Peak Shape Problems: Sensitivity Loss

We, 1.10.2025
| Original article from: Phenomenex
Learn how to identify and fix sensitivity loss in GC. Discover key causes, from column contamination to injector leaks, and apply remedies for reliable, accurate results.
<p><strong>Phenomenex:</strong> Peak Shape Problems: Sensitivity Loss</p><p>&nbsp;</p>

Phenomenex: Peak Shape Problems: Sensitivity Loss

 

When sensitivity loss occurs, it can significantly impact the accuracy and reliability of analytical results. Several factors can contribute to sensitivity loss, such as column contamination, sample matrix effects, injector and detector issues, or improper sample preparation. This technical note aims to guide analysts in identifying the common causes behind sensitivity loss and implementing effective strategies to address the issue.

Symptom: Decreased Response

Some or all peaks are displaying decreased response.

Possible Cause  /  Suggested Remedy

  • Contamination of column and/ or liner can lead to loss of sensitivity for active compounds.: 
    • Clean liner.
    • Bake out the column. Solvent rinse or replace the column.
  • Injector leaks reduce the peak height of the most volatile components of a sample.: 
    • Find and fix any leaks
  • If initial column temperature is too high for splitless injection, this prevents refocusing of sample. This affects the more volatile components most.: 
    • Lower temperature below the boiling point of the solvent. Decrease the initial column temperature, or use a less volatile solvent
  • Inlet discrimination: injector temperature is too low. Later eluting and less volatile compounds have low response.: 
    • Increase the injection temperature or use on-column injection with direct connect liner
  • Issues with the sample: 
    • Check the sample concentration, any sample preparation procedures, and shelf life. Prepare a fresh sample ensuring the proper concentration

In conclusion, sensitivity loss in GC is characterized by decreased response in one or more peaks, which can hinder accurate and reliable analysis. It is crucial to maintain a clean column and liner, address any injector leaks, optimize column conditions, and ensure proper sample preparation. Remember to regularly perform maintenance tasks and quality checks to ensure the ongoing performance of the GC system.

By implementing these countermeasures, you can overcome sensitivity loss and achieve more accurate and reliable analytical results.

Phenomenex
LinkedIn Logo
 

Related content

Run Smarter, Go Greener: Smart Energy Saving In Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Gas Chromatograph Mass Spectrometer GCMS-TQ RX Series

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Shimadzu
Industries
Other

Hydrogen Carrier Gas Method Translation in Comprehensive Two- Dimensional Gas Chromatography for Sustainable Nontargeted Analysis

Scientific articles
| 2026 | William & Mary
Instrumentation
GCxGC, GC/MSD, GC/MS/MS, GC/TOF, GC/HRMS
Manufacturer
LECO
Industries
Other

Novel data processing workflow including in-silico derivatization for integrated CI & EI MS data acquired with the GC-ecTOFinstrument

Posters
| 2026 | Bruker (ASMS)
Instrumentation
GC/MS/MS, GC/MSD, GC/TOF, GC/HRMS
Manufacturer
Bruker
Industries
Metabolomics

Is Your Next Sip Safe? A Simple, Rapid Method for Measuring volatile PFAS in Juices.

Posters
| 2026 | Shimadzu (ASMS)
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, HeadSpace
Manufacturer
Shimadzu
Industries
Food & Agriculture
 

Related articles

Peak Shape Problems: Broad Solvent Peaks/Fronts
Article | Academy

Peak Shape Problems: Broad Solvent Peaks/Fronts

Learn how to troubleshoot and fix broad solvent peaks/fronts in GC analysis with practical tips from Phenomenex to improve peak shape and accuracy.
Phenomenex
tag
share
more
Peak Shape Problems: No Peaks
Article | Academy

Peak Shape Problems: No Peaks

Missing peaks in GC runs? This technical tip reviews key causes like syringe, inlet, detector, or column issues, with practical solutions to restore results.
Phenomenex
tag
share
more
Water Injections in GC: Is it Safe?
Article | Academy

Water Injections in GC: Is it Safe?

