GC & GC/MS Method Development Quick Reference Guide

Th, 22.4.2021
| Original article from: ChromSolutions
This GC & GC/MS guide offers practical advice which we hope will assist users avoid many common problems encountered from the results of poor method development, translation or optimisation.
<p>Unsplash/Julia Koblitz: GC &amp; GC/MS Method Development Quick Reference Guide</p>

Unsplash/Julia Koblitz: GC & GC/MS Method Development Quick Reference Guide

This can range from inaccurate results to instruments with significant downtime and high maintenance costs. The objective being to develop automated reliable identification and quantitative results that are fit for purpose.

C3 -C15 Hydrocarbons Method Translation Example.jpgC3 -C15 Hydrocarbons Method Translation Example.jpg

Before Commencement

  • Ensure the new or existing instrument is configured and installed with the correct supplies, gases, column(s) and injector(s), valve(s), liner(s), detector, interfaces and software options for detection of the target analytes.

  • The instrument has been installed to manufactures specification and all maintenance logs/service history are current up to date.

Initial Testing

  • Ensure data capture method has sufficient speed to comply with accuracy requirement (8 points min across a peak for 5-10% repeatability).

  • Adjust column flows and oven temperature profile to achieve the required sample throughput and separation for a mid range standard.

  • Optimise detector conditions for best response and minimal interference.

  • Optimise injector parameters for best response, chromatographic resolution and minimal peak distortion.

  • Apply data processing method to obtain the correct integration parameters and identification windows for the chromatography.

  • Adjust auto sampler settings and to obtain optimal repeatability (min 6 injections) on a mid range standard.

  • If the method criteria cannot be reached for the mid range standard consider increasing data capture speed and/or the addition of internal standard(s).

SQC & Sample Testing

  • Inject a matrix matched standard or a sample (min 6 injections) to check identification, integration and repeatability.

  • If the chromatograms/results differ significantly from the initial testing or than expected/acceptable, consider a more appropriate matrix elimination strategy.

Carry Over Testing

  • Inject high std +20% followed by a blank and measure carry over and carry over re-peatability. Also ensuring overloaded peak shapes is kept to a minimum.

  • Modify autosampler parameters and wash solvents/purge gas to minimise carry over.

  • Modify data processing method to ensure correct integration of high level standard and blanks.

  • Run a sample or independent QC standard six times with new auto sampler conditions to compare chromatography and repeatability data to the initial SQC and sample testing. If not comparable re-adjust modifications made in carry over tests.

Calibration/QC/Validation Considerations

  • Initial Calibration for external/internal standard and mole %/normalised %.

  • Run calibration range from low std 3*limit of detection and top std max level +20% (ideally 7 point). Run 3 points at each level, 6 points at LOQ, 6 points at mid point and 6 points at highest level. Plot calibration curve and assess linearity criteria and calculate % rsd at levels containing 6 points.

  • If any adjustments are made to the method the sample or independent QC/sample spikes need to be recalculated or re-injected.

  • Devise a QC/validation protocol with independent QC and calibrations to ensure method robustness and the control of the frequency of the maintenance procedures. Typical protocols include injection of low and high level standards/SQC’s/sample spikes measuring peak areas, retention times and peak shapes. In addition baseline noise and offsets from chromatogram start and end are typically measured.

Conclusion

If these guidelines are adhered to you will avoid undesirable outcomes and have a fully utilized robust method providing accurate and reliable results fit for purpose.

ChromSolutions Ltd

What we offer at ChromSolutions is our wealth of experience in analytical instrument sales and support (over 110 years distributed through the members of our company). We can help you from defining your requirements to the implementation of a robust gas analytical method fit for purpose.

Many webinars focused on the development and optimization of GC and GCMS methods can also be found in the Webinars section.
ChromSolutions
LinkedIn Logo
 

Related content

Alcohol Determination of Sanitizer Gel Using Nexis GC-2060

Applications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Pharma & Biopharma

Sensitive and Reproducible Determination of Nitrosamines by GC/MSD

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/SQ
Manufacturer
Agilent Technologies
Industries
Environmental

High-Throughput BTEX Analysis in Nail Products by SPME and GC/TQ

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, SPME
Manufacturer
Agilent Technologies
Industries
Materials Testing

Accurate multi-component blast furnace gas analysis maximizes iron production and minimizes coke consumption

