(E)-2-Nonenal in Beer with the Atomx and GC/MS SIM Detection
Applications | 2015 | Teledyne LABSInstrumentation
Aged beer develops an off-flavor described as cardboard or papery, primarily due to the formation of (E)-2-nonenal. Monitoring this marker is critical for quality control in breweries and for ensuring product consistency over shelf life.
This study evaluates the performance of the Teledyne Tekmar Atomx automated VOC preparation system using a modified soil method to extract and concentrate (E)-2-nonenal from beer. Detection and quantification are performed with GC/MS in Selected Ion Monitoring mode across five commercial pale lager lots.
Sample Preparation
SIM total ion chromatograms of unspiked and spiked beer showed clear (E)-2-nonenal peaks with minimal interference. Quantified concentrations in five lots ranged from 0.082 to 0.463 ppb, demonstrating good sensitivity at mid-ppt levels. Triplicate analyses confirmed method reproducibility.
Advancements may include integration with headspace SPME, real-time monitoring via portable GC/MS, and expansion to other volatile off-flavor compounds. Data analytics and predictive models could further enhance shelf-life forecasting.
The Teledyne Tekmar Atomx with a modified soil method coupled to GC/MS SIM delivers a robust approach for detecting (E)-2-nonenal in beer at sub-ppb levels. Its automation and sensitivity make it a valuable tool for maintaining flavor stability in brewing operations.
GC/MSD, Purge and Trap, GC/SQ
IndustriesFood & Agriculture
ManufacturerTeledyne LABS
Summary
Significance of the Topic
Aged beer develops an off-flavor described as cardboard or papery, primarily due to the formation of (E)-2-nonenal. Monitoring this marker is critical for quality control in breweries and for ensuring product consistency over shelf life.
Objectives and Study Overview
This study evaluates the performance of the Teledyne Tekmar Atomx automated VOC preparation system using a modified soil method to extract and concentrate (E)-2-nonenal from beer. Detection and quantification are performed with GC/MS in Selected Ion Monitoring mode across five commercial pale lager lots.
Methodology and Instrumentation
Sample Preparation
- Beer samples: 10 mL degassed, spiked where indicated, with 6 g NaCl to enhance volatility.
- Standard solutions: (E)-2-nonenal in 5% ethanol at 0.27–5.4 ppb.
- Trap material: Tenax in Trap 1A.
- Purge/desorb cycle: 200 °C valve oven, 210 °C transfer line, purge at 70 °C for 10 min, desorb at 225 °C, bake at 230 °C.
- Flows: purge 25 mL/min, dry purge 50 mL/min, bake 200 mL/min.
- Column: Rtx-VMS (20 m×0.18 mm, 1 µm film), He constant flow at 0.9 mL/min.
- Inlet/transfer line: 230 °C, split ratio 60:1.
- Oven program: 95 °C initial, ramp 3.5 °C/min to 145 °C, then 15 °C/min to 240 °C, 3 min hold.
- MS: source 230 °C, quad 150 °C; SIM ions at m/z 55, 70, 83, 96, 111; dwell 100 ms.
Key Results and Discussion
SIM total ion chromatograms of unspiked and spiked beer showed clear (E)-2-nonenal peaks with minimal interference. Quantified concentrations in five lots ranged from 0.082 to 0.463 ppb, demonstrating good sensitivity at mid-ppt levels. Triplicate analyses confirmed method reproducibility.
Benefits and Practical Applications
- High sensitivity and selectivity for trace-level off-flavor markers.
- Automated sample prep reduces manual steps and variability.
- Applicability in routine quality control in breweries.
Future Trends and Possibilities
Advancements may include integration with headspace SPME, real-time monitoring via portable GC/MS, and expansion to other volatile off-flavor compounds. Data analytics and predictive models could further enhance shelf-life forecasting.
Conclusion
The Teledyne Tekmar Atomx with a modified soil method coupled to GC/MS SIM delivers a robust approach for detecting (E)-2-nonenal in beer at sub-ppb levels. Its automation and sensitivity make it a valuable tool for maintaining flavor stability in brewing operations.
Reference
- Scherer R., Wagner R., Kowalski C.H., Godoy H.T. (2010) Determination of (E)-2-nonenal in Brazilian beers using headspace SPME-GC-MS. Cienc. Tecnol. Aliment. 30(Suppl.1):161–165.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Italian Legislative Decree 152/2006 with the Tekmar Atomx XYZ and Thermo Scientific™ TRACE™ 1310 GC and ISQ™ 7000 MS with an Advanced Electron Ionization (AEI) Source
2019|Thermo Fisher Scientific|Applications
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1910 P+T VOC ENV Italian Legislative Decree 152/2006 with the Tekmar Atomx XYZ and Thermo Scientific™ TRACE™ 1310 GC and ISQ™ 7000 MS with an Advanced Electron Ionization (AEI) Source Amy Nutter, Applications Chemist; Teledyne Tekmar…
Key words
temp, tempbake, bakepurge, purgerinse, rinsesweep, sweepatomx, atomxdesorb, desorbxyz, xyztime, timetekmar, tekmarteledyne, teledynevariable, variableflow, flowoff, offcondensate
USEPA Method 524.4 Using the Teledyne Tekmar Atomx and Thermo Scientific™ GC/MS System with New Helium Saver Module
2016|Thermo Fisher Scientific|Applications
USEPA Method 524.4 Using the Teledyne Tekmar Atomx and Thermo Scientific™ GC/MS System with New Helium Saver Module Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract Increasing concerns over the availability of helium, and…
Key words
temp, tempbake, bakepurge, purgerinse, rinsetert, tertsweep, sweepether, etherdesorb, desorbcondensate, condensateatomx, atomxrinses, rinsestime, timesample, samplelpir, lpirmethanol
US EPA Method 624 – Teledyne Tekmar Atomx XYZ and Agilent 7890B GC/5977A MSD
2019|Agilent Technologies|Applications
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1902 P+T VOC ENV US EPA Method 624 – Teledyne Tekmar Atomx XYZ and Agilent 7890B GC/5977A MSD Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA Method 624 was used…
Key words
temp, tempbake, bakepurge, purgerinse, rinsesweep, sweepdesorb, desorbxyz, xyzcondensate, condensateatomx, atomxtime, timemethanol, methanolppb, ppbvariable, variableflow, flowneedle
US EPA Method 624 using the Teledyne Tekmar Atomx XYZ and Thermo Scientific™ ISQ™ 7000 Mass Spectrometry (MS) System Coupled with a Thermo Scientific™ TRACE™ 1310 Gas Chromatograph (GC)
2020|Thermo Fisher Scientific|Applications
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN2003 P+T VOC ENV US EPA Method 624 using the Teledyne Tekmar Atomx XYZ and Thermo Scientific™ ISQ™ 7000 Mass Spectrometry (MS) System Coupled with a Thermo Scientific™ TRACE™ 1310 Gas Chromatograph (GC) Amy Nutter, Applications…
Key words
temp, tempbake, bakepurge, purgerinse, rinsesweep, sweepatomx, atomxdesorb, desorbtime, timecondensate, condensatexyz, xyzteledyne, teledynetekmar, tekmarvariable, variablemethanol, methanolppb