Injecting water into a GC system can cause backflash, poor vaporization, detector issues, and stationary phase degradation. Learn how to minimize these risks with the right column and injection setup.
Phenomenex
tag
share
more
GC Columns - Troubleshooting & Maintenance
Article | Academy

GC Columns - Troubleshooting & Maintenance

A practical guide to GC column troubleshooting. Learn how to identify and resolve peak tailing, loss of resolution, ghost peaks, baseline drift, sensitivity issues, and retention time shifts.
Phenomenex
tag
share
more
 

Related content

Run Smarter, Go Greener: Smart Energy Saving In Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Gas Chromatograph Mass Spectrometer GCMS-TQ RX Series

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Shimadzu
Industries
Other

Hydrogen Carrier Gas Method Translation in Comprehensive Two- Dimensional Gas Chromatography for Sustainable Nontargeted Analysis

Scientific articles
| 2026 | William & Mary
Instrumentation
GCxGC, GC/MSD, GC/MS/MS, GC/TOF, GC/HRMS
Manufacturer
LECO
Industries
Other

Novel data processing workflow including in-silico derivatization for integrated CI & EI MS data acquired with the GC-ecTOFinstrument

Posters
| 2026 | Bruker (ASMS)
Instrumentation
GC/MS/MS, GC/MSD, GC/TOF, GC/HRMS
Manufacturer
Bruker
Industries
Metabolomics

Is Your Next Sip Safe? A Simple, Rapid Method for Measuring volatile PFAS in Juices.

Posters
| 2026 | Shimadzu (ASMS)
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, HeadSpace
Manufacturer
Shimadzu
Industries
Food & Agriculture
 

Related articles

Peak Shape Problems: Broad Solvent Peaks/Fronts
Article | Academy

Peak Shape Problems: Broad Solvent Peaks/Fronts

Learn how to troubleshoot and fix broad solvent peaks/fronts in GC analysis with practical tips from Phenomenex to improve peak shape and accuracy.
Phenomenex
tag
share
more
Peak Shape Problems: No Peaks
Article | Academy

Peak Shape Problems: No Peaks

Missing peaks in GC runs? This technical tip reviews key causes like syringe, inlet, detector, or column issues, with practical solutions to restore results.
Phenomenex
tag
share
more
Water Injections in GC: Is it Safe?
Article | Academy

Water Injections in GC: Is it Safe?

Injecting water into a GC system can cause backflash, poor vaporization, detector issues, and stationary phase degradation. Learn how to minimize these risks with the right column and injection setup.
Phenomenex
tag
share
more
GC Columns - Troubleshooting & Maintenance
Article | Academy

GC Columns - Troubleshooting & Maintenance

A practical guide to GC column troubleshooting. Learn how to identify and resolve peak tailing, loss of resolution, ghost peaks, baseline drift, sensitivity issues, and retention time shifts.
Phenomenex
tag
share
more
 

Related content

Run Smarter, Go Greener: Smart Energy Saving In Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Gas Chromatograph Mass Spectrometer GCMS-TQ RX Series

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Shimadzu
Industries
Other

Hydrogen Carrier Gas Method Translation in Comprehensive Two- Dimensional Gas Chromatography for Sustainable Nontargeted Analysis

Scientific articles
| 2026 | William & Mary
Instrumentation
GCxGC, GC/MSD, GC/MS/MS, GC/TOF, GC/HRMS
Manufacturer
LECO
Industries
Other

Novel data processing workflow including in-silico derivatization for integrated CI & EI MS data acquired with the GC-ecTOFinstrument

Posters
| 2026 | Bruker (ASMS)
Instrumentation
GC/MS/MS, GC/MSD, GC/TOF, GC/HRMS
Manufacturer
Bruker
Industries
Metabolomics

Is Your Next Sip Safe? A Simple, Rapid Method for Measuring volatile PFAS in Juices.