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD
Manufacturer
Thermo Fisher Scientific
Industries
Energy & Chemicals

Gas Chromatograph Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other
 

Related articles

Broadband Laser-Based Infrared Detector for Gas Chromatography
Article | Scientific article

Broadband Laser-Based Infrared Detector for Gas Chromatography

In the article, the researchers described using Cantilever-enhanced photoacoustic spectroscopy coupled with gas chromatography to quantitatively analyze a mixture of alcohols in a quasi-online manner.
LabRulez
tag
share
more
Introduction of a small volume ethyl acetate based liquid-liquid extraction procedure for analysis of polycyclic aromatic hydrocarbons in wastewater by atmospheric pressure gas chromatography-mass spectrometry and evaluation of method greenness
Scientific article | Environment

Introduction of a small volume ethyl acetate based liquid-liquid extraction procedure for analysis of polycyclic aromatic hydrocarbons in wastewater by atmospheric pressure gas chromatography-mass spectrometry and evaluation of method greenness

The goal of this study is to evaluate and validate a rapid, eco-friendly SVEA-LLE method with APGC-MS/MS for the sensitive and accurate analysis of 17 PAHs in water and wastewater.
LabRulez
tag
share
more
News from LabRulezGCMS Library - Week 12, 2025
Article | Application

News from LabRulezGCMS Library - Week 12, 2025

This week we bring you application notes by Agilent Technologies, Shimadzu and Thermo Fisher Scientific
LabRulez
tag
share
more
GC & GC/MS Relocation Quick Reference Guide
Article | Academy

GC & GC/MS Relocation Quick Reference Guide

Relocating GC or a GC/MS is a meticulous task that requires careful planning and execution to ensure the integrity and functionality of the instruments.
ChromSolutions
tag
share
more
 

Related content

Alcohol Determination of Sanitizer Gel Using Nexis GC-2060

Applications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Pharma & Biopharma

Sensitive and Reproducible Determination of Nitrosamines by GC/MSD

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/SQ
Manufacturer
Agilent Technologies
Industries
Environmental

High-Throughput BTEX Analysis in Nail Products by SPME and GC/TQ

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, SPME
Manufacturer
Agilent Technologies
Industries
Materials Testing

Accurate multi-component blast furnace gas analysis maximizes iron production and minimizes coke consumption

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD
Manufacturer
Thermo Fisher Scientific
Industries
Energy & Chemicals

Gas Chromatograph Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other
 

Related articles

Broadband Laser-Based Infrared Detector for Gas Chromatography
Article | Scientific article

Broadband Laser-Based Infrared Detector for Gas Chromatography

In the article, the researchers described using Cantilever-enhanced photoacoustic spectroscopy coupled with gas chromatography to quantitatively analyze a mixture of alcohols in a quasi-online manner.
LabRulez
tag
share
more
Introduction of a small volume ethyl acetate based liquid-liquid extraction procedure for analysis of polycyclic aromatic hydrocarbons in wastewater by atmospheric pressure gas chromatography-mass spectrometry and evaluation of method greenness
Scientific article | Environment

Introduction of a small volume ethyl acetate based liquid-liquid extraction procedure for analysis of polycyclic aromatic hydrocarbons in wastewater by atmospheric pressure gas chromatography-mass spectrometry and evaluation of method greenness

The goal of this study is to evaluate and validate a rapid, eco-friendly SVEA-LLE method with APGC-MS/MS for the sensitive and accurate analysis of 17 PAHs in water and wastewater.
LabRulez
tag
share
more
News from LabRulezGCMS Library - Week 12, 2025
Article | Application

News from LabRulezGCMS Library - Week 12, 2025

This week we bring you application notes by Agilent Technologies, Shimadzu and Thermo Fisher Scientific
LabRulez
tag
share
more
GC & GC/MS Relocation Quick Reference Guide
Article | Academy

GC & GC/MS Relocation Quick Reference Guide

Relocating GC or a GC/MS is a meticulous task that requires careful planning and execution to ensure the integrity and functionality of the instruments.
ChromSolutions
tag
share
more
 

Related content

Alcohol Determination of Sanitizer Gel Using Nexis GC-2060

Applications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Pharma & Biopharma

Sensitive and Reproducible Determination of Nitrosamines by GC/MSD

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/SQ
Manufacturer
Agilent Technologies
Industries
Environmental

High-Throughput BTEX Analysis in Nail Products by SPME and GC/TQ

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, SPME
Manufacturer
Agilent Technologies
Industries
Materials Testing

Accurate multi-component blast furnace gas analysis maximizes iron production and minimizes coke consumption

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD
Manufacturer
Thermo Fisher Scientific
Industries
Energy & Chemicals

Gas Chromatograph Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other
 

Related articles

Broadband Laser-Based Infrared Detector for Gas Chromatography
Article | Scientific article

Broadband Laser-Based Infrared Detector for Gas Chromatography

In the article, the researchers described using Cantilever-enhanced photoacoustic spectroscopy coupled with gas chromatography to quantitatively analyze a mixture of alcohols in a quasi-online manner.
LabRulez
tag
share
more
Introduction of a small volume ethyl acetate based liquid-liquid extraction procedure for analysis of polycyclic aromatic hydrocarbons in wastewater by atmospheric pressure gas chromatography-mass spectrometry and evaluation of method greenness
Scientific article | Environment

Introduction of a small volume ethyl acetate based liquid-liquid extraction procedure for analysis of polycyclic aromatic hydrocarbons in wastewater by atmospheric pressure gas chromatography-mass spectrometry and evaluation of method greenness

The goal of this study is to evaluate and validate a rapid, eco-friendly SVEA-LLE method with APGC-MS/MS for the sensitive and accurate analysis of 17 PAHs in water and wastewater.
LabRulez
tag
share
more
News from LabRulezGCMS Library - Week 12, 2025
Article | Application

News from LabRulezGCMS Library - Week 12, 2025

This week we bring you application notes by Agilent Technologies, Shimadzu and Thermo Fisher Scientific
LabRulez
tag
share
more
GC & GC/MS Relocation Quick Reference Guide
Article | Academy

GC & GC/MS Relocation Quick Reference Guide

Relocating GC or a GC/MS is a meticulous task that requires careful planning and execution to ensure the integrity and functionality of the instruments.
ChromSolutions
tag
share
more
 

Related content

Alcohol Determination of Sanitizer Gel Using Nexis GC-2060

Applications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Pharma & Biopharma

Sensitive and Reproducible Determination of Nitrosamines by GC/MSD

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/SQ
Manufacturer
Agilent Technologies
Industries
Environmental

High-Throughput BTEX Analysis in Nail Products by SPME and GC/TQ

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, SPME
Manufacturer
Agilent Technologies
Industries
Materials Testing

Accurate multi-component blast furnace gas analysis maximizes iron production and minimizes coke consumption

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD
Manufacturer
Thermo Fisher Scientific
Industries
Energy & Chemicals

Gas Chromatograph Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other
 

Related articles

Broadband Laser-Based Infrared Detector for Gas Chromatography
Article | Scientific article

Broadband Laser-Based Infrared Detector for Gas Chromatography

In the article, the researchers described using Cantilever-enhanced photoacoustic spectroscopy coupled with gas chromatography to quantitatively analyze a mixture of alcohols in a quasi-online manner.
LabRulez
tag
share
more
Introduction of a small volume ethyl acetate based liquid-liquid extraction procedure for analysis of polycyclic aromatic hydrocarbons in wastewater by atmospheric pressure gas chromatography-mass spectrometry and evaluation of method greenness
Scientific article | Environment

Introduction of a small volume ethyl acetate based liquid-liquid extraction procedure for analysis of polycyclic aromatic hydrocarbons in wastewater by atmospheric pressure gas chromatography-mass spectrometry and evaluation of method greenness

The goal of this study is to evaluate and validate a rapid, eco-friendly SVEA-LLE method with APGC-MS/MS for the sensitive and accurate analysis of 17 PAHs in water and wastewater.
LabRulez
tag
share
more
News from LabRulezGCMS Library - Week 12, 2025
Article | Application

News from LabRulezGCMS Library - Week 12, 2025

This week we bring you application notes by Agilent Technologies, Shimadzu and Thermo Fisher Scientific
LabRulez
tag
share
more
GC & GC/MS Relocation Quick Reference Guide
Article | Academy

GC & GC/MS Relocation Quick Reference Guide

Relocating GC or a GC/MS is a meticulous task that requires careful planning and execution to ensure the integrity and functionality of the instruments.
ChromSolutions
tag
share
more
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike