbramd, bramdpyridinamine, pyridinamineunknowns, unknownsdelivering, deliveringdibromo