Posters
| 2026 | Shimadzu (ASMS)
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, HeadSpace
Manufacturer
Shimadzu
Industries
Food & Agriculture
 

Related articles

Peak Shape Problems: Broad Solvent Peaks/Fronts
Article | Academy

Peak Shape Problems: Broad Solvent Peaks/Fronts

Learn how to troubleshoot and fix broad solvent peaks/fronts in GC analysis with practical tips from Phenomenex to improve peak shape and accuracy.
Phenomenex
tag
share
more
Peak Shape Problems: No Peaks
Article | Academy

Peak Shape Problems: No Peaks

Missing peaks in GC runs? This technical tip reviews key causes like syringe, inlet, detector, or column issues, with practical solutions to restore results.
Phenomenex
tag
share
more
Water Injections in GC: Is it Safe?
Article | Academy

Water Injections in GC: Is it Safe?

Injecting water into a GC system can cause backflash, poor vaporization, detector issues, and stationary phase degradation. Learn how to minimize these risks with the right column and injection setup.
Phenomenex
tag
share
more
GC Columns - Troubleshooting & Maintenance
Article | Academy

GC Columns - Troubleshooting & Maintenance

A practical guide to GC column troubleshooting. Learn how to identify and resolve peak tailing, loss of resolution, ghost peaks, baseline drift, sensitivity issues, and retention time shifts.
Phenomenex
tag
share
more
 

Related content

Run Smarter, Go Greener: Smart Energy Saving In Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Gas Chromatograph Mass Spectrometer GCMS-TQ RX Series

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Shimadzu
Industries
Other

Hydrogen Carrier Gas Method Translation in Comprehensive Two- Dimensional Gas Chromatography for Sustainable Nontargeted Analysis

Scientific articles
| 2026 | William & Mary
Instrumentation
GCxGC, GC/MSD, GC/MS/MS, GC/TOF, GC/HRMS
Manufacturer
LECO
Industries
Other

Novel data processing workflow including in-silico derivatization for integrated CI & EI MS data acquired with the GC-ecTOFinstrument

Posters
| 2026 | Bruker (ASMS)
Instrumentation
GC/MS/MS, GC/MSD, GC/TOF, GC/HRMS
Manufacturer
Bruker
Industries
Metabolomics

Is Your Next Sip Safe? A Simple, Rapid Method for Measuring volatile PFAS in Juices.

Posters
| 2026 | Shimadzu (ASMS)
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, HeadSpace
Manufacturer
Shimadzu
Industries
Food & Agriculture
 

Related articles

Peak Shape Problems: Broad Solvent Peaks/Fronts
Article | Academy

Peak Shape Problems: Broad Solvent Peaks/Fronts

Learn how to troubleshoot and fix broad solvent peaks/fronts in GC analysis with practical tips from Phenomenex to improve peak shape and accuracy.
Phenomenex
tag
share
more
Peak Shape Problems: No Peaks
Article | Academy

Peak Shape Problems: No Peaks

Missing peaks in GC runs? This technical tip reviews key causes like syringe, inlet, detector, or column issues, with practical solutions to restore results.
Phenomenex
tag
share
more
Water Injections in GC: Is it Safe?
Article | Academy

Water Injections in GC: Is it Safe?

Injecting water into a GC system can cause backflash, poor vaporization, detector issues, and stationary phase degradation. Learn how to minimize these risks with the right column and injection setup.
Phenomenex
tag
share
more
GC Columns - Troubleshooting & Maintenance
Article | Academy

GC Columns - Troubleshooting & Maintenance

A practical guide to GC column troubleshooting. Learn how to identify and resolve peak tailing, loss of resolution, ghost peaks, baseline drift, sensitivity issues, and retention time shifts.
Phenomenex
tag
share
more
